This is a Statement of Work for NASA Ames Research Center procuring a Short-Wave Infrared (SWIR) camera system. NASA requires one Lytid SIRIS Deep Cooled Scientific SWIR Camera or equivalent, with delivery and performance at NASA Ames Research Center, MS 230-2, Moffett Field, California 94035. The period of performance is 30 days after receipt of order (ARO).

The camera must meet specific technical specifications including a minimum 640 x 512 pixel array with a minimum image area of 6 x 6 mm and sensitivity range of 900-1650 nanometers. Required performance parameters include quantum efficiency greater than 70% at 900 nm when operated at 77K, programmable integration time with minimum gate time of 1.0 microsecond or less, rise time less than 0.1 microsecond, single shot externally triggered acquisition capability, and minimum 14-bit analog-to-digital conversion. The detector must operate at a minimum temperature of 77K using liquid nitrogen or a cryocooler, maintain a working distance of less than 1.0 inch, achieve a minimum 44 dB dynamic range, support full well capacity greater than 17,000 electrons, maintain sensor noise below 350 electrons, include multiple gain settings and a software development kit, and feature a 100% fill factor with no built-in cold filter.

View the file

Other files

File Type Posted
SATPC0042837 Tab 10 1 RFQ.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

STATEMENT OF WORK

1. Objective/Requirements

NASA Ames Research Center has a requirement for the Lytid Short-Wave Infrared Camera;

the specifics are listed below. OR EQUIVALENT

2. Characteristics, Scope, and Specs

Description OR EQUIVALENT QTY

Lytid SIRIS - Deep Cooled Scientific SWIR Camera, 0.9 - 1.7 μmLytid SIRIS -

Deep Cooled Scientific SWIR Camera, 0.9 - 1.7 μm

The specification of the NIR FPA Camera include:

•Minimum 640 x 512 pixel array

•Minimum image area of 6 x 6 mm

•Range of sensitivity to include at minimum 900-1650 nm

•Quantum Efficiency >70% at 900nm when operated at 77K

•Peak Quantum Efficiency >70%

•Programmable integration time with minimum gate time of 1.0 us or less

•Rise time less than 0.1 us

•Single shot externally triggered acquisition

•Reproducible single shot behavior with no pre-trigger

•Minimum operating temperature (LN2 or Cryocooler) of 77K

•Working distance from detector to front face of detector <1.0 inch

•Minimum 14-bit A/D conversion

•Multiple gain settings

•Software development kit

•Minimum 44 dB dynamic range

•Full well capacity >17 ke-

•Sensor noise <350 e-

•No built in cold-filter

•100% fill factor

3. Place of Performance

NASA Ames Research Center

MS 230-2 Moffett Field , CA 94035 United States

4. Period of Performance

30 Days ARO