Salient Characteristics.doc

DOC document 77 KB Posted

Attached to
TOTAL 100% SMALL BUSINESS SET-ASIDE FOR Benchtop Parallel Bioreactor System
Solicitation number
W911QX-25-Q-0044
Issued by

About this file

This is a Salient Characteristics document outlining the technical requirements for a benchtop parallel bioreactor system being procured by the Department of the Army Materiel Command at Aberdeen Proving Ground. The system must support working volumes up to 250mL, operate at least four vessels in parallel, be compatible with reusable/single-use vessels, and include features such as liquid-free temperature control, pH and dissolved oxygen control, foam/level detection, and liquid-free exhaust condensation. The system must operate at temperatures up to 60°C, fit on a lab bench with 21-inch depth, and support batch, fed batch, and continuous processing for bacteria and yeast.

The system must include license-free monitoring software with capabilities for parameter control, feedback loops, feeding strategies, guided calibration, user management, Excel reporting, and configurable charting. Delivery requirements include FOB destination to Army Research Laboratory-South at UT Austin, onsite installation and training, one-year warranty, readily available commercial replacement parts, and 12 months of software and hardware technical support. This procurement is set aside 100% for small businesses under solicitation number W911QX-25-Q-0044.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics

Commercial Availability Certification:

I have evaluated the commercial marketplace and determined that all of the salient characteristics documented below are available in products/services offered in the commercial marketplace.

Signature (typed name): ___ __________________________

Technical Point of Contact (TPOC) or Contracting Officer’s Representative (COR)

The Contractor shall provide a benchtop parallel bioreactor system with the following characteristics:

· Working volumes up to 250 milliliters (mL)

· Parallel operation of at least four (4) vessels in parallel

· Compatibility with reusable or single-use vessels

· Liquid free temperature control

· Standard Potential of Hydrogen (pH) and Dissolved Oxygen (DO) control

· Foam/level detection

· Liquid free exhaust condensation

· Capable of batch, fed batch, and continuous processing for bacteria and yeast

· Temperature control up to 60 degree Celsius (°C)

· Shall fit on a lab bench with a 21 inch depth restriction The Contractor shall provide license-free monitor and control software with the following characteristics:

· Control of process parameters

· Control of feedback loops

· Control of feeding strategies

· Step by step guided calibration

· User management and audit trails

· Reporting for Microsoft Excel

· Configurable charting

· Pre-defined cultivaton templates

· Graphical representation of progress And lastly, the major end item listed above shall have the following minimum specifications:

· Shall be delivered FOB Destination to Army Research Laboratory-South, University of Texas at Austin, Texas

· Shall provide onsite installation and training

· Shall provide at least a one (1) year warranty on all parts

· Vessels, sensors, and replacement parts shall be readily available on commercial vendor websites

· Shall provide maintenance and service options for one (1) year coverage increments

· Shall provide on-going software/reporting technical support for 12 months

· Shall provide on-going hardware technical support for 12 months Contracting Officer: