Salient Characteristics.docx
DOCX document 48 KB Posted
- Solicitation number
- 36C26125Q0009
About this file
This document is a "Brand Name or Equal Worksheet" that outlines the salient characteristics for a Sit-Stand Adjustable Height Workstation for Telemetry Viewing at the San Francisco VA Medical Center.
The key details include:
- The workstation must have separate, independent, electrically adjustable surfaces for the monitor and keyboard that meet specific height requirements. The surfaces must be stable, able to support heavy loads, and have safety features like collision detection.
- The workstation must include a parabolic monitor rack with adjustable focal length, cable management, and quick-release mounts. It must also have a keyboard surface data connection center, acoustical panel system, CPU enclosures, and optional rotating resource guide.
- The workstation must be GREENGUARD Gold certified for indoor air quality.
- This is a sources sought notice to determine if vendors can meet these requirements for a future solicitation. Interested parties should provide capability statements and pricing information by August 12, 2024. The NAICS code is 337214 with a small business size standard of 1,100 employees.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C26125Q0009.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Brand Name or Equal Worksheet (“Salient Characteristics”) v 2.0
Brand Name Item: Sit-Stand Adjustable Height Workstations for Telemetry Viewing
1) Essential/significant physical, functional, or performance characteristics.
1. Adjustable Height Workstations
1. Stability – Function
1. Requires completely separate independent electric floor supported, adjustable work surfaces for both monitor and keyboard.
1. Console must demonstrate stability at full extension. Maximum deflection of ½” is allowed when a horizontal load of 100 lbs. is applied to the center of each work surface.
1. Each work surface (input and monitor) shall have a separate lifting equipment weight capacity of 300lbs. minimum which does not include weight of worksurface.
1. Console must meet Operator Clearance requirements of ANSI/HFES 100-2007 page 79 section 8.3.2.1.1 Method 2
1. Test requirements: Respondent shall include copies of independent test laboratory results indicating compliance with ANSI/BIFMA X5.5 – 2014.
1. Keyboard/Input Surface - Sitting/Standing
Keyboard Surface height requirements are considered paramount to proper ergonomic positioning.
1. Keyboard surface shall have static load capacity of 500 lbs to prevent damage from users sitting / leaning on or using the surface as an aid in standing.
1. Keyboard surface must lower to 5th percentile seated female elbow height dimension (22") according to ANSI/HFES 100 -2007 8.3.2.4.3, Page 85 from the floor to the top of the keyboard surface.
1. Keyboard surface must raise to at least standing elbow height for 95th percentile male user (46.5") from the floor to the home row of keyboard. ANSI/HFES 100 -2007 8.3.2.4.3, Page 85
Keyboard Surface width must accommodate multiple keyboards or other input devices and still provide room for note taking on either right or left hand side
1. Keyboard surface must be separately electrically adjustable with electronic controller having digital readout in one (1) cm increments to enable precise position replication and must have provisions for connection to network/computer for software used to control console from computer/device.
1. Monitor Surface - Sitting/Standing
Monitor Surface height requirements are considered paramount to proper ergonomic positioning.
1. Monitor Surface must lower to allow positioning of monitor so that the gaze angle of the 5th percentile seated female to the center of the screen ranges between -15° and -25° from horizontal eye level. (ANSI/HFES100 section 8.3.2.2 – 8.3.2.3 pages 82-83)
1. Monitor Surface must raise to 48" to allow positioning of monitor so that the gaze angle of the 95th percentile standing male to the center of the screen ranges between -15° and -25° from horizontal eye level (69¼”) using a 22" monitor (ANSI/HFES100 section 8.3.2.2 – 8.3.2.3 pages 82-83)
1. Monitor Surface must be wide and deep enough to accommodate up to 8 GE Carescape central stations monitors, and 1 VA OIT Imaged computer, or the number/size specified in equipment inventory. If monitors are stacked, monitor surface must be able to lower below the keyboard surface to maintain proper viewing angles for most users.
1. Monitor surface must be separately electrically adjustable with electronic controller having digital readout in one (1) cm increments to enable precise position replication and must have provisions for connection to network/computer for software used to control console from computer/device.
1. Monitor surface must be engineered to anticipate the possibility of technicians needing to stand or kneel on console to service equipment and shall be capable of supporting a 500 lb. static load.
1. Adjustment Devices
1. Adjustment speed to be minimum 1.5"/second.
1. Independent keyboard surface to be separately adjustable with electronic controller.
1. Manually operated controller shall not be located where it can be damaged by chair arms. Location under front edge of keyboard surface is unacceptable. Option of digitally integrated control system to provide secondary source for height adjustment controls. System to have ability to save unlimited number of users’ desired heights and personal control adjustments.
1. Safety finger clearance of 1½” minimum between stationary returns and moving surfaces or between moving surfaces of double surface tables is required.
1. Controller system shall include collision detection technology which will detect sudden changes in load to identify obstructions in the path of the moving surface. Collision detection technology to cause work surface to stop on detection of obstruction and reverse direction approximately 10cm to avoid entrapment of obstruction. Collision detection shall function in both upward and downward directions.
1. Laminate – Console Tops and Side Surfaces
1. Keyboard, Monitor and fixed height surfaces must be non-glare, 3-D Laminate.
1. Edge shall be continuous from top through to the bottom of the surface.
1. Front edge shall be a chamfer style edge.
1. No seams between laminate and edge will be acceptable.
1. No edge banding or T-Mold edging will be acceptable.
1. Laminate surfaces to be fully balanced construction with Greenguard certified laminate on top surface and Greenguard certified backer sheet on bottom.
1. Backer sheet on the bottom of the surface must be a light color to improve visibility for technicians.
1. Laminate – CPU Cabinets, Drawer Pedestals and other casework
1. Casework shall be Laminate to be Thermofused Melamine (TFM) in standard colors, 3/4" thickness finished both sides or approved equal. Doors will be 3-D Laminate with continuous face through edge material.
1. Cut edges, where exposed, shall be covered with edge banding, T-mold or sprayed sealant to encapsulate particle board core and limit formaldehyde emissions.
1. Console Control Electronics
1. All electronic control boxes which power the height adjustment of the surfaces shall be mounted in easily accessible location for trouble shooting. Dispatchers and service technicians shall not be required to crawl under consoles and look up at the bottom of the surface.
1. All console control electronics shall be labeled for easy identification. This includes cables, electric boxes and pathways.
1. Drawer Pedestals
1. Drawer Configuration - 1 File drawer (12”) and 1 File drawer (12") - 16” w x 28”h x 22”d. Optional drawer configurations / depths to be available.
1. All drawers to have full extension 100lb rated - soft close - steel ball bearing drawer slides.
1. Drawer pedestals shall have gang locking mechanism with master keyed removable core locks.
1. Each pedestal to include pencil tray insert and side filing conversion bar with capability to hang letter or legal-size hanging files.
1. Test requirements: Respondent shall include copies of independent test laboratory results indicating compliance with ANSI/BIFMA X5.9 – 2012..
1. Acoustical Panel System – Specified to provide for cable management, visual separation of tasks and both sound barrier and sound absorptive functions. Consoles without panel divider systems do not meet base bid requirements.
1. Panel frames shall be minimum 14 ga. cold rolled steel – with powder coat paint finish and slotted uprights to support components at 1” centers or equal.
1. Internal Cable management within the panel frame system is required. Please state your panel frame internal CAT5e cable capacity.
1. Stackable panel frames - Panel heights shall be vertically modular - The system shall be constructed in a manner to allow additional 18” segments to be “stacked” on base panel frames to change panel heights for future change or reconfiguration. “Stackable” components shall meet all specifications of 3.1.10
1. Panel Top Caps - Flush mounted design to be removable without tools. Top caps shall be available in standard powder coated aluminum.
1. Panel Segments are to be user removable/replaceable without tools. Segments construction - 22 ga. Min. cold rolled steel casing - Class A interior finish flame spread/smoke developed certification.
1. Acoustical panel construction – All panel segment tiles above the worksurface height shall be of acoustical construction. Acoustical panels shall have 22 ga. min. Cold rolled steel casing with ½” compressed formaldehyde-free fiberglass insert and shall have a minimum .55 NRC (noise reduction coefficient) rating and a Class A flame spread/smoke developed certification.
1. Optional Premium acoustical panel construction –Tiles shall be covered with specified fabric and located as indicated on the drawings. From an independent, certified laboratory according to ASTM procedure C-423-08 shall be included with RFP in appendix D compliance table and shall have a minimum .75 NRC (Noise Reduction Coefficient) rating.
1. Test Requirements: Respondent shall include copy of independent test lab results indicating compliance with E84-09c Surface Burning characteristics of Building Materials in Appendix D.
1. Optional Panel segment types shall be - Powder coat paint finish, fabric over steel, airflow, and clear or frosted glass.
1. Test requirements: Respondent shall include copies of independent test laboratory results indicating compliance with ANSI/BIFMA X5.6 – 2016. Attach test results to Appendix D Compliance Table.
1. Shelving / Rackmount Enclosures
1. Optional An open shelf shall be located under the front edge of the Monitor Surface for material / equipment storage in the primary reach zone. This shelf can be converted to a rackmount for electronic equipment mounting.
1. Optional Desktop E.I.A. 19” Rackmount enclosures shall be available in a variety of heights from 3U up to 15U (1U = 1.75”)
1. Rackmount Enclosures shall have ventilated back panels for heat dissipation.
1. Cable Management
1. Cable pathways must be easy for the tech to access from the front of the console. Consoles which require rear access will not be considered.
1. Monitor and keyboard cables must have separate pathways from the computer to the end point.
1. Cable entry path from computer cabinet to the console must have opening large enough for all cables and a hand to fit through. Minimum of 2.5” in height by 10” wide.
1. Cable bridge shall support cables from cabinet to console. Must have separate pathways for low voltage and high voltage.
1. Cabling shall be guided from CPU cabinet or panel enclosure to the monitor surface of the adjustable table in an energy chain with easy flip-up cable channel access.
1. Keyboard cabling shall be guided from cable bridge through energy chain to keyboard surface.
1. Cabling shall be guided through a 3rd energy chain from the back of the monitor surface to the focal depth platform to keep cables organized during focal depth adjustments.
1. “J” Channel under the monitor surface shall have enough internal room to hold all the cables and any power transformers. “J” channel around back of monitor surface aligned with grommets for management of cables / transformers and cable connections.
1. Supplier shall provide premium quality extension cables as required to connect monitors, keyboards, mice and all devices to CPUs.
1. CPU Enclosures
1. CPU enclosures shall be available in sizes to accommodate mini-tower cases up to 8” wide x 18” high and be available in 1, 2, 3, 4, 5, or 6 unit capacity.
1. CPU enclosures shall be a minimum of 29” deep (deeper is preferred) in order to accommodate CPUs and cabling.
1. CPU enclosure shall have at a minimum of two of the three front, rear and top access points depending on layout.
1. Work surface height CPU enclosures shall have option for the surface top to be hinged at the back and fold up to a safe resting position against the panel system. This provides top access to the back of the computers.
1. CPU enclosures shall be available in vertical “technology tower” or horizontal under the work surface configurations.
1. CPU enclosures shall have an internal motion sensing LED service light as standard equipment.
1. CPU enclosures shall be equipped as standard with active ventilation using a minimum of two (2) quiet, 28db (decibel) 45 cfm exhaust fans to keep electronic equipment cool.
1. CPU enclosures shall not be located underneath a height adjustable surface due to potential crush zones.
1. CPU enclosures shall be tested for strength and durability to ANSI/BIFMA X5.9-2012. Respondent shall include copies of independent test laboratory results indicating compliance with ANSI/BIFMA X5.9 – 2012.
1. Supplier shall provide premium quality extension cables as required to connect monitors, keyboards, mice and all devices to CPUs.
1. Rotating Resource Guide
1. Shall be available in 36” or 41” diameter.
1. Capacity to hold 2 levels of 3 ring binders.
1. Shall have a weight capacity of 500 pounds.
1. Optional rotational stop and continuous cable pathway shall be available.
1. Optional paper form shelving shall be available.
1. Optional Casters shall be available
1. Rotating Resource Guide shall have the ability to be positioned under a peninsula top style work surface.
1. Personal Controls
1. Personal controls shall operate fan, LED task lights, heaters, LED RGBW lights for the following: Bias lighting, footwell lighting, panel system lighting, and monitor arc lighting.
1. PC human interface can be accessed through a Windows 10 device. Additional methods of PC interface shall be through a android tablet, or both Android and Apple devices.
1. Integrated control system shall be downloadable for Windows 10 or through appropriate app store.
1. Windows 10 will work on existing customer computer.
1. Optional LED RGBW light controls with presets for up to four lights to include: panel system acrylic, footwell lighting, down bias lighting, and arc acrylic lighting. Each of these is an option and not required.
1. Integrated control system shall incorporate a motion detector which will shut down all selected functions when workstation is unoccupied for fifteen (15) minutes. All previously selected functions will resume when motion detector senses movement in the workstation.
1. Desktop Airflow shall allow the user to select the airflow rate delivered to the desktop area through one desktop fan/filter unit. Fan speed shall be user selectable from no airflow to a maximum of no less than 50cfm. per fan.
1. Two LED task lights shall be at each station. Light shall be dimmable and have 2 arm adjustment for user comfort.
1. Status Indicator “Help” Light shall be controlled via integrated control system. Option to provide external switch as alternative control method.
1. Two, 250-watt forced air heaters located under the monitor surface shall be provided. Heaters must be able to rotate to blow heat on hands or feet. Heater to be controlled by a switch on the integrated software system. Option to provide external switch as alternative control method.
1. Adjustable Monitor Rack
1. Parabolic monitor rack shall be curved to match the cockpit shape of the workstation and achieve as close to equal focal lengths from the user’s eyes to the face of each monitor as possible and also position monitors for a view angle perpendicular to the screen.
1. Parabolic monitor rack shall be available in various sizes to accommodate up to 50” monitors and an individual weight of 70 pounds on a single or dual level.
1. Monitor rack platform shall be mounted on a movable platform which will permit a 10” focal length adjustment – from 19.7” (50cm) to 29.7” (75cm) with a maximum of 5 lbs. push/pull effort required to move all monitors simultaneously.
1. Mounting rail shall permit unrestricted horizontal adjustment of LCD mounts anywhere on the rail.
1. Quick release monitor mounts shall allow removal of monitors from the front of the station. An appropriate length service loop of the cables lets the tech remove monitors and set them on the keyboard surface for a quick swap when needed.
1. LCD mounts shall provide for all VESA monitor mount sizes.
1. LCD mounts shall adjust 8” vertically to allow centering dissimilar monitor sizes on horizontal rail.
1. Monitor rack shall be designed to permit “stacking” of a single monitor up to a complete additional row without removal or disassembly of any existing monitors or any portion of the base unit.
1. Monitor rack shall have simple and clean cable management on the back.
1. Vendor shall supply premium quality cable extensions as required to reach CPUs in CPU cabinets and provide adequate length to extend CPUs for service.
1. Keyboard Surface Data Connection Center (KSDCC)
1. A KSDCC shall be located along the back edge of the keyboard surface to provide an easy, instantly accessible location to plug in keyboards, mice and touch screen monitors if so equipped.
1. The KSDCC shall have at least eight (8) locations which can be configured with either USB or PS2 ports, two (2) locations for RJ11, RJ45 or DB9 ports.
1. The KSDCC shall incorporate an integral wire management channel along the back side of the keyboard surface to provide a place for mouse and keyboard cables to be stored and easily retrieved by the user to re-position keyboards or mice for either right or left hand use and to prevent cables from hanging down into knee/foot space.
1. Vendor shall supply appropriate number of extension cables of sufficient length to reach CPUs with each KSDCC specified.
1. Environmental
1. GREENGUARD Gold Certification for indoor air quality/emissions must be for the entire console. http://www.greenguard.org. Please provide copy of certification in your response. Include GreenGuard certification in Appendix D – Specification Compliance Table.
1. GREENGUARD Certification for only individual component parts of the console is not acceptable.
2) Complete generic identification – Sit-Stand Worksurface in Corner Dual Format
3) Applicable model/make/catalog number – All part numbers listed in IGCE
4) Manufacturer name – Xybix Systems, Inc.
Per VAAR 811.104 and 811.105, I certify the following regarding the above information:
· The salient characteristics listed above are essential to the Government’s requirements.
· To my knowledge, the restrictions listed above will not limit the acquisition to an item peculiar to one manufacture.
Name, Date: Gabriela Orteza, 6/10/2024
File details come from the government source that posted it. Updated .