SON_Redacted.pdf
PDF 136 KB Posted
- Attached to
- Hemispheric Scatterometer
- Solicitation number
- FA9401-24-Q0020
About this file
This document is a Statement of Need (SON) for a Hemispheric Scatterometer Instrument. The key requirements are:
The scatterometer must be capable of performing Bidirectional Reflectance Distribution (BRDF) measurements of materials at angles of incidence from 0 to 60 degrees, with a step size of 1 degree and an angular resolution of 1 degree or better. It must have three discrete light sources in the blue, green, and red wavelength ranges, and be capable of exporting BRDF data in .txt format to a computer system. The dimensions of the instrument must be 3 ft or less in width, length, and height, and it must be encapsulated in a dark box/cabinet to eliminate stray light.
This SON is associated with a formal solicitation by the Department of the Air Force Materiel Command Nuclear Weapons Center for a Hemispheric Scatterometer, with a solicitation number of FA9401-24-Q0020.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| FAQs for FA940124Q0020.pdf | ||
| FAQs for FA940124Q0020.pdf | ||
| FA940124Q0020_Combo_Sol..pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF NEED
Requirements Document
Hemispherical Scattering Measurement Instrument
1. Overview:
requires a scatterometer that will be used for locally acquiring low-cost Bidirectional Reflectance Distribution (BRDF) measurements of materials that can be compared to measurements gathered from outside facilities.
‐ Product must perform measurements at angles of incidences from 0° to ≥ 60°.
‐ The step size in change of angle of incidence must be ≤ 1°.
‐ Product must have an angular resolution ≤ 1°.
‐ Product must be capable of exporting BRDF data from the instrument to a computer system in the form of
.txt file.
‐ Product must have three discrete light sources in the blue, green, and red wavelength range.
‐ Dimensions of product must be ≤ 3 ft in width, length, and height.
‐ Instrumentation must be encapsulated in a dark box/cabinet to eliminate stray light interfering with measurements.
2. Requirements:
Item Description Required Hemispheric Scatterometer Instrument capable of measuring
BRDF measurements of materials