This document is a Salient Characteristics Brand Name or Equal (BNOE) specification for a digital fundus camera system, specifically identifying the Nikon Optos Silverstone P200TxE as the reference equipment (M5510). The specification establishes mandatory minimum requirements for image capture, including nine imaging modalities with ultra-wide field (UWF) capability and integrated Swept Source OCT technology to facilitate comprehensive retinal-vitreous-to-sclera examination.
The camera must operate non-mydriatic with a minimum 2mm pupil and deliver UWF imaging across eight modalities: Color RG, Color RGB, Sensory (red-free), Choroidal, Autofluorescence Green, Autofluorescence Blue, Fluorescein, and Indocyanine Green. Specified laser wavelengths include 488 nm (blue), 532 nm (green), 635 nm (red), and 802 nm (infrared), with exposure time under 0.4 seconds and optomap resolution of 20 µm standard or 14 µm for optomap plus. OCT imaging must utilize Swept Source technology at 1050 nm wavelength with axial resolution less than 7 microns, transverse resolution less than 20 microns, scan depth up to 2.5 mm, and A-Scan rates up to 100k cycles per second. The unit must include a table with space requirements of 887 mm width by 600 mm depth with height adjustable from 725 to 1205 mm. The camera system itself measures 23"D x 22"W x 28"H, with a quantity requirement of one unit.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C24W26Q0147_3.pdf | ||
| 36C24W26Q0147_2.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SALIENT CHARACTERISTICS BRAND NAME OR
EQUAL (BNOE)
M5510 – Camera, Fundus, Digital – Nikon, Optos Silverstone P200TxE
a) Unit must include minimum Image Capture Features:
a. Nine imaging modalities and UWF with integrated Swept Source OCT, facilitate detailed examination of the retina-vitreous to sclera
b. UWF guided, Swept Source OCT, images pathology anywhere on the optomap
c. 1050 nm OCT light source, provides deeper tissue penetration for clear, detailed choroidal imaging
d. Non-mydriatic, with a minimum 2mm pupil.
e. cSLO imaging through most cataracts.
f. 4-in-1 Color Depth ImagingTM provides important clinical data from the retinal surface through the choroid
b) UWF Imaging:
a. Imaging Modalities: Color RG, Color RGB, Sensory (red-free), Choroidal, Autofluorescence Green (Green AF), Autofluorescence Blue (Blue AF), Fluorescein (FA), Indocyanine Green (ICG)
b. plus:
c. Laser Wavelengths: Blue Laser: 488 nm (for FA), Red laser: 635 nm, Green laser: 532 nm (for AF), Infra-red: 802 nm (for ICG)
d. Exposure Time: Less than 0.4 seconds
c) OCT Imaging:
a. Signal Type: Optical scattering from tissue
b. Signal Source: Swept Source OCT, Wavelength 1050 nm
c. Axial Resolution: < 7 micron
d. Transverse Resolution: < 20 micron
e. Scanners: Galavanometric X, Y pair
f. Scan Depth: Up to 2.5 mm
g. A-Scan Rate: Up to 100k cycles/sec
h. Scan Types: Line Scans Width: 6, 14, 23 mm; Volume Scan Height: Min 3.5 mm; Max 9 mm; Volume Scan Width: Min 6.0 mm; Max 14 mm
d) Must include table (Space Requirements: Width: 887 mm / 35 in, Depth: 600 mm / 24 in, Height:
725 to 1205 mm / 29 - 48 in)
e) Dimension: 23”D x 22”W x 28”H
f) Quantity: 1