TECHNICAL REQUIREMENTS OR SALIENT FEATURES:

“BRAND NAME OR EQUAL TO UNETIXS”

A. Salient Features for vascular lab system

• Must have software running under Windows 10

• Must have DICOM Modality Worklist, compatible with VistA EMR (Electronic

Medical Record)

• Must be able to edit case/report and patient information before and after sent to

PACS

• Capability to store at least 50,000 patient cases/reports

• Capability to edit history questions and testing sequences

• No printer

• Monitor screen must be at least 24” wide touch screen with swivel/rotation capabilities

• Remote control with capabilities including but not limited to:

o Move up and down o Volume up, down, off o Save, hold/freeze o Inflate and deflate o Select cuff o Enter and exit o Gain increase/decrease o Base/waveform up and down

• Capability to eliminate false positive studies/readings

• Capability to sensor appropriate cuff wrapping for PVR (pulse volume recording) waveforms

• Calibrated PVR

• Cart with ergonomic features and adjustable height range 45”- 55”

• Maximum weight of machine 145lbs

• Dimensions: maximum width 22”, maximum depth 31”

• Minimum two storage drawers

• Keyboard: ergonomic and easy to clean/disinfect

• Battery power lasting 6-8 hours. Must be able to run on battery for back-to-back portable exams.

• Pressure cuffs for:

o calf/ankle sites-10cm o toes- 2.5cm o above knee and thigh sites-12cm o extra length for large thighs-12L

• Photoplethysmography (PPG) probes

• Doppler probes: 8MHz and 4-5MHz

• Digit clips

• Onsite installation

• Annual onsite preventative maintenance during and out of warranty period

• Onsite training

• Software upgrades

• Warranty: define warranty for system, software, pressure cuffs, hoses, all items with system, etc.

• List any added value