SON_Redacted.pdf

PDF 136 KB Posted

Attached to
Hemispheric Scatterometer
Solicitation number
FA9401-24-Q0020
Issued by

About this file

This document is a Statement of Need (SON) for a Hemispheric Scatterometer Instrument. The key requirements are:

The scatterometer must be capable of performing Bidirectional Reflectance Distribution (BRDF) measurements of materials at angles of incidence from 0 to 60 degrees, with a step size of 1 degree and an angular resolution of 1 degree or better. It must have three discrete light sources in the blue, green, and red wavelength ranges, and be capable of exporting BRDF data in .txt format to a computer system. The dimensions of the instrument must be 3 ft or less in width, length, and height, and it must be encapsulated in a dark box/cabinet to eliminate stray light.

This SON is associated with a formal solicitation by the Department of the Air Force Materiel Command Nuclear Weapons Center for a Hemispheric Scatterometer, with a solicitation number of FA9401-24-Q0020.

View the file

Other files

File Type Posted
FAQs for FA940124Q0020.pdf PDF
FAQs for FA940124Q0020.pdf PDF
FA940124Q0020_Combo_Sol..pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF NEED

Requirements Document

Hemispherical Scattering Measurement Instrument

1. Overview:

requires a scatterometer that will be used for locally acquiring low-cost Bidirectional Reflectance Distribution (BRDF) measurements of materials that can be compared to measurements gathered from outside facilities.

‐ Product must perform measurements at angles of incidences from 0° to ≥ 60°.

‐ The step size in change of angle of incidence must be ≤ 1°.

‐ Product must have an angular resolution ≤ 1°.

‐ Product must be capable of exporting BRDF data from the instrument to a computer system in the form of

.txt file.

‐ Product must have three discrete light sources in the blue, green, and red wavelength range.

‐ Dimensions of product must be ≤ 3 ft in width, length, and height.

‐ Instrumentation must be encapsulated in a dark box/cabinet to eliminate stray light interfering with measurements.

2. Requirements:

Item Description Required Hemispheric Scatterometer Instrument capable of measuring

BRDF measurements of materials