SALIENT CHARACTERISTICS - CYSTOSCOPY.docx
DOCX document 21 KB Posted
- Attached to
- 6515--Q&A AMENDMENT l BLUE LIGHT CYSTOSCOPY Federal contract opportunity
- Solicitation number
- 36C26226Q1282
About this file
This is a Salient Characteristics document for a blue light cystoscopy system procurement by the VA Long Beach Healthcare System. The document specifies a complete endoscopic imaging solution from Karl Storz that combines Photodynamic Diagnosis (PDD) and standard white-light endoscopy capabilities. The system includes multiple components: a 4mm x 30cm autoclavable Hopkins II telescope, fiber optic light cables with locking mechanisms (minimum 230cm length), four pendulum camera heads with full HD capability, a dual-LED light source with 30,000-hour service life, a 4K/32-inch monitor, medical-grade printer, and a surgical cart. All reusable components must be autoclavable or suitable for medical reprocessing.
The system must deliver specific technical performance characteristics including adjustable blue light fluorescence in the 390-430nm range for PDD imaging, white light with CHROMA color enhancement across the 400-700nm range, 4K and HD video outputs, and integrated camera control unit connectivity. Additional requirements include a pendulum camera head with programmable user buttons, high-intensity LED light source compatible with PDD, fluorescence imaging technology (OPAL1), and foot switch controls for handsfree operation of lighting functions. The solicitation is issued by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 22 under solicitation number 36C26226Q1282 as a combined synopsis/solicitation for commercial products prepared in accordance with FAR Part 12.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| RFQ AMENDMENT - 36C26226Q1282 0001 - CYSTOSCOPY.docx | DOCX document | |
| 36C26226Q1282 0001.docx | DOCX document | |
| 36C26226Q1282_1.docx | DOCX document | |
| RFQ - 36C26226Q1282 - CYSTOSCOPY.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
VA Long Beach Healthcare System Blue Light Cystoscopy
1. BRAND NAME OR EQUAL
| BRAND |
| PRODUCT DESCRIPTION |
| PRODUCT # |
| QTY |
| Karl Storz |
| HOPKINS II 30° PDD TELESCOPE WITH |
ENLARGED VIEW, 4MM X 30CM,
AUTOCLAVABLE
27005BIA
| Karl Storz |
| LIGHT CABLE, 3.5MM X 230CM (7.6') |
AUTOCLAVABLE, WITH LOCKING
MECHANISM
495NAC-UG
| Karl Storz |
| IMAGE1 S™ HX-P SAPHIRA ONE-CHIP FULL |
HD PENDULUM CAMERA HEAD, FOR USE W/
TC305US
TH113US
| Karl Storz |
| HIGH-DEFINITION CAMERA STERILIZATION |
CONTAINER
KSZ-39301HCTS-ZBZ
| Karl Storz |
| IMAGE1 S™ SAPHIRA |
CONNECT II, KIT
TC205US-KT
| Karl Storz |
| IMAGE1 S™ SAPHIRA LINK, LINK MODULE, |
FOR USE W/ ONE-CHIP CAMERA HEAD
TH113US
TC305US
| Karl Storz |
| POWER LED SAPHIRA™, OPAL1® PDD, |
COLD LIGHT FOUNTAIN POWER LED
SAPHIRA™
TL410US
| Karl Storz |
| 32" 4K UHD LED MONITOR 16:9 ASPECT |
RATIO W/HDMI,12G-SDI, DP 1.2
TM343
| Karl Storz |
| SONY UP-DR80MD MEDICAL GRADE |
PRINTER FOR USE W/USB
WU1333
| Karl Storz |
| SLIM DESIGN SURGICAL CART W/OPEN |
SIDES KIT WITH CO2 TANK HOLDER
XUSM0008
UG130U-KT
| Karl Storz |
| LETTER SIZE COLOR PRINT PACK F/ |
UPDR80MD 2 RIBBONS, 2 ROLL PAPER
YIELD 100 PRINTS
WU1278
2. SALIENT CHARACTERISTICS
Obtain a blue light cystoscopy system that provides the following characteristics:
1. The system provides a complete endoscopic imaging solution for Photodynamic Diagnosis (PDD) and standard white-light endoscopy, including telescopes, camera heads, light source, image processing, documentation, viewing, and printing.
2. The system components must be designed to be autoclavable or be suitable for reprocessing in a medical environment.
3. The system must have a blue light with adjustable fluorescence intensity for photodynamic diagnosis (PDD) in the with a minimum and maximum 390–430 nm range.
4. The system must have a dual‑LED light source with an expected service life of 30,000 hours.
5. The system must have a white light with CHROMA color enhancement for improved visualization in within the 400–700 nm range.
6. The system must have user interface and network connectivity interface.
7. The system must have a pendulum camera head design with programmable buttons for customization of user controls.
8. The system must have 4K video outputs
9. The system must have HD video outputs
10. The system must have camera control unit (CCU) connect and link
11. The system must have a light source to assist with user operation.
12. The system must have a fiber optic light cable with a minimum length of 230 cm.
13. The system must have a foot switch/pedal to allow for hands‑free control of the blue and white light function.
14. The system must have a telescope with the following characteristics 4 mm diameter, autoclavable, available in 0°, 12°, 30°, and 70°.
15. The system must have a high‑intensity LED light source compatible with Photodynamic Diagnosis (PDD).
16. The system must have a fluorescence and optical imaging technology, i.e., OPAL1, that works in conjunction with the PDD to improve the visibility samples.
17. Must have Light Cables autoclavable, with locking mechanism.
File details come from the government source that posted it. Updated .