SATPC0037950 Tab 04 4 SOW..pdf
PDF 126 KB Posted
- Attached to
- Mensor CPG2500 Precision Pressure Indicator & CPR2550 Pressure Sensor Federal contract opportunity
- Solicitation number
- 80NSSC25895843Q
About this file
This Statement of Work (SOW) details NASA Glenn Research Center's requirement for the procurement of Mensor precision pressure measurement equipment for the 10x10 Supersonic Wind Tunnel. The specific equipment includes four (4) CPG2500 Precision Pressure Indicators, ten (10) CPR2550 Standard Pressure Sensors, three (3) CPT6180 Precision Pressure Transducers, and four (4) accessories for pressure controllers and measuring devices.
The SOW specifies detailed technical requirements for each equipment type, including compatibility with Mensor systems, communication interfaces (RS-232, IEEE-488, USB, Ethernet), accuracy specifications ranging from 0.008% to 0.01%, and support for multiple pressure ranges (15, 30, 45, and 115 psia). The accessories must be compatible with the Mensor CPG2500 and standard 19" server rack. Performance is to be completed within 8 weeks after receipt of order (ARO) at the NASA Glenn Research Center located in Cleveland, Ohio.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| SATPC0037950 Tab 10 RFQ Open Market Template (2).pdf | ||
| SATPC0037950 Tab 06 BNJ..pdf | ||
| SATPC0037950 Tab 10 RFQ Open Market Template.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
1. Objective/Requirements The NASA Glenn Research Center (GRC) requires the purchase of Mensor CPG2500 Precision Pressure Indicator, CPR2550 Standard Pressure Sensor, CPT6180 Precision Pressure Transducer & Accessories which will be used at the 10x10 Supersonic Wind Tunnel at GRC.
2. Characteristics, Scope, and Specs
Four (4) Mensor CPG2500 Precision Pressure Indicators
Salient Specifications:
• The vendor shall provide indicators compatible for use with Mensor CPR2550 precision pressure transducers of various ranges.
• The vendor shall provide indicators compatible for use with Mensor CPT6180 precision pressure transducers of various ranges.
• The vendor shall provide indicators that will contain an on-board barometric pressure standard with accuracy of 0.01% rdg or better.
• The vendor shall provide indicators that will support simultaneous 3 channel pressure readings.
• The vendor shall provide indicators that will support transducer accuracies up to
0.008% of IS and precision of up to 0.004% FS.
• The vendor shall provide indicators that will have color touchscreen.
• The vendor shall provide indicators that will support communication via RS-232, IEEE-488, USB or Ethernet.
• The vendor shall provide communication commands and queries that will be consistent with Mensor command set.
Ten (10) Mensor CPR2550 Standard Pressure Sensors
Salient Specifications:
• The vendor shall provide sensors that will be compatible with Mensor CPG2500.
• The vendor shall provide sensors that will be able to interface with internal ports and connectors in chassis of Mensor CPG2500.
• The vendor shall provide sensors that will be available in various ranges including
15, 30, 45, and 115 psia.
• The vendor shall provide sensors that will be accurate to 0.008% IS 50 or better.
Three (3) Mensor CPT6180 Precision Pressure Transducers
Salient Specifications:
• The vendor shall provide transducers that will be compatible with Mensor
CPG2500.
• The vendor shall provide transducers that will be able to interface with remote sensor connector on rear of Mensor CPG2500 chassis.
• The vendor shall provide transducers that will be available in various ranges including 15 psia.
• The vendor shall provide transducers that will be accurate to 0.01% IS 50 or better.
Four (4) Mensor Accessories for Pressure Controller and Pressure Measuring Devices
Salient Specifications:
• The vendor shall provide mounting kits that will be compatible with Mensor
CPG2500.
• The vendor shall provide mounting kits that will be compatible with standard 19” server rack.
3. Place of Performance NASA Glenn Research Center (GRC) Cleveland; OH 44135
4. Period of Performance 8 Weeks ARO
File details come from the government source that posted it. Updated .