S02 - Pre-Sol Notice - 36C26024Q0931.pdf
PDF 149 KB Posted
- Solicitation number
- 36C26024Q0931
About this file
This document is a Presolicitation Notice for a federal contract opportunity related to Carestream RVG 6200 Digital X-Ray Sensors or equal items.
The key details are:
- Solicitation Number: 36C26024Q0931
- Set-Aside: SDVOSBC (Service-Disabled Veteran-Owned Small Business Concern)
- Product/Service Code: 6520
- NAICS Code: 339114
- Response Deadline: 09-20-2024 12:00 PM Pacific Time
- Archival Period: 60 days after response deadline
- The VA is seeking Carestream RVG 6200 Size 1 and Size 2 Intraoral Digital X-Ray sensors, or equal items meeting specific technical requirements detailed in the document, including minimum sensor resolutions, dimensions, grayscale depth, connectivity, and infection control features.
- The contract will be for 1 Size 1 sensor kit, 12 Size 2 sensor kits, 2 user manuals, and 1 service manual, all to be delivered electronically.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Presolicitation Notice
SUBJECT* Carestream RVG 6200 Digital X-Ray Sensors OR EQUAL ITEM
GENERAL INFORMATION
CONTRACTING OFFICE’S ZIP CODE* 98661
SOLICITATION NUMBER* 36C26024Q0931
RESPONSE DATE/TIME/ZONE 09-20-2024 12:00 PM PACIFIC TIME, LOS ANGELES, USA
ARCHIVE 60 DAYS AFTER THE RESPONSE DATE
SET-ASIDE SDVOSBC
PRODUCT SERVICE CODE* 6520
NAICS CODE* 339114
PLACE OF PERFORMANCE
POSTAL CODE
COUNTRY
CONTACT INFORMATION
CONTRACTING OFFICE ADDRESS NCO 20 Network Contracting Office
ATTN: V4CONT
1601 E. Fourth Plain Blvd Bldg.17, Suite B428 Vancouver WA 98661
POINT OF CONTACT* scott.reed2@va.gov
ADDITIONAL INFORMATION
AGENCY’S URL
URL DESCRIPTION
AGENCY CONTACT’S EMAIL ADDRESS Scott Reed EMAIL DESCRIPTION scott.reed2@va.gov
DESCRIPTION
ITEM INFORMATION
ITEM
NUMBER
DESCRIPTION OF
SUPPLIES/SERVICES QUANTITY UNIT UNIT PRICE AMOUNT
0001 1.00 EA _______________ _________________
Digital Intraoral X-ray sensor
CARESTREAM DENTAL RVG 6200 SIZE 1 COMPLETE KIT
(Brand name or equal)
0002 12.00 EA _______________ _________________
Digital Intraoral X-ray sensor
CARESTREAM DENTAL RVG 6200 SIZE 2 COMPLETE KIT
(Brand name or equal)
0003 2.00 EA _______________ _________________
USER MANUAL ELECTRONIC DELIVERY
0004 1.00 EA _______________ _________________
SERIVCE MANUAL ELECTRONIC DELIVERY
This requirement is for Carestream RVG 6200 Size 1 and Size 2 Intraoral Digital X-Ray sensors or equal item. For an equal item to be acceptable, it must meet the following minimum specifications:
Digital image sensor with CMOS, Scintillator and Optical Fiber
Size 1 and Size 2 available
Size 1 sensor must have a resolution of 1168 X 1562 pixels
Size 2 sensor must have a resolution of 1402 X 1874 pixels
Size 1 sensor must be no bigger than 27.6mm W X 37.7mm H X 7.3mm D
Size 2 sensor must be no bigger than 32.2mm W X 44.2mm H X 7.3mm D
Gray scale must be 4096 or 12 bits
USB 2.0 high speed connectivity with Type A plug
5V DC input voltage
Hygienic sleeve for infection control
Must be able to be disinfected with quaternary ammonium compound w/ alcohol
File details come from the government source that posted it. Updated .