Seta-PM-93-35000-0.pdf
PDF 2 MB Posted
- Attached to
- Flashpoint Tester
- Solicitation number
- W50S6Z24QA011
- Issued by
-
Department of the Army National Guard
About this file
This document appears to be a technical specification for the Seta PM-93 Automated Pensky-Martens Flashpoint Tester. The tester is designed to provide fast, accurate, and safe flashpoint analysis for a variety of petroleum and chemical products including diesel, biodiesel, heating oil, turbine fuels, lubricating oils, paints, and bitumen. Key features include electric or gas ignition, a large color touchscreen, storage for 2,000 test results, and an optional integrated fire extinguisher system. The tester complies with ASTM D93, IP 34, and ISO 2719 test methods. It is available through the federal contract opportunity W50S6Z24QA011 for the Department of the Army National Guard, which is soliciting the purchase of the Seta PM-93 tester and related accessories such as sample cups and ignitors. Funds are not yet available for this contract, and the government's obligation is contingent on future appropriations.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| W50S6Z24QA011.pdf | ||
| Parts List-Updated.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Seta PM-93 35000-0 Automated Pensky-Martens Flash Point
ASTM D93 Procedures A, B & C; IP 34 Procedures A, B & C; ISO 2719 Procedures A, B & C
Diesel • Biodiesel (FAME) • Heating Oil • Turbine Fuels Lubricating Oils • Paints & Varnishes • Residual Fuel Oils • Bitumen
• Fast, accurate & safe
• Easy operation
• Single action raising & lowering of lid
• Superior safety
• Patented ‘SafeFlash’ auto fire extinguisher
• Complies with ISO 2719 annex b
• Gas or electronic ignition
• Unique test profiles
• Large touch screen
• Memory storage for 2000 results
• Compact footprint
• Statistical Quality Control software
PM-93 is the new solution for automatic Pensky- Martens flash point determinations, providing the ultimate level of fast, accurate and safe analysis.
Designed to provide operators with a high level of functionality combined with a class leading ease of use, robustness and safety.
Flash points can be determined up to 400°C using standard mode, where the flash point is known, or search mode.
Fully compliant with international standard test methods and supplied pre-programmed with ASTM D93 test protocols for procedures A, B and C.
All routine functions are accessed by an intuitive menu system on a super-size industrial 8.4” colour touch screen. Test progress is clearly displayed, with large format characters for long range viewing.
“Safeflash” fire extinguisher system is available as a factory option for the latest ISO 2719 annex b requirements. An integrated and sealed CO cylinder punctures automatically and discharges over the sample cup and lid, eliminating the need for external gas lines.
PM-93 Statistical Quality Control (SQC) software provides detailed analysis of test results, calculated mean, standard deviation, standard error, repeatability and plot of results in accordance with ASTM D6299 Statistical Standard.
Seta PM-93 Automated Pensky-Martens Flash Point ASTM D93 Procedures A, B & C; IP 34 Procedures A, B & C; ISO 2719 Procedures A, B & C
PM-93, three easy steps
Typical Applications
Operator Interface
> Enter test details > Run test > View test results
Touch screen operation
Electric hot-wire or gas igntition
Draught screen
USB port
Single action lifting pod
Load Lower GO
CO
cylinder
(optional)
Operator Safety
• A unique ‘SafeFlash’ system automatically and immediately activates an extinguisher in the event of excess flame or fire around the test cup
• If discharged, user is prompted to re-activate the extinguisher by replacing the integral CO canister
• Additional safety features include PRT checks, safety dip before starting, over-temperature cut-out, gas flame relight, gas shut off and emergency stop
• Complies with ISO 2719 annex b
Precision and Accuracy
• Automatic flash point testing ensures test repeatability
• Unique alignment design ensures lid and cup locate perfectly every time
• Password protected ‘calibration’ mode allows verification of instrument performance
• SQC software to maintain instrument accuracy and enable product quality monitoring
Enhanced Test Throughput
• Unique test profiles enable the test to be set up and repeated quickly without re-entering the same information
• Once a test profile is stored to memory, a user can select this option without the need to program test parameters
• The lid module can be cleaned in-situ to minimise down time between tests
• Fast warm up and high capacity air cooling minimises cycle time
Ease Of Use
• Single raising or lowering action when loading or removing test cup
• Simple 3 step test operation - load, lower and go
• Intuitive user menu with colour touch screen operation
• Real time display of test progress
• Large capacity memory allows storage of test profiles, operator names, test methods and results
• Test status and results are graphically displayed
• Data format is compatible with most PC spreadsheets and a USB port allows results to be saved to portable memory devices
• QR code data transfer
Seta PM-93
ASTM D93 Procedures A, B & C; IP 34 Procedures A, B & C; ISO 2719 Procedures A, B & C
Seta PM-93
ASTM D93 Procedures A, B & C; IP 34 Procedures A, B & C; ISO 2719 Procedures A, B & C
Seta PM-93 35000-0 Technical Specifications Operation
Test modes Expected flash or search
Ignition system Electric hot-wire or gas flame
Flash detection Thermal
Cooling Forced air (integral fan)
Heating rate Fast heating mode (>10°C/min) and standard rate 5.50C/min, 3°C/min, 1.3°C/min, 1°C/min
Calibration Calibration dates/data stored to memory, easy retrieval and password protected
Application range Ambient to +5°C to +400°C
Measurement
Sample temperature PT 100 stainless steel probe
Units of temperature °C or °F (user selectable)
Barometric pressure correction
Automatic correction with built-in pressure sensor
Data Management
Information Real-time display on screen of test progress and results
Internal memory -parameters
30 programmable test profiles, sample ID’s, operator names and test methods
Internal memory - results In excess of 2000 test results
SQC ASTM D6299 Statistical Standard
Safety
Fire detection Thermal with integral CO fire extinguisher
(factory fitted option)
External fire alarm relay contact
240Vac max, 1A max resistive. External warning: 12V dc signal output
Gas Supply (optional)
Gas type Butane, propane or natural gas source
Gas supply 30mbar (3kPa)
Power Requirements
Voltage 230 Vac +/-15% or 110Vac +/-15%. Frequency 50/60 Hz. Auto sensing - no selector switch required
Power 1.1 kW
Interface Specifications
Display 8.4” LCD colour touchscreen (resistive - can be used with gloves), 800 x 600 pixel, USB keyboard, mouse and bar code scanner
Data input/output Connection to LIMS via; Ethernet RJ45, USB Type A (x2 front and back RS232C, test results can be emailed, saved to memory stick or transferred via QR code
Printer options USB, Ethernet or RS232C
Language User selectable - English (default), Russian, German
Environmental conditions
Operating temperature 5 to 35°C (50 to 104°F)
Relative humidity Up to 80% at 35°C (not condensing)
Altitude 2000m maximum
Physical
Dimensions (HxWxD) 38.5 x 24 x 50cm
Weight 25kg
Minimal Servicing
Monthly:
• Check operation of shutter and stirrer
6 Monthly:
• Service fire extinguisher system
• Pressure calibration
Yearly:
• Temperature calibration
• Instrument verification using Certified Reference Material
Part No Description
35002-0
Safeflash Fire Extinguisher System Note: Factory fit option only - must be ordered at time of purchase or the unit must be returned to the factory for installation. Comprises extinguisher, manifold assembly, regulator, fire sensor and remote operating button
35003-0 Barcode Scanner USB interface barcode scanner for 1D barcodes, connects in place of keyboard
81002-3 Printer RS 232, thermal printer, with power supply and connecting cable
99851-0 Multi-Test Verification Material, ASTM D93-IP34 Gas Oil Flash point range 56 to 80°C, 500ml
99852-0 Multi-Test Verification Material, ASTM D93-IP34 Fuel Oil Flash point range 92 to 122°C, 500ml
99853-2 Multi-Test Verification Material, ASTM D93-IP34 Lubricating Oil Flash point range 196 to 210°C, 500ml
Optional Accessories
File details come from the government source that posted it. Updated .