SATPC0041803 Tab 04 4 SOW.pdf
PDF 119 KB Posted
- Solicitation number
- 80NSSC26938239Q
About this file
This is a Statement of Work (SOW) for the procurement of a closed-cycle helium cryocooler system for NASA's ICEE (Integrated Cryogenic Equipment and Experimentation) facility at their Ames Research Center.
The cryocooler must achieve a minimum temperature of 9 K or lower with a maximum temperature of 355 K and temperature stability of 3 mK or less. The system requires a cooling power of 60 Watts or greater at 77K on the first stage and 4 Watts or greater at 10 K on the second stage, with an expander weight of 13.1 kg or less. The unit must operate at pressures of 10-11 Torr or lower and include a flat flange configuration with an 8-inch CF flange and four 2.75-inch CF instrumentation feedthroughs. Specific technical requirements include ¾-inch by 20-foot helium superflex hoses with 90-degree elbows, a nickel-plated OFHC copper radiation shield, and a cold finger extension flange adjustable between 10 and 24 inches. The UHV instrumentation package must include a 10-pin UHV feedthrough, wire-wrapped thermofoil heater, curve-matched silicon diode sensor at the cooler tip, and a calibrated silicon diode sensor with 4-inch free length. Vibration requirements specify a maximum of 140 nanometers at the sample with sample displacement limited to ±5 nm total displacement. The system must include an integrated temperature controller with dual sensor and heater control capabilities, IEEE-488 and USB compatibility, PID control zones with autotune, and 75 Watts heater power. All deliverables include helium superflex hoses, installation kit, technical manuals, packing, and shipping. The period of performance extends from contract award through December 20, 2026. The NASA point of contact is Dr. Andrew Mattioda at the Ames Research Center.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work for UHV Low Vibration Cryocooler
Background:
The ICEE facility is a state-of-the-art planetary science research facility consisting of three separate stations: a Raman Microscope, an Ultra High Vacuum (UHV) chamber, and a Fourier Transform Infrared and Raman Experiments (i.e., FIRE) station. The ICEE Facility is a unique NASA resource capable of replicating the wide range of extraterrestrial environments needed for planetary science research. The purchase of a low temperature, low vibration helium cryo-cooler will enable additional ice experiments,as UHV conditions are needed for accurate ice experiments. The low vibration cryostat will enable imaging of the sample during experimental runs.
Scope and Specifications:
The closed cycle cryo-cooler shall
1. Be capable of a 9 K, or lower, minimum temperature.
2. Be capable of a 355 K maximum temperature.
3. Temperature stability shall be 3 mK or less.
4. Capable of 10-11 Torr or lower pressure
5. Have an expander weight or 13.1 Kg or less, for easy manipulation.
6. Have a cooling power @ 60Hz of 60 Watts or greater at 77K, for the first stage
7. Have a cooling power @ 4 Watts or greater at 10 K, for the second stage
8. Be a flat flange configuration, with an 8 inch CF flange with four (4) 2.75” CF instrumentation feedthroughs.
9. Have a low water flow rate.
10. Have ¾” [19mm] by 20’ [6 m] helium superflex hoses (per hose pair) with 90 degree elbows for connection at the expander.
11. Have a Nickle Plated OFHC Copper terminated radiation shield that is 0.125” short of the cold tip.
12. Have a cold Finger Extension Flange to tip length of up to L = 24 inches and be adjustable to 10 inches, final length to be determined upon order.
13. The UHV (ultra-high vacuum) instrumentation for sample temperature control shall have
a. 10-Pin UHV feedthrough on a 1.33” Mini-CF flange
b. 1x 50 Ohm Wire Wrapped Durable metallized thermofoil heater.
c. 1x Curve Matched Silicon Diode sensor installed at the cooler tip for temperature control.
d. 1x calibrated silicon diode sensor installed with 4” [102 mm] free length for accurate sample temperature measurements.
14. Have vibration damping direct mounts for the interface, mounts the cryocooler directly to the interface.
15. The vibrations at the sample shall be 140 nm or less.
16. Displacement of the sample, due to vibrations, shall be +/- 5 nm (10 nm total).
17. Have a temperature controller integrated with the cryostat:
a. Interconnecting cables to the cryostat are included.
b. Temperature controller capable of dual sensor input.
c. Temperature controller has dual heater control loops.
d. Temperature controller shall be IEEE-488 and USB compatible.
e. Temperature controller shall have PID control zones, and autotune.
f. Temperature controller shall have 75 Watts heater power.
18. Shall include installation kit and technical manuals.
Statement of Work for UHV Low Vibration Cryocooler
19. Price shall include packing.
20. Price shall include shipping.
Period of Performance:
The period of performance shall be from award to 12/20/2026.