Salient Characteristics.doc

DOC document 88 KB Posted

Attached to
One (1) Digital Microscope - Amendment 001
Solicitation number
W911QX24Q0088
Issued by

About this file

This document is a Salient Characteristics Commercial Availability Certification for a digital microscope required under the federal contract opportunity W911QX24Q0088. The key details are:

The Contractor must provide one (1) digital microscope meeting the following minimum requirements: C-mount or equivalent camera, HD camera resolution, contrast modes including brightfield and darkfield, 5x to 100x zoom, motorized X-Y and Z stage control, fluorite/semi-apochromat optics, image stitching and color correction software, and 24-inch HD display. The microscope must be compatible with 110/120V, 60Hz US electrical standards. The Contractor must certify that all these salient characteristics are commercially available. The related federal contract opportunity is an upcoming solicitation from the Department of the Army Materiel Command for the required digital microscope, with a submission deadline to be considered for award.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics

Commercial Availability Certification:

I have evaluated the commercial marketplace and determined that all of the salient characteristics documented below are available in products/services offered in the commercial marketplace.

Signature (typed name): Wendy L. Sarney Technical Point of Contact (TPOC) or Contracting Officer’s Representative (COR) The Contractor shall provide one (1) Digital Microscope and shipping and handling in accordance with the minimum requirements of the Government stated below:

a. Shall have a C-mount or equivalent for digital camera on eyepiece tube.

b. Shall have a digital camera with the following specifications:

i. Compatible with C-mount or equivalent

ii. Possessing High Definition (HD) resolution (1920 x 1080)

c. Shall have the following contrast modes:

i. Brightfield

ii. Darkfield

iii. Upgradable option to differential interference contrast

d. Shall have a nosepiece/turret capable of rotating between five (5) different objects or equivalent zoom settings to include five (5) times zoom (x), ten (10)x, 20x, 50x, 100x.

e. Shall have an illuminator capable of providing adequate lighting for brightfield and darkfield contrast modes.

f. Shall have a 100 millimeter (mm) x 100 mm stage, or larger, capable of holding samples from one (1) centimeter (cm) x one (1) cm Si coupons to 100 mm diameter Si wafers.

g. Shall have equipped motorized X-Y state controllable with computer digital interface.

h. Shall have equipped motorized Z stage/nosepiece focus controllable with computer digital interface.

i. Shall have equipped fluorite/semi-apochromat or better five (5)x, ten (10)x, 20x, 50, 100x (or better)

j. Shall have a computer or equivalent embedded system for microscope stage, focus, and camera control.

k. Shall have a software package capable of the following:

i. Image stitching

ii. Image color corrections

iii. Micrograph file saving

l. Shall have a display/monitor with an HD resolution and 24 inch (in) diagonal.

m. Shall be compatible with United States (US) electrical standards listed below:

i. 110/120 volts (v)

ii. 60 Hertz (Hz)

iii. Type B outlet with ground Contracting Officer:

Template last updated 01/8/2019