This document is a Salient Characteristics statement for an inhaled nitric oxide (iNO) therapy system procurement by Brooke Army Medical Center. The system serves as a selective pulmonary vasodilator intended to improve oxygenation and blood flow in patients, particularly those with pulmonary hypertension, with the primary goal of reducing reliance on extracorporeal membrane oxygenation (ECMO).

The required system must be FDA-approved for treating term and near-term neonates (greater than 34 weeks gestation) with hypoxic respiratory failure associated with pulmonary hypertension. Key technical specifications include: operator-determined nitric oxide concentration that remains constant throughout the respiratory cycle; continuous monitoring of inspired nitric oxide (NO), nitrogen dioxide (NO2), and oxygen (O2) levels; monitoring capability for patient partial pressure of oxygen in arterial blood (PaO2) and methemoglobin levels; prevention of NO2 concentrations exceeding 20 ppm; compatibility with additional ventilatory support therapies; backup battery power supply; and a reserve nitric oxide delivery system for contingency use. The nitric oxide gas must be supplied as a blend of nitric oxide and nitrogen in concentrations such as 800 ppm (0.08% NO, 99.92% N2) or 100 ppm (0.01% NO, 99.99% N2), delivered in aluminum cylinders as compressed gas at 2000 pounds per square inch gauge (psig). Specifically, size 88 cylinders must contain at least 1963 liters at standard temperature and pressure (delivering 1918 liters) of 800 ppm nitric oxide.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SALIENT CHARACTERISTIS

INHALED NITRIC OXIDE

The following outlines the essential characteristics for an inhaled nitric oxide (iNO) therapy system for use at the Brooke Army Medical Center. The primary purpose of this system is to act as a selective pulmonary vasodilator to improve oxygenation and blood flow in patients, particularly those with pulmonary hypertension.

List of Characteristics

• The system shall improve patient oxygenation and reduce the need for extracorporeal membrane oxygenation (ECMO).

• Be approved for treating term and near-term neonates (>34 weeks gestation) with hypoxic respiratory failure associated with pulmonary hypertension.

• The inhaled gas Shall act as a vasodilator.

• Compatibility Must be usable with additional ventilatory support therapies to maximize patient oxygen delivery.

• Shall allow for operator-determined concentration of nitric oxide, which must remain constant throughout the respiratory cycle.

• Shall be able continuously monitor inspired levels of Nitric Oxide (NO), Nitrogen

Dioxide (NO2), and Oxygen (O2).

• Shall facilitate the monitoring of patient’s partial pressure of oxygen in arterial blood

(PaO2) and methemoglobin levels during administration.

• Shall prevent the generation of inhaled Nitrogen Dioxide (NO2) in concentrations exceeding 20 ppm.

• This shall include a backup battery power supply and a reserve nitric oxide delivery system for use in case of system or power failure.

• Shall meet all U.S. Food and Drug Administration (FDA) required specifications for this type of medical device.

• Shall be a blend of nitric oxide and nitrogen, available in concentrations such as 800 ppm (0.08% NO, 99.92% N2) or 100 ppm (0.01% NO, 99.99% N2), administered by inhalation.

• It shall be supplied in aluminum cylinders as a compressed gas under high pressure (2000 pounds per square inch gauge [psig]).

• Shall be available in a size 88 cylinder (or equivalent) containing at least 1963 liters at STP (delivering 1918 liters) of 800 ppm nitric oxide."