This document is a combined synopsis/solicitation for a commercial item contract to provide two (2) Fresenius AquaC 15 Amp reverse osmosis water purification systems or equal for the Clarksburg VA Medical Center. The solicitation number is 36C24524Q0882 and the response date is September 5, 2024. The government intends to award a firm-fixed price contract as a result of this solicitation. The statement of work outlines the required specifications for the reverse osmosis systems, including minimum permeate capacity, yield, water supply input, and other technical requirements. The contractor must provide an OEM authorized distributor letter if they are not the manufacturer. Delivery is expected within 60 days of contract award and the contractor must provide training on operating and disinfecting the equipment. The government will select the quote that represents the best benefit at a reasonable price.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

B.3 STATEMENT OF WORK

STATEMENT OF WORK

CLARKSBURG VAMC

STATEMENT OF WORK AND GENERAL SPECIFICATIONS FOR

FRESENIUS AQUAC 15 AMP RO OR EQUAL

Program Manager/Service Line Representative (Requestor): angela.bush@va.gov Biomed Representative:

Logistics Representative: AcquisitionUtilizationSpecialistTeamClarksburg@va.gov FMS Representative:

Contracting Officer’s Representative (COR):

1. Background

VA Clarksburg has a need for two (2) replacement Reverse Osmosis Water Purification Systems for hemodialysis. Existing units are in constant need of repair.

2. Specifications

Please see below for salient characteristics.

Fresenius AquaC 15 Amp Reverse Osmosis Water Purification System or

Equal – QTY of 2 EA

Permeate Capacity: Minimum of 85 L/h @ 15° C

Yield: Up to 70%

Water Supply Input: Minimum of 3.0 L/Minute

Water Temperature: +5 to +30°C

Drain Operation: Minimum of 1.1 L/Minute

Peak Flow (Rinse): Minimum of 2.9 L/Minute

User Interface Display: Unit must be able to display, at a minimum, the date, time, total run time and time since last disinfection.

mailto:angela.bush@va.gov mailto:AcquisitionUtilizationSpecialistTeamClarksburg@va.gov

Temperature Measurement – Operating Range: 0 to 150°C

Reverse Osmosis Membrane: pH Tolerance Range of 2 to 11

Power Supply: 15 Amp

Product Flow Output: Minimum of 1.4 L/Minute

Noise: Less than 50 dB during normal operation

Chemical Disinfect: Must have an Integrated chemical port (No additional mixing container needed)

Equipment is required to have an automated rinse and standby heat program

Equipment must be mobile and on lockable casters

3. Installation

VA Clarksburg requires that the equipment, upon receipt, be assembled by the vendor prior to being inspected by our Biomedical Engineering staff and put into use.

4. Training

VA Clarksburg requires the vendor to provide training on operating/disinfecting the equipment upon receipt or before the equipment is put into operation. Option to have an additional five (5) healthcare professionals trained on the operation/disinfecting of the equipment after the equipment has been put into operation is also required.

5. Delivery Location

Attn: (Angela Bush)

Louis A. Johnson Medical Center 1 Medical Center Drive Clarksburg, WV 26301

Prior to delivery, Contracting Officer Representative and/or Service Line Representative shall receive shipping information from Contractor and forward to Logistics using the email(s) listed above.

The Contractor must comply and provide all necessary information to VA.

Clarksburg. Otherwise, the delivery may be refused and require redelivery at no additional cost to the government.

6. Requested Delivery Date

Delivery of equipment is expected within 60 days ARO.

7. Safety Data Sheet (SDS)

If applicable, contractor shall provide three (3) copies of each Safety Data Sheet for every product, chemical, etc. SDS sheets shall be provided for any material on the same day those materials arrive on VA property.

8. Delivery Coordination

Normal business hours for acceptance of deliveries are 7:00am-3:30pm, Monday thru Friday excluding Federal Holidays. Deliveries must be coordinated in advance with COR or Service Line Representative once a ship date is established. A Logistics Representative shall be the only VA entity signing for freight received on this requirement. The VA-issued purchase order number must be included in the packing slip. All items included in the requirement shall be shipped and delivered to site at same time. If deliveries must be broken into several shipments due to size, this must be coordinated in advance with the COR or Service Line Representative.

Failure to coordinate shipment with COR or Service Line Representative may result in failure to deliver and cause re-scheduling of delivery/freight at no charge to the government.