This is a Justification for Other than Full and Open Competition (JOFOC) document from the Naval Research Laboratory (NRL) for a sole source procurement of a Xeuss 3.0 or Pro HR x-ray scattering system from Xenocs Inc. under FAR 13.5 simplified procedures.
The NRL requires a high brilliance microfocus rotating anode x-ray generator system with specific capabilities including: sample probing features from 1Å to 300nm, continuous vacuum operation, Cu anode with 70μm or smaller focal spot size, support for Mo/Cr/Co anodes, automated data collection software, chamber dimensions of at least 480x375x450mm, vacuum/air/gas ambient support, sample stage with precise positioning, hybrid photon counting detector, and various sample holders and stages including GISAXS capabilities up to 900°C. Xenocs Inc. was determined to be the only vendor capable of meeting these requirements, as confirmed through market research and a Sources Sought Notice (N00173-24-SSN-TL21) posted June-August 2024. The only other potential supplier, Anton Paar, explicitly stated they cannot provide the required integrated rotating anode generator. The anticipated contract value was redacted, with delivery expected within one year of award. A Notice of Intent to sole source was posted October 2-9, 2024 with no challenges received.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
JA_13.5_template_0519_xenocs_draft pg. 2
3.1.1. Graphical user interface and ability to create scripts for automated sequential performance of commands
3.1.2. Automated system alignment and change of scattering geometry
3.1.3. Real-time display of scattering pattern during data collection
3.1.4. Correction of scattering data for cosmic rays
3.1.5. Export data to various non-proprietary file formats
3.2. Must include data analysis software with the following capabilities:
3.2.1. Comprehensive suite of tools for data reduction and analysis
3.2.2. Export of high quality, publication ready figures
4. Chamber
4.1. Must have minimum height, length, and depth dimensions of 480 mm, 375 mm, and 450 mm, respectively
4.2. Must support vacuum, air, and specific gas ambients
4.3. Must have both front and rear access doors
4.4. Must have vacuum feedthroughs to support in situ/in operando experiments
4.5. The sample area must be able to be isolated in non-vacuum, atmospheric environments with a minimized gap for samples that cannot be exposed to vacuum
4.6. Must provide visualization of sample along the beam axis with no parallax error for sample alignment
5. Sample stage assembly
5.1. Must have two translational degrees of freedom perpendicular to beam axis for sample positioning, with a minimum travel range of 100 mm per direction and resolution of 1 µm or better
5.2. Must be capable of translating loads weighing at least 4 Kg
5.3. Samples must remain in the same translational plane during measurements
5.4. Must have an automatic sample environment recognition system that can identify different sample holders (ambient holders, heaters, tensile stages, etc.) when mounted to the sample stage assembly;
vendor must provide support for integrating custom sample environments with the recognition system
6. Detector
6.1. The detector must be a vacuum compatible hybrid photon counting detector with the following capabilities/specifications:
6.1.1. Energy range covering at least 5.4 keV to 20 keV
6.1.2. Active area comprised of at least 1028 pixels x 1028 pixels with pixel size of or smaller
6.2. Detector must support programmable, fully in vacuo movement along beam axis across entire range of sample-to-detector distances without requiring sample translation or rotation
6.3. Detector must support programmable movement in both directions perpendicular to beam axis to fill in gap regions between detector panels and virtually expand measurement area by a factor of at least 2 per direction
6.4. Beamstop-free, window-less intensity measurements
7. Accessories
7.1. Must include holders for simultaneous mounting of multiple samples, including solutions, solids, and powders
7.2. Must include a grazing incidence small-angle x-ray scattering (GISAXS) stage with 3 degrees of rotational freedom and an azimuthal rotational range of at least 180 degrees
7.3. Must include a GISAXS-compatible temperature stage which can cool to cryogenic temperatures and can heat to at least 350 °C
7.4. Must include a high temperature sample stage with maximum temperature of at least 900°C
8. Services
8.1. Vendor must provide shipping, installation, and commissioning of system
JA_13.5_template_0519_xenocs_draft pg. 5 emerges, or new technology is developed which advances the state of the art and provides the capabilities NRL requires, NRL will assess whether competition for future requirements is feasible. In accordance with FAR 13.501(a)(iii), this J&A will be made publicly available on www.SAM.gov within 14 days after contract award.