SATPC0041949 Tab 04 4 SOW.pdf
PDF 143 KB Posted
- Attached to
- Purchase of Thermo Scientific Nicolet Apex FTIR Spectrometer with a RaptIR+ IR Microscope
- Solicitation number
- 80NSSC26939343Q
About this file
This is a Statement of Work (SOW) from NASA Ames Research Center's Center for Life Detection Facility (CLD) for the procurement of a specialized analytical instrument system.
The CLD Facility requires a high-sensitivity, cleanroom-compatible Fourier Transform Infrared (FTIR) spectrometer integrated with an FTIR microscope featuring Attenuated Total Reflectance (ATR) capability. The system will be used to measure and characterize organic contamination on facility hardware and equipment within the nested ISO-8, ISO-6, and ISO-5 cleanroom environment. Based on market research, NASA has determined that the Thermo Scientific Nicolet Apex FTIR Spectrometer with RaptIR+IR Microscope uniquely meets facility requirements, indicating a single-source procurement approach.
The FTIR spectrometer must deliver signal-to-noise better than 50,000:1, 0.25 cm⁻¹ resolution, and 7,800–350 cm⁻¹ spectral range, with full processing and analysis tools, KBR external sample beam port, and pass-through mirror assembly. The integrated FTIR microscope must include 4x and 15x objectives, 1 micron visual and 5 micron infrared spatial resolution, motorized positioning stage with sub-2-micron software control reproducibility, thermoelectrically cooled detector, 350 micron Ge Tip ATR with automated contact alert, digital color video camera, comprehensive automation software with ATR spectra library, and vendor-provided user training. Delivery is required to Building N239 Room 432 at NASA Ames Research Center in Moffett Field, California, with a desired delivery date of September 30, 2026.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
NASA Ames Research Center
Planetary Science and Exobiology Division (Code ST) Center for Life Detection Facility (CLD)
Technical POC: Richard.C.Quinn@nasa.gov
1. Objective/Requirements The ARC CLD Cleanroom Facility requires a high-sensitivity cleanroom compatible Fourier transform infrared (FTIR) spectrometer and FTIR microscope with Attenuated Total Reflectance (ATR) capability for measurement and characterization of facility and hardware organic contamination. The new CLD Facility is a nested ISO-8, ISO-6, and ISO-5 cleanroom environment designed for the development of in-situ life detection technologies and science verification. The facility and its projects have strict biological and organic chemical contamination-control requirements, which demand on-going monitoring of contamination levels to ensure that science and technology products meet required cleanliness levels. The FTIR-spectrometer and -microscope, is required for the measurement of facility, equipment, and product surface organic contamination levels. After market research, it has been determined (see attached Recommendation and Determination to Solicit from One Source) that the Thermo Scientific Nicolet Apex FTIR Spectrometer with a RaptIR+IR Microscope uniquely meets the CLD facility needs and requirements.
2. Characteristics, Scope, and Specs
The instrument must meet the following criteria:
• Compatible for use in ISO-8 Cleanroom Environments
• FTIR spectrometer integrated with FTIR microscope
FTIR Spectrometer with:
• Signal-to-noise better than 50,000:1 (1 minute)
• 0.25 cm-1 resolution
• 7,800–350 cm-1 range
• ATR crystal clean detection to provide visual indication of crystal cleanliness
• Full spectral processing and analysis tools (baseline correction, spectral math, peak area, peak height) for quantification
• KBR external sample beam port window and compartment with purge system
• Pass through mirror assembly external measurements
FTIR Microscope with:
• 4x glass refracting objective
• 15x 0.65 N.A. diffraction limited objective, with built-in dovetail mount for ATR
• 15x 0.65 N.A. diffraction limited condenser, permanently aligned and motorized
• 1 micron visual spatial resolution
• Better than 5 micron IR spatial resolution
• Infrared aperture x-y adjustments and rotation control
• Both Auto-focus and auto-illumination and manual-focus and manual-illumination
• Motorized positioning stage with software control better than 2-micron reproducibility
• Microscope integrated with FTIR microscope right side assembly configuration
• Thermoelectrically cooled (TEC) detector
• 350 micron Ge Tip ATR with dovetail mount to IR objective
• ATR pressure sensor, for automated contact alert
• <1 micron reproducibility, 40 mm working distance, 5 kg maximum loading
• Reflectance / Transmission / ATR Sampling Modes
• Digital Color Video Camera (5 Megapixels)
• Joystick for motorized positioning control
• Computer with 22" monitor
• Full infrared microscopy automation and control software with comprehensive
ATR spectra library for data analysis
• User training provided by vendor field scientist
See Thermo-Electron quote CPQ-00874867 (attached) for full specifications.
3. Place of Performance/Deliver to
Center for Life Detection Laboratory, NASA Ames Research Center, Building N239 Room 432, Moffett Field, CA 94035
4. Period of Performance
Desired delivery by September 30, 2026.
| 1. Objective/Requirements |
| 2. Characteristics, Scope, and Specs |
| 3. Place of Performance/Deliver to |
| 4. Period of Performance |