Salient Characteristics Commercial Availability Certification:
I have evaluated the commercial marketplace and determined that all of the salient characteristics documented below are available in products/services offered in the commercial marketplace.
The Government requires one (1) large aperture scintillometer, in accordance with the below minimum specifications:
· Suitable for a measurement range that includes point five to five (0.5 - 5.0) kilometers (km), without saturation of scintillation.
· Capable of measuring atmospheric turbulence, heat flux, and crosswind over the above range of point five to five (0.5 - 5.0) km.
· Capable of measuring the inner scale of turbulence or performing separate wind measurements.
· Contain one or two transmitters that have light-emitting diode (LED) arrays for easy alignment and a reciever that is equipped with a means of sighting the transmitter(s) for the purpose of alignment.
· Scintillometer transmitter(s) and receiver to be water resistant, stable against winds up to 25 miles per hour (mph), capable of operating in outdoor temperatures ranging from 30 to 95 degrees.
· Includes a tripod stand for both the transmitter and reciever.
· Delivery to include shipping FOB destination.