SEM Requirements_31 Jul 25_jro.pdf
PDF 72 KB Posted
- Attached to
- Scanning Electron Microscope
- Solicitation number
- FA4819NK0001
About this file
This document is a Requirements List for a Scanning Electron Microscope (SEM) that details specific technical and operational specifications. The requirements include a minimum magnification of 50X with a resolution of 10 nm or less at 0.5kV, Windows PC-based interface with an easy-to-use GUI, and computer-assisted optics system operation. Additional key specifications are a battery backup power supply lasting eight hours, a goniometer sample stage with Z-position tilt, stage camera with visible spectra live-imaging, and an Energy Dispersive Spectroscopy (EDS) system larger than 60mm² that is manufacturer-serviced.
The document also outlines service and support requirements, including multiple market-located service technicians, on-site manufacturer training for hardware and software, and a minimum one-year warranty covering all parts and labor. The vendor must be capable of responding to warranty calls and repairing the instrument within 20 business days. Unique technical capabilities requested include live mapping while imaging, 3D mapping/reconstruction, EDS overlay with individual elements, and the ability to operate without network connectivity for maintenance or repair.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Scanning Electron Microscope Requirements List
• Multiple service technicians located in market
• A battery backup power supply sufficient to maintain the system for a minimum of eight hours
• Magnification o Lower magnification bound of 50X or more o Resolution of 10 nm or less at 0.5kV
• All microscope functions shall be controlled through a Windows PC based interface.
o The GUI shall be easy to understand and use o Operation of optics system shall be computer assisted
• Must not require network connectivity for operation, maintenance, or repair
• Goniometer sample stage with tilt at all Z-positions
• Stage camera with visible spectra live-image of sample stage
• EDS >60mm² operated via SEM software, and serviced by SEM manufacturer technicians
• Live mapping while imaging, updating as sample is navigated
• 3D mapping / reconstruction
• EDS must show overlay with individual elements
• On-site training provided by manufacturer for both hardware and software
• Minimum of 1 year warranty that includes all parts and labor.
• Vendor has capability for technician to respond to warranty calls/repair the instrument within 20 business days of the requested service call