S02 - Pre-Sol Notice - 36C26024Q0931.pdf

PDF 149 KB Posted

Attached to
6520--Carestream RVG 6200 Digital X-Ray Sensors OR EQUAL ITEM
Solicitation number
36C26024Q0931
Issued by

About this file

This document is a Presolicitation Notice for a federal contract opportunity related to Carestream RVG 6200 Digital X-Ray Sensors or equal items.

The key details are:

  • Solicitation Number: 36C26024Q0931
  • Set-Aside: SDVOSBC (Service-Disabled Veteran-Owned Small Business Concern)
  • Product/Service Code: 6520
  • NAICS Code: 339114
  • Response Deadline: 09-20-2024 12:00 PM Pacific Time
  • Archival Period: 60 days after response deadline
  • The VA is seeking Carestream RVG 6200 Size 1 and Size 2 Intraoral Digital X-Ray sensors, or equal items meeting specific technical requirements detailed in the document, including minimum sensor resolutions, dimensions, grayscale depth, connectivity, and infection control features.
  • The contract will be for 1 Size 1 sensor kit, 12 Size 2 sensor kits, 2 user manuals, and 1 service manual, all to be delivered electronically.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Presolicitation Notice

SUBJECT* Carestream RVG 6200 Digital X-Ray Sensors OR EQUAL ITEM

GENERAL INFORMATION

CONTRACTING OFFICE’S ZIP CODE* 98661

SOLICITATION NUMBER* 36C26024Q0931

RESPONSE DATE/TIME/ZONE 09-20-2024 12:00 PM PACIFIC TIME, LOS ANGELES, USA

ARCHIVE 60 DAYS AFTER THE RESPONSE DATE

SET-ASIDE SDVOSBC

PRODUCT SERVICE CODE* 6520

NAICS CODE* 339114

PLACE OF PERFORMANCE

POSTAL CODE

COUNTRY

CONTACT INFORMATION

CONTRACTING OFFICE ADDRESS NCO 20 Network Contracting Office

ATTN: V4CONT

1601 E. Fourth Plain Blvd Bldg.17, Suite B428 Vancouver WA 98661

POINT OF CONTACT* scott.reed2@va.gov

ADDITIONAL INFORMATION

AGENCY’S URL

URL DESCRIPTION

AGENCY CONTACT’S EMAIL ADDRESS Scott Reed EMAIL DESCRIPTION scott.reed2@va.gov

DESCRIPTION

ITEM INFORMATION

ITEM

NUMBER

DESCRIPTION OF

SUPPLIES/SERVICES QUANTITY UNIT UNIT PRICE AMOUNT

0001 1.00 EA _______________ _________________

Digital Intraoral X-ray sensor

CARESTREAM DENTAL RVG 6200 SIZE 1 COMPLETE KIT

(Brand name or equal)

0002 12.00 EA _______________ _________________

Digital Intraoral X-ray sensor

CARESTREAM DENTAL RVG 6200 SIZE 2 COMPLETE KIT

(Brand name or equal)

0003 2.00 EA _______________ _________________

USER MANUAL ELECTRONIC DELIVERY

0004 1.00 EA _______________ _________________

SERIVCE MANUAL ELECTRONIC DELIVERY

This requirement is for Carestream RVG 6200 Size 1 and Size 2 Intraoral Digital X-Ray sensors or equal item. For an equal item to be acceptable, it must meet the following minimum specifications:

Digital image sensor with CMOS, Scintillator and Optical Fiber

Size 1 and Size 2 available

Size 1 sensor must have a resolution of 1168 X 1562 pixels

Size 2 sensor must have a resolution of 1402 X 1874 pixels

Size 1 sensor must be no bigger than 27.6mm W X 37.7mm H X 7.3mm D

Size 2 sensor must be no bigger than 32.2mm W X 44.2mm H X 7.3mm D

Gray scale must be 4096 or 12 bits

USB 2.0 high speed connectivity with Type A plug

5V DC input voltage

Hygienic sleeve for infection control

Must be able to be disinfected with quaternary ammonium compound w/ alcohol