S02. RFI SOURCE SOUGHT.docx
DOCX document 22 KB Posted
- Attached to
- PORTABLE REVERSE OSMOSIS
- Solicitation number
- 36C26326Q0305
About this file
This is a Request for Information (RFI) and Sources Sought notice issued by the Department of Veterans Affairs, Network 23 Contracting Office for portable reverse osmosis equipment. The RFI number is 36C26326Q0305, and responses must be received by 10:00 p.m. CST on Monday, February 9, 2026, submitted to becky.delossantos@va.gov with the RFI number in the subject line. The NAICS code is 339113 with no Non-Manufacturer Rule Class waiver available, and the PSC code is 6515. This is market research only and does not constitute a solicitation; no solicitation currently exists, and responses will not be considered adequate responses to any future solicitation.
The VA seeks suppliers for the Evoqua W3T591409 Kit, RO Eon Rover ALX SCB or equivalent portable reverse osmosis system with a fully integrated endotoxin removal filter capable of meeting ANSI/AAMI 11663 and AAMI-Grade water standards. Key specifications include multi-stage filtration (carbon, reverse osmosis, and ultra-filtration), maximum product flow rate of 0.4 gpm at 77°F, inlet pressure range of 5–120 psi, heat-based disinfection, color touch screen LCD control panel, and a rust-proof aluminum cart with dual swivel casters. The equipment must be FDA-approved as a 510(K) medical device and certified to UL 61010-1 and IEC standards. A secondary item, the Evoqua W3T578738 Kit Booster, MHX/WRO for low-pressure inlet situations, is also requested. Delivery is to Fargo VA Health Care System, Fargo, North Dakota. Respondents must be OEM authorized dealers, distributors, or resellers providing new equipment only with full OEM warranty and support, and must provide company information, business classification (including socio-economic status), SAM UEI number, and any existing schedule contracts.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Sources Sought: 36C26326Q0305
The DEPARTMENT OF VETERANS AFFAIRS, Network 23 Contracting Office is conducting research and is issuing this Request for Information to learn of the availability and capability of business sources qualified to provide the items listed below as Brand-Name or Equal. This is NOT a solicitation.
SUBJECT: PORTABE REVERSE OSMOSIS, Brand-Name or Equal.
NAICS CODE: 339113 (there is NOT a Non-Manufacturer Rule Class waiver for this NAICS code).
PSC: 6515
INTERESTED SOURCES SHALL RESPOND BY:
All responses must have the VA RFI Number in the Subject Line of the email!!!
All responses must be received no later than 10:00pm CST, Monday, February 9, 2026 All responses must be submitted to this email address: becky.delossantos@va.gov
POTENTIAL SOURCES SHALL PROVIDE THE FOLLOWING INFORMATION IN THEIR RESPONSE:
1. Company name, address, phone number, primary contact(s), e-mail address.
2. State if your firm distributes the Brand-Name items or Equal items.
3. State if your firm has the required items on a Schedule (GSA, FSS, NAC, etc). If so, include contract number and expiration date.
4. SAM UEI number for your firm
5. State if your firm is a large business, small business, service-disabled-veteran-owned small business, veteran-owned small business, HUB zone 8(a), women-owned small business or a small-disadvantaged business.
6. If your firm is a Socio-economic small business (WOSB, VOSB, SDVOSB…), state if your firm is a Non-Manufacturer under the Non- Manufacturing Rule. Information regarding the Non-Manufacturer Rule can be found at these links:
a. Nonmanufacturer rule | U.S. Small Business Administration (sba.gov)
-OR-
b. https://www.sba.gov/partners/contracting-officials/small-business-procurement/nonmanufacturer-rule#:~:text=What%20is%20the%20nonmanufacturer%20rule,if%20it%20is%20a%20nonmanufacturer
BRAND-NAME OR EQUAL MODEL AND TECHNICAL SPECIFICATIONS:
Salient Characteristics Brand Name or Equal:
Evoqua W3T591409 Kit, RO Eon Rover ALX SCB Fully integrated endotoxin removal filter Ultrafilter assists in preparing water or bicarbonate concentrate to meet the requirements for the production of ultrapure dialysate according to the ANSI/AAMI 11663 standard. The filter’s unique proprietary single-stage design removes bacteria, viruses, endotoxins and other contaminants greater than 0.005 micron (5 nanometers) Multi-Stage Filtration System, including carbon filtration, reverse osmosis (RO), and ultra-filtration in one unit Features easy-to-use controls and instant access to key parameters Prevents RO shutdown when the inlet water has low PSI Uses heat to disinfect rather than chemicals Continuous flow of hot water with no time constraints Provides single heat, storage heat Multiple Leak Detector options, internal or remote sensor with user alert or shutdown Color touch screen LCD control panel, clear display of operational functions Recirculation product line Filter and membrane access – removable back panel, machine disassembly not required Certified to UL 61010-1, IEC 61010-1, CAN/CSA C22.2 No: 61010-1, IEC 61326-1 FDA approved 510(K) Medical Device Must be designed to meet AAMI-Grade water standards according to ANSI/AAMI 13959 standards Leak Detection System
| Maximum Product Flow Rate at 77°F | Up to 0.4 gpm (1.5 lpm) | |||
| Inlet Temperature | 40-104°F (4-40°C) | |||
| Inlet Pressure (Dynamic) | Minimum: 5 psi (0.34 bars) | Maximum: 120 psi (8.3 bars) | ||
| Inlet Water Flow/Water Feed | Minimum: 1.5 gpm (5.7 lpm) | |||
| Drain Capacity | Minimum 2.0 gal/min | |||
| Total Hardness | < 10 grains (pH dependent) | |||
| Silt Density Index (SDI) | < 5 | |||
| Total Chlorine | < 0.1 ppm |
CART:
Rust-proof and corrosion resistant. Aluminum frame, including built-in rubber bumper guard Large 6” dual swivel casters with locking features Rear directional locking casters with integrated braking mechanism Centrally located monitoring area that features pressure and temperature gauges, and sample ports with drip tray Includes booster pump for low-pressure inlet situations Open design to allow for easy access to filter and other areas of maintenance Complies with ETL Tip and Threshold Testing Standards, and meets and adheres to the latest ANSI, AAMI, ISO Standards provides hose and cord management options All-in-one water treatment delivery system
Brand Name or Equal:
Evoqua W3T578738 Kit, Booster, MHX/WRO, Rover Family Includes booster pump for low-pressure inlet situations System Design Small footprint for ease of use in vicinity of patients without interrupting patient care due to limited space in patient area Includes installation and start up System orientation provided
Delivery: Fargo VA Health Care System 2101 N. Elm St Fargo, ND 58102
NOTICE TO ALL RESPONDENTS:
The Vendor shall be an Original Equipment Manufacturer (OEM) authorized dealer, authorized distributor, or authorized reseller for the proposed equipment/system such that the OEM warranty and service/support pass through to the VA customer in totality. All warranty and service associated with the equipment/system shall be in accordance with the OEM terms and conditions. The vendor shall provide new items ONLY; no remanufactured or "gray market" items. All items must be covered by the manufacturer's warranty.
DISCLAIMER AND ADDITIONAL NOTICES:
Responses may be used as research to better understand the marketplace or for creating a potential future acquisition. Response is strictly voluntary; no reimbursement will be made for any costs associated with providing information in response to this or future information requests.
AT THIS TIME NO SOLICITATION EXISTS. (DO NOT REQUEST A COPY OF THE SOLICITATION).
This notice does not obligate the Government to award a contract or otherwise pay for the information provided in response. The Government reserves the right to use information provided by respondents for any purpose deemed necessary and legally appropriate. Any organization responding to this notice should ensure that its response is complete and sufficiently detailed to allow the Government to determine the organization's qualifications to perform the work. Respondents are advised that the Government is under no obligation to acknowledge receipt of the information received or provide feedback to respondents with respect to any information submitted. After a review of the responses received, a solicitation may be published in Federal Business Opportunities. However, responses to this notice will not be considered adequate responses to a solicitation.
No Proprietary, classified, confidential, or sensitive information should be included in your response. The Government reserves the right to use any non-proprietary technical information in any resultant solicitation(s.)
File details come from the government source that posted it. Updated .