This document outlines requirements for a cryogen-free magnetotransport probe station solicitation. The United States Army intends to solicit Lake Shore Cryotronics, Inc. as the only responsible source to provide one cryogen-free magnetotransport probe station to support research at the Army Research Laboratory and Sensors and Electron Devices Directorate Power Components Branch. The probe station must have a 2.5 Tesla or greater magnetic field at base temperature, continuously variable temperature measurement probes from 10 to 400 Kelvin, a load-lock assembly for rapid sample exchange of approximately one hour, a minimum of six micromanipulated probe arms, a sample space greater than two square centimeters, and vibration at the sample of less than one micron. While sole-source procured from Lake Shore Cryotronics, other sources may submit proposals for consideration if received by the closing date of the notice. The Department of the Army Materiel Command Army Contracting Command at Aberdeen Proving Ground is the contracting agency.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics
Commercial Availability Certification:
I have evaluated the commercial marketplace and determined that all of the salient characteristics documented below are available in products/services offered in the commercial marketplace.
Signature (typed name): Owen Vail Technical Point of Contact (TPOC) or Contracting Officer’s Representative (COR)
Technical Requirements for Cryogen-Free MagnetoTransport Probe Station (CRX-VF) The Army Research Laboratory (ARL) and Sensors and Electron Devices Directorate (SEDD) Power Components Branch in service of the 2019 Office of the Secretary of Defense (OSD) Applied Research for Advancement of Priorities (ARAP) program on Topologically Enabled Devices requires one (1) Cryogenic MagnetoTransport Probe Station (CRX-VF):
· Shall be a dry (cryogen-free) system (completely closed-cycle operation)
· 2.5 Tesla (T) or greater magnetic field at base temperature
· Continuously variable temperature (CVT) measurement probes
· Load-lock assembly for rapid sample exchange (approximately one (1) hour)
· Minimum of six (6) micromanipulated probe arms
· Sample space greater than (>) two (2) centimeter (cm) squared (2cmx2cm)
· Temperature ranges from ten (10) to four hundred (400) Kelvin (K)
· Vibration at sample less than 1 (<1) micron (µm)
· Electrical noise of less than ten nanovolts per root Hertz (<10 nV/√Hz)
· Capable of back side voltage biasing
· Complete Hall measurement package, including instrumentation and software
· Properly grounded for mesoscopic measurement Contracting Officer: