This Statement of Work (SOW) details a procurement for a Clavis EMG/STIM device for the Baltimore VA Medical Center. The device is needed to help physicians treat medical conditions including strabismus, dystonia, torticollis, blepharospasm, and spasticity. Key technical requirements include: a single device with both EMG and stimulation functions, muscle/nerve localization capabilities, battery operation, touch screen interface, portability (less than 500 grams), 8+ hour battery life, sampling rate of at least 1000Hz, and noise level below 1v. The device must operate continuously for 8 hours with an EMG performance amplifier bandwidth of 600Hz-2.5kHz and speaker bandwidth of 200Hz-14kHz. The contract will be a firm-fixed price, small business set-aside, with delivery to Fran Jeter/Dr. Hoppe at 10 N. Greene St, Room 2D116, Baltimore, MD 21201.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
B.3 STATEMENT OF WORK
STATEMENT OF WORK (SOW)
CLAVIS EMG/STIM DEVICE BALTIMORE VAMC
SECTION 1: BACKGROUND
1.1
Why does your service require this equipment – Handheld Estim/EMG device is needed to allow Physician to improve and precisely treat strabismus, dystonia, torticollis, blepharospasm and spasticity.
SECTION II PURPOSE
2.1 Product, Service or Outcome Needed.
What will the vendor do? Delivery, installation, assembly, start up? The vendor will only deliver the items and will not be needed for installation. Providers are already familiar with Estim/EMG devices.
The device does not have any type of software that will require connection to the VA network
2.2 Scope of the Product, Service, or Outcome.
Salient Characteristics:
• Must have EMG and Stim function in one device.
• Must provide muscle and nerve localization for accurate guide and monitoring injection of medical products
• Device should have a high sampling rate, low noise level and advanced filtering capabilities. Sampling rate of at least 1000Hz and noise level below 1v
• The device must be battery operated for portability
• The interface should be touch screen, clear display and have the ability to customize settings and provide visual feedback.
Should have at least these minimal technical data points:
• Power Supply: internally powered by battery
• The device must be portable, lightweight and compact. Less than 500 grams and with a battery life of at least 8 hours.
• Mode of operation: Continuous operation for at least 8 hours, EMG performance, amplifier band width: EMG mode 600Hz-2.5kHz
• Volume output: Speaker band width 200Hz-14kHz
DELIVERY TO:
Fran Jeter/ Dr. Hoppe 10 N. Greene St, Rm 2D116 Baltimore, MD 21201 Contact phone number: Fran Jeter 410-340-4732
2.3 Contract Type.
• Firm-Fixed Priced Contract