This document is a Statement of Work (SOW) for the procurement of a Dual Channel CAPS NOx instrument. The instrument must have a measurement range of 0-1,000 ppbv NOx and NO2, noise less than 0.5 ppbv, physical response time less than 2 seconds, and a sample flow rate of 2.5 lpm total. The instrument must include RS232, USB, and Ethernet connectivity, and meet specific power requirements.
The related federal contract opportunity is a Special Notice for the procurement of the Aerodyne Cavity Attenuated Phase Shift (CAPS) Monitor instrument, with the Statement of Work provided in the attached document. The Solicitation Number is 80NSSC24878899Q, and the Contracting Agency is the National Aeronautics and Space Administration Shared Services Center.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
1. Objective/Requirements
The information listed below is required to complete this purchase.
2. Characteristics, Scope, and Specs
LINE PART NO. DESCRIPTION QTY.
1 1000-6486 Dual Channel CAPS NOx instrument: (NOx + NO2)
• Range: 0-1,000 ppbv NOx and NO2
• Noise (2σ): < 0.5 ppbv NOx and NO2 in 1 second integration time
• Physical response time (10-90%): < 2 seconds
• Sample Flow Rate: 1.25 lpm per channel (2.5 lpm total)
• Included I/O: RS232, USB, Ethernet
• Power requirements: <100 W; 100-250 VAC (50-60 Hz)
• Dimensions (LxHxW): 24" x 8.75" x 17" (5U 19” rack mount); 61cm x 22.2cm x
43.2cm
• Weight: ~40 lbs (20 kg)
• Exposed materials: Stainless steel, PFA, and Nafion
Included items:
• Front panel display
• Nafion gas dryer
• Internal scrubber, including Purafil medium and 3-way valve (eliminates need for
NOx-free air supply)
• Baseline trigger input and baseline sync signal output
• Laptop for instrument control with custom DAQ program
• On-board data storage ~5 years at 1 second sampling
• Reusable hard case for instrument shipping
Attributes:
• Autonomous operation
• Data stream includes all data underlying calculation of final product
• • Baseline: Automated or manual trigger
2 1000-5174 1000-5174 CAPS Two year extended warranty
Two additional years (three years total) of Service Warranty Coverage (excluding pump and laptop) can be purchased at an additional cost. This extended warranty does include a 50% discount on a pair of replacement mirrors, if damaged during use.
3. Place of Performance
NASA Langley Research Center
21 Langley Blvd.
Hampton, VA 23681
4. Period of Performance
Estimated delivery lead-time is 4 months ARO.