Salient Characteristic Request Form updated 11.5.2025 (THA).pdf

PDF 183 KB Posted

Attached to
Total Hydrocarbon Analyzers Federal contract opportunity
Solicitation number
N3239826Q5174
Issued by
Department of the Navy Naval Supply Systems Command

About this file

This is a Request for Salient Characteristics dated 2/17/26 for the procurement of four (4) total hydrocarbon analyzers (THAs). The requirement is identified as "Brand Name or Equal" and is designated as a sole source acquisition. The analyzers must be delivered by Fiscal Year 2026 Quarter 3 (FY26 Q3).

Each analyzer must meet the following minimum specifications: capability to monitor total hydrocarbons with a flame ionization detector; four concentration ranges spanning 0 to at least 10,000 ppm at the high end; RS-232 digital output; ability to utilize 100% hydrogen as fuel; accuracy and repeatability of at least 1% or less of full scale; internal sample pump; footprint less than 494 square inches (width x depth); height less than 6 inches; weight less than 60 pounds; sample flow rate less than 2 L/min; warm-up time less than 2 hours after ignition; 115-120VAC @ 60 Hz single phase power supply; single or multi-point sampling capability; real-time analysis with touch screen display; chemically inert sample lines constructed of PTFE or stainless steel; no vendor training required for setup and operation. No assembly or setup is required upon delivery. Warranty shall be provided as stated by the manufacturer on quote. The security classification is Unclassified. Technical Point of Contact is Nathan Gargas, NAMRU-D EHEL, 937-904-9459, and the End User Point of Contact is Brian Sharits, NAMRU-D EHEL, 385-6355-7288.

View the file

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Request for Salient Characteristics

TYPE OF REQUIRMENT

DATE

1. Requirement Type:

Salient Characteristics Included: Brand Name or Equal:

Sole Source:

The Name of Directives and Department require state needed requirement. The following are required to meet the needs (state the quantity needed beside each component of the required product/service):

2. ASSEMBLY / SETUP REQUIRED?

3. WARRANTY / MAINTENANCE INFORMATION

4. DELIVERABLES

5. SECURITY

Unclassified

6. DELIVERY SCHEDULE / DATE / PERIOD OF PERFORMANCE

Delivery Lead Time Identified:

7. TECHNICAL POINT OF CONTACT (TPOC)

8. END USER POINT OF CONTACT

YES/NO:
Text28: Four (4) total hydrocarbon analyzers (THAs). Each analyzer must have a minimum of the following capabilities:

o One analyzer with capability to monitor total hydrocarbons o A flame ionization detector o Four (4) concentration ranges with 0 to at least 10,000 ppm at the high end o RS-232 Digital Output o Ability to utilize 100% hydrogen as a fuel o Accuracy of at least 1% or less of full scale o Repeatability of at least 1% or less of full scale o Must have an internal sample pump o Footprint less than 494 square inches (Width x Depth) o Height less than 6 inches o Weight must be less than 60 pounds o Sample flow rate less than 2 L/min o Warm-up time: less than 2 hours after ignition o Power: 115-120VAC @ 60 Hz single phase o Single or Multi Point Sampling o Realtime analysis with touch screen for viewing o Chemically inert sample lines, PTFE or stainless Steel o No vender training required for setup and operation o Delivery by FY26 Q3

Text29:
Text30: Nathan Gargas, NAMRU-D EHEL, 937-904-9459
Text31: Brian Sharits, NAMRU-D EHEL, 385-6355-7288
Text32:
Text33: Four (4) Total Hydrocarbon Analyzers
Text34: Manufacturer's warranty as stated on quote.
Text35: No
Brand Name or Equal:
Date36_af_date: 2/17/26
Text37: Total hydrocarbon analyzers

File details come from the government source that posted it. Updated .