Salient.pdf
PDF 138 KB Posted
- Attached to
- PROBE, EDDY CURRENT
- Solicitation number
- SPMYM226Q0108
About this file
This document is a specification of eddy current probes and related components required for federal contract opportunities JML 61596930-39 and 61606930.
The procurement specifies 11 distinct eddy current probe models with varying specifications for conducting bobbin and array inspection on condenser tubing. All probes must feature a 36-pin Amphenol connector and be compatible with the existing Eddy Current Acquisition System. The required probes include: one CXB4-024 straight section array bobbin probe (0.490" diameter), one XP12-005 flexible model array bobbin probe (0.480" diameter), two CXB4-012 straight section probes (0.570" diameter), one XP12-012 flexible model probe (0.560" diameter), one CXB6-001 mag biased array probe (0.490" diameter), one CXB6-006 mag biased array probe (0.570" diameter), one CXB4-032 straight section probe (0.480" diameter), one CXB6-008 mag biased probe (0.480" diameter), and six CXB4RPF-001 straight section array bobbin probes (0.490" diameter). Additionally, the procurement requires 30 replaceable centering feet kits (RPFCXB-001) and 12 replaceable feet captivation screws (P/N 70002884). All probes are 50 feet in length with Poly Type 3/8T cable configuration. The manufacturer must perform quality verification checks on each probe to ensure proper operational capability.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 26Q0108 PROBE EDDY CURRENT.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Eddy Current Probes Salients
JML 61596930-39 and 61606930
The probes listed below will be utilized to perform eddy current bobbin and array inspection on condenser tubing. Manufacturer shall ensure a manufacturing quality check is performed on each probe to ensure proper operation. All probes will have a 36 pin Amphenol connector and be capable of being detected and identified by the existing Eddy Current Acquisition System.
1 each P/N CXB4-024 – Straight Section CXB4 Array Bobbin Probe, Titanium Sleeve and Centering Feet, Diameter: 0.490”, used to inspect tube ID of 0.527”, Coil: 2x14 Array/Mid Frequency Bobbin; Poly Type 3/8T, Length: 50’
1 each P/N XP12-005 – Flexible Model XP12 Array Bobbin Probe, Diameter: 0.480”, used to inspect tube ID of 0.527”, Coil: 2x14 Array/Mid Frequency Bobbin; Poly Type 3/8T, Length: 50’
2 each P/N CXB4-012 – Straight Section CXB4 Array Bobbin Probe, Titanium Sleeve and Centering Feet, Diameter: 0.570”, used to inspect tube ID of 0.620”, Coil: 2x16 Array/Low Frequency Bobbin; Poly Type 3/8T, Length: 50’
1 each P/N XP12-012 – Flexible Model XP12 Array Bobbin Probe, Diameter: 0.560”, used to inspect tube ID of 0.620”, Coil: 2x16 Array/Low Frequency Bobbin; Poly Type 3/8T, Length: 50’
1 each P/N CXB6-001 - Straight Section CXB6 Mag Biased Array and Mag Biased Bobbin Probe, Titanium Sleeve and Centering Feet, Diameter: 0.490”, used to inspect tube ID of 0.527”, Coil: 2x14 Array/Low Frequency Bobbin; Poly Type 3/8T, Length: 50’
1 each P/N CXB6-006 - Straight Section CXB6 Mag Biased Array and Mag Biased Bobbin Probe, Titanium Sleeve and Centering Feet, Diameter: 0.570”, used to inspect tube ID of 0.620”, Coil: 2x16 Array/Low Frequency Bobbin; Poly Type 3/8T, Length: 50’
1 each P/N CXB4-032 – Straight Section CXB4 Array Bobbin Probe, Titanium Sleeve and Centering Feet, Diameter: 0.480”, used to inspect tube ID of 0.527”, Coil: 2x14 Array/Mid Frequency Bobbin; Poly Type 3/8T, Length: 50’
1 each P/N CXB6-008 - Straight Section CXB6 Mag Biased Array and Mag Biased Bobbin Probe, Titanium Sleeve and Centering Feet, Diameter: 0.480”, used to inspect tube ID of 0.527”, Coil: 2x14 Array/Mid Frequency Bobbin; Poly Type 3/8T, Length: 50’
6 each P/N CXB4RPF-001 – Straight Section CXB4/RPF Array Bobbin Probe, Titanium Sleeve and Centering Feet, Diameter: 0.490”, used to inspect tube ID of 0.527”, Coil: 2x14 Array/Low Frequency Bobbin; Poly Type 3/8T, Length: 50’
30 each P/N RPFCXB-001 – replaceable centering feet kit for CXB2/RPF, CXB4/RPF or CXB6/RPF Probes, Diameter: 0.490”, used to inspect tube ID 0.527”
12 each P/N 70002884 - Replaceable feet captivation screw for CXB4 probes used in a 0.527" ID tube
File details come from the government source that posted it. Updated .