Salient Characteristics - Oligonucleotide.doc
DOC document 77 KB Posted
- Solicitation number
- W911QX25Q0055
About this file
This is a Salient Characteristics document outlining technical specifications for two oligonucleotide-related systems being procured by the Department of the Army Materiel Command. The procurement is structured as a small business set-aside.
The requirements specify one oligonucleotide cleavage and deprotection oven/chamber that must feature programmable temperature control up to 85°C, user-controlled heat and pressure settings, compatibility with various deprotection methods (ammonia-based and ammonia-free), and input/output gas controls. Additionally, one oligonucleotide purification system is required that can process eight different oligonucleotides per run cycle, dispense from five reagent bottles, utilize reusable commercial purification columns, and allow flexible programming of purification conditions. Both systems must include user training (onsite or virtual), complete documentation, and shipping/installation services.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Combination Synopsis-Solicitation for Commercial Purchases (Under SAT) - Oligonucleotide.docx | DOCX document | |
| Adelphi Local List of Special Instructions - Ogligonucleotide.doc | DOC document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics
Commercial Availability Certification:
I have evaluated the commercial marketplace and determined that all of the salient characteristics documented below are available in products/services offered in the commercial marketplace.
Signature (typed name): To Be Announced After Award Technical Point of Contact (TPOC) or Contracting Officer’s Representative (COR)
The contractor shall provide the following requirements for the systems, as described below:
One (1) Oligonucleotide cleavage and deprotection oven/chamber shall include the following Salient characteristics:
· Programmable control of chamber temperature of up to 85 degrees Celsius.
· Must facilitate the timed control of reaction conditions.
· Must allow for user control of heat and pressure within the reaction chamber.
· Must be compatible with oligonucleotide cleavage and deprotection via standard methods under a variety of conditions including ammonia-based, ammonia-free, and liquid deprotection.
· Must include input and output gas controls for the reaction chamber.
· Onsite or Virtual user training including complete user manuals, available standard protocols, and routine maintenance procedures.
· Shipping and Installation included.
One (1) Oligonucleotide purification system shall include the following Salient characteristics:
· Must be able to purify at least eight (8) different oligonucleotides per run cycle.
· Must be able to dispense from at least five (5) attached reagent bottles.
· Must automatically dispense reagents to the purification columns or cartridges to facilitate the purification of oligonucleotides.
· Must utilize reusable commercially available oligonucleotide purification columns or cartridges to purify oligonucleotides to high purity and quality.
· Must allow for flexible programming of purification conditions.
· Must allow for desalting, purification and collection of synthetic oligonucleotides.
· Onsite or Virtual user training including complete user manuals, available standard protocols, and routine maintenance procedures.
· Shipping and Installation included.