Electromagnetic Induction Ground Conductivity Meters

SPECIFICATIONS

• Provides measurement of both the quad-phase (conductivity) and in-phase (magnetic susceptibility) components within two distinct depth ranges, all simultaneously, without any requirement for soil-to-instrument contact.

• Maximum effective depth of exploration of 1.5 m, applications in agriculture, archaeology and general soil sciences.

• Includes two receiver coils, separated by 1 m and 0.5 m from the transmitter, providing data from effective depth ranges of 1.5 m and 0.75 m respectively when positioned in the vertical dipole orientation, and 0.75 m and 0.375 m respectively when in the horizontal dipole orientation.

• Coil technology, supported by temperature compensation circuitry, to improve temperature-related drift.

• Collapsible calibration stand to support automation of the instrument calibration procedure. Once positioned within the stand, the instrument is to be calibrated within seconds, without any requirement for iterative adjustments.

• Support both walking and trailer-mounted survey methods.

• External power sources to be connected to the instrument for extended field operations.

• Lightweight, rechargeable battery pack with 25 hours of continuous operation.

• Data collection is to be supported by the DAS70-AR Data Acquisition System, or other appropriate field computer, connected by either cable or Bluetooth wireless technology.

• MEASURED QUANTITIES

1: Apparent conductivity in millisiemens per metre (mS/m)

2: In-phase ratio of the secondary to primary magnetic field in parts per thousand (ppt)

• INTERCOIL SPACING

1 and 0.5 metres

• OPERATING FREQUENCY

14.5 KHz

• MEASURING RANGE

Conductivity: 1000 mS/m In-phase: ± 29 ppt for 1 m In-phase: ± 7 ppt for 0.5 m

• MEASUREMENT RESOLUTION

± 0.1 % of full scale

• MEASUREMENT ACCURACY

± 5 % at 30 mS/m

• NOISE LEVELS

Conductivity: 0.5 mS/m;

In-phase: 0.02 ppt

• DATA OUTPUT

RS-232 serial port, Bluetooth

• POWER SOURCE

9 V disposable battery, external rechargeable battery pack

• BATTERY LIFE

Up to 5 h continuous for MN1604; 12 h continuous for L522; 25 h for rechargeable battery pack

• OPERATING TEMPERATURE

-40°C to + 50°C

• DIMENSIONS and WEIGHT

107 x 17 x 8 cm

5.4 kg