This document is a Salient Characteristics Commercial Availability Certification that defines the minimum technical requirements for a Ferroelectric Test System to be procured under the related federal contract opportunity.
The key details are:
The contractor shall provide a Ferroelectric Test System with the following minimum characteristics: frequency range to 250 kHz, minimum pulse widths of 1 microsecond, output range of +/-30V, polarization measurement under AC and DC magnetic fields, 18-bit resolution, capture rate of 0.5 microseconds, simultaneous capture of polarization and output voltage, charge sensitivity as low as 1 femtocoulomb, hysteresis loop frequency up to 250 kHz, minimum of two channel options for controlling high voltage amplifiers, provision of two external inputs with a range of +/- 10V, and user interface software. Shipping and handling to the Adelphi Lab Center is also required. The related federal contract opportunity is a Total 100% Small Business Set-Aside for One (1) Ferroelectric Test System, solicitation number W911QX24Q0151.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Combination Synopsis-Solicitation_W911QX24Q0151 Final.docx | DOCX document | |
| Provisions and Clauses_W911QX24Q0151.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics
Commercial Availability Certification:
I have evaluated the commercial marketplace and determined that all of the salient characteristics documented below are available in products/services offered in the commercial marketplace.
Signature (typed name): ______Wendy L. Sarney___________
Technical Point of Contact (TPOC) or Contracting Officer’s Representative (COR)
The Contractor shall provide the following minimum characteristic for one (1) Ferroelectric Test System:
1. Frequency range extending to 250 kilohertz (KHz).
2. Minimum pulse widths of one (1) microsecond.
3. Output range spanning +/-30 volts (V).
4. Polarization measurement under both alternating current (AC) & direct current (DC) magnetic fields.
5. 18-bit resolution (analog to digital converters).
6. Capture rate of 0.5 microseconds with 0.1 microsecond interlace facility.
7. Simultaneous capture of polarization and output voltage.
8. Charge sensitivity as low as one (1.00) femtocoulomb (fC).
9. Hysteresis loop frequency reaching up to 250 kHz.
10. Minimum of two (2) channel options for controlling high voltage amplifiers.
11. Provision of two (2) external inputs with a range of +/- ten (10) V.
12. Inclusion of user interface software.
13. Shipping and handling to the Adelphi Lab Center.