This is a Performance Work Statement (PWS) for a full-service maintenance and repair contract for specialized respiratory diagnostic equipment at the USPHS Lawton Indian Hospital in Lawton, Oklahoma.

The Indian Health Service (IHS), operating under the Department of Health and Human Services, requires comprehensive maintenance, inspection, calibration, and repair services for one MGC Diagnostics Platinum RTD Elite Series with Body Plethysmograph system (Serial No. 242000547). The contractor must provide all necessary labor, certified technicians, original equipment manufacturer (OEM) replacement parts, calibration standards, testing equipment, travel, and shipping to maintain the system in optimal, certified operating condition. All services must strictly adhere to Medical Graphics Corporation (MGC) manufacturer specifications, applicable Federal regulations, and Joint Commission accreditation standards. The contract requires one annual preventative maintenance and recertification visit scheduled within ±30 calendar days of the anniversary date, including comprehensive physical inspection, electronic calibration of all sensors and analyzers, system leak checks, baseline electronic alignment, and affixing a certification sticker. For unscheduled repairs, the contractor must respond on-site within 24 to 48 hours for critical/emergency failures and within 3 to 5 business days for non-critical issues. The contractor must supply 100% OEM parts with a minimum one-year warranty, provide toll-free technical phone and remote support (6:00 AM to 8:00 PM CST, Monday through Friday, with one-hour response time), and submit itemized Field Service Reports within ten calendar days of service completion. All technicians must be factory-trained and certified by MGC on plethysmography systems. The period of performance consists of a twelve-month Base Year (December 2, 2026 to December 1, 2027) with three optional twelve-month extensions through December 1, 2030.

View the file

Other files

File Type Posted
RFQ-246-27-Q-0002.pdf PDF
Pricing Schedule.pdf PDF

On GovTribe

Work with this file

  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Department of Health and Human Services Indian Health Service

Performance Work Statement (PWS) Lawton Service Unit – Respiratory Therapy MGC Diagnostic Platinum RTD

Elite Series Service Contract 11/03/2026

1. PURPOSE / BACKGROUND

The Indian Health Service (IHS), an operating division within the U.S. Department of Health and Human Services (HHS), is responsible for providing federal health services to American Indians and Alaska Natives. The mission of IHS is to raise the physical, mental, social, and spiritual health of American Indians and Alaska Natives to the highest level and to assure that comprehensive, culturally acceptable personal and public health services are available and accessible to members of the 574 federally recognized Tribes across the United States.

The Lawton Service Unit (LSU) is a Joint Commission–accredited federal healthcare facility in Oklahoma. The mission of the Lawton Service Unit is to deliver quality care in order to attain positive health outcomes for all valued patients with emphasis on cultural sensitivity and human dignity. To support this mission, the Respiratory Therapy Department utilizes specialized pulmonary diagnostic equipment to assess and treat cardiopulmonary conditions.

The purpose of this Performance Work Statement (PWS) is to establish a full-service maintenance, inspection, calibration, and repair agreement for the Respiratory Therapy Department’s MGC Diagnostics Platinum Elite DL Series w/ Body Plethysmograph system located at the USPHS Lawton Indian Hospital. This PWS defines the required outcomes; except where specific technical or manufacturer standards are mandated, the methods used to achieve those outcomes are the responsibility of the Contractor.

2. SCOPE / REQUIRED OUTCOMES

The Contractor shall provide all necessary labor, certified technicians, original equipment manufacturer (OEM) replacement parts, calibration standards, testing equipment, travel, and shipping necessary to maintain the MGC Diagnostics Platinum RTD Elite Series with Body Plethysmograph system in optimal, certified operating condition.

All preventative maintenance, calibrations, diagnostic tests, and system recertifications shall strictly adhere to Medical Graphics Corporation (MGC) manufacturer specifications, applicable Federal regulations, and The Joint Commission accreditation standards.

Location of Service:

USPHS Lawton Indian Hospital Respiratory Therapy Department 1515 Lawrie Tatum Road Lawton, Oklahoma 73507

3. PERFORMANCE REQUIREMENTS AND STANDARDS

The Contractor’s performance will be evaluated against the following standards. (See Section 10 for the Quality Assurance / Surveillance Matrix.)

# Required Service (Outcome)

Performance Standard

Acceptable Quality Level

(AQL)

Method of Surveillance

3.1

Annual Scheduled Preventative Maintenance

(PM) &

Recertification

Complete annual factory-level PM, comprehensive multi-point inspection, calibration, and full system recertification per MGC specifications.

Performed within ±30 calendar days of the scheduled annual anniversary date;

100% compliant with OEM criteria.

Detailed Field Service Report & Biomedical verification.

3.2

Unscheduled Emergency & Corrective Repairs

On-site response and diagnostic correction for unexpected system malfunctions or hardware failures using OEM components.

On-site technician arrival within 24 to 48 hours from initial LSU service call; system restored to operational readiness.

LSU Bio-Med service ticket logs & Contractor work orders.

# Required Service (Outcome)

Performance Standard

Acceptable Quality Level

(AQL)

Method of Surveillance

3.3 Replacement Parts & Hardware Warranty

All replacement components, circuit boards, and sensors must be new OEM parts manufactured or authorized by Medical Graphics Corporation

(MGC).

100% OEM parts;

minimum 1-year (12-month) warranty on all installed parts and associated labor.

Work order itemization, parts receipts, and invoice review.

3.4

Technical Phone & Remote Diagnostics Support

Direct access to certified technical specialists for real-time troubleshooting, system error resolution, and operational guidance.

Priority toll-free phone and remote assistance available 6:00 AM

– 8:00 PM CST,

Monday through Friday (excluding federal holidays).

Call response 1 hour.

LSU staff trouble logs and vendor teleconference records.

3.5

Service Documentation & Joint Commission Compliance

Comprehensive, itemized Field Service Reports detailing inspection items, as-found/as-left test results, calibrations, replaced parts, and system clearance.

100% of reports submitted to LSU Biomedical Engineering and Respiratory Therapy within 10 calendar days of service completion.

Document review against Joint Commission standards.

4. CONTRACTOR RESPONSIBILITIES

4.1. Preventative Maintenance and Recertification

The Contractor shall perform one (1) comprehensive annual Preventative Maintenance inspection and system recertification visit per contract year, scheduled within 30 days prior to or 30 days after the annual service date.

PM procedures must conform to Medical Graphics Corporation (MGC) guidelines and include, at a minimum:

Comprehensive physical and mechanical inspection of the body plethysmograph cabin, door seals, latching mechanisms, and structural components.

Electronic calibration of pressure transducers, pneumotachographs, flow sensors, gas analyzers, and automated valve circuits.

System leak checks, baseline electronic alignment, and verification of software diagnostic routines.

Affixing a visible, legible inspection sticker indicating date inspected, next due date, technician signature, and pass/fail certification.

# Required Service (Outcome)

Performance Standard

Acceptable Quality Level

(AQL)

Method of Surveillance

3.6 Technician Qualifications & Competency

All field service engineers and technicians performing on-site service or calibration must be factory-trained and certified on MGC plethysmography systems.

100% compliance;

valid manufacturer certifications on file prior to performing service.

Verification of training certificates and credentials.

4.2. Unscheduled and Corrective Repairs

Critical / Emergency Failure (Tier 1): Defined as any catastrophic malfunction, primary sensor/analyzer failure, electronic breakdown, or plethysmograph cabin failure that renders clinical diagnostic testing impossible or compromises patient safety. Upon notification by the LSU Biomedical Engineering Department or the Contracting Officer’s Representative (COR), the Contractor must respond on-site within 24 to 48 hours.

Non-Critical / Minor Failure (Tier 2): Defined as an operational defect, peripheral malfunction, auxiliary software error, or cosmetic/secondary hardware issue where the system remains safe and clinically viable for basic pulmonary measurements.

Upon notification by LSU Biomedical Engineering or the COR, the Contractor shall coordinate and respond on-site within 3 to 5 business days, avoiding unnecessary emergency dispatch costs.

The full-service agreement covers all associated costs: certified technician travel, on-site labor, diagnostic bench time, freight, and shipping of parts.

Prior Authorization: Any service requirement or repair determined to fall outside the standard scope of repair requires written authorization/modification from the Contracting Officer (CO) (coordinated through the COR/Biomedical Technician) prior to initiating billable work.

4.3. Replacement Parts and Warranty

The Contractor shall maintain direct access to authentic OEM MGC replacement parts and sub-assemblies. No rebuilt, salvaged, or third-party unauthorized components shall be installed without express written consent from the LSU Biomedical Department.

Warranty: Any and all parts installed under this contract shall carry a minimum of a one

(1) year full-replacement warranty, effective from the date of system acceptance by the government.

4.4. Technical Telephone and Remote Troubleshooting Support

The Contractor shall provide toll-free telephone support and secure remote diagnostic capabilities to assist LSU clinicians and biomedical staff with immediate troubleshooting.

Operational hours: 6:00 AM to 8:00 PM CST, Monday through Friday, excluding federal holidays.

5. LEVEL OF EFFORT AND EQUIPMENT INVENTORY

The equipment covered under this agreement is operated by the Respiratory Therapy Department at the Lawton Indian Hospital:

6. PERIOD OF PERFORMANCE

The period of performance shall consist of a twelve (12) month Base Year with three (3) consecutive twelve (12) month Option Years, exercised at the discretion of the Government:

Base Year: December 02, 2026 to December 01, 2027 Option Year 1: December 02, 2027 to December 01, 2028 Option Year 2: December 02, 2028 to December 01, 2029 Option Year 3: December 02, 2029 to December 01, 2030

7. SPECIAL REQUIREMENTS

7.1. Facility Coordination & Biomedical Notification

The Contractor’s service representative must contact and notify the Lawton Indian Hospital Biomedical Engineering Department prior to traveling to the facility or performing any on-site service.

Upon arrival at the hospital, the service engineer must report to the Security Desk, obtain a visitor/vendor badge, and coordinate immediate check-in with the Biomedical Department and Respiratory Therapy Department.

Item

Equipment Description Model / Series

Serial Number Location

Coverage Level

MGC

Diagnostics Platinum RTD Elite Series w/ Body Plethysmograph

USPHS

Full Comprehensive Service (Annual

PM,

Recertification, Unscheduled Repairs, Parts, Calibration, Phone/Remote Support)

Pulmonary Lawton Function Indian

001 Testing System 242000547 Hospital with Body (Respiratory Plethysmograph Therapy

Dept.)

7.2. Smoke- and Tobacco-Free Environment

The USPHS Lawton Service Unit campus (including all buildings, parking areas, sidewalks, and federal grounds) is strictly a tobacco-free and smoke-free environment.

Use of any tobacco, vaping, or electronic nicotine products is strictly prohibited.

7.3. Security, Badging, and Safety

Contractor personnel must present a valid government-issued photo ID at check-in.

Contractor personnel must abide by all LSU Infection Control, Safety, and Environment of Care policies while on premises. Escort by LSU personnel will be provided when required in clinical or secure areas.

7.4. Confidentiality, Privacy, and Information Security

The Contractor and its personnel shall respect patient privacy and strictly comply with the Health Insurance Portability and Accountability Act (HIPAA) and federal privacy standards.

Contractor personnel shall not disclose, copy, or disseminate any sensitive hospital, operational, or patient data encountered during maintenance or calibration.

All inquiries, comments, or media requests shall be immediately referred to the Government Designated Representative.

7.5. Technician Qualifications

All technicians assigned to perform maintenance, calibration, or repairs must be fully qualified, factory-trained, and certified by Medical Graphics Corporation (MGC) to service the Platinum RTD Elite plethysmograph series. Documentation of current technician qualifications must be provided upon request.

8. DELIVERABLES AND REPORTING REQUIREMENTS

8.1. Itemized Service Reports

Following any maintenance, repair, or calibration visit, the Contractor must furnish a comprehensive written Field Service Report.

Delivery: The service report must be delivered upon completion of on-site service or submitted electronically within ten (10) calendar days to:

Biomedical Equipment Technician

Respiratory Therapy Supervisor Report Content: The report must explicitly document:

1. Date and duration of service.

2. Equipment model and serial number (242000547).

3. Detailed list of all inspections, adjustments, and calibrations performed.

4. As-found and as-left diagnostic measurements/calibration values.

5. Itemized list of all OEM replacement parts installed (including part numbers and serial numbers).

6. Final operational status (Certified / Pass / Fail) meeting Joint Commission accreditation standards.

7. Any follow-up recommendations for clinical staff.

8.2. Invoicing and Warranty Documentation

Itemized invoices must be submitted in accordance with contract terms. Copies of warranty documentation and proof of OEM replacement parts used during repairs must accompany the invoice.

8.3. Contractor Point of Contact (POC)

The Contractor shall designate a single primary POC and provide a dedicated toll-free customer service phone number (answered during core business hours, 6:00 AM to 8:00 PM CST, Monday through Friday) for dispatch, scheduling, and contract administration.

9. CONTRACT ADMINISTRATION & GOVERNMENT FURNISHED

RESOURCES

9.1. Points of Contact

Biomedical Engineering Point of Contact:

Biomedical Technician USPHS Lawton Indian Hospital 1515 Lawrie Tatum Rd, Lawton, OK 73507

Department Technical Representative / Supervisor:

Respiratory Therapy Supervisor USPHS Lawton Indian Hospital 1515 Lawrie Tatum Rd, Lawton, OK 73507

Contracting Officer’s Representative (COR) / LSU Facilities & Engineering:

Facilities Engineer USPHS Lawton Indian Hospital 1515 Lawrie Tatum Rd, Lawton, OK 73507

9.2. Government-Furnished Property, Facilities, and Services

The Government shall supply:

safe, climate-controlled working environment with adequate commercial electric power for calibration tools and test instrumentation.

Reasonable physical access to the equipment during pre-arranged service windows.

The use of government facility space and equipment is strictly limited to the performance of services defined under this contract.

9.3. Contractor-Furnished Property

The Contractor shall furnish all specialized diagnostic tools, calibration gases/standards, flow meters, electronic analyzers, and safety equipment necessary to complete service on the plethysmograph system. All contractor-provided calibration gases must be medical-grade and traceable to the National Institute of Standards and Technology (NIST).

9.4. Changes to the Performance Work Statement

Any changes, scope modifications, or equitable adjustments to this PWS must be authorized and approved exclusively through written formal correspondence from the Contracting Officer (CO). Costs incurred through the actions or directions of any party other than the CO shall be borne entirely by the Contractor.

10. QUALITY ASSURANCE / SURVEILLANCE & PERFORMANCE MATRIX

The following checklist shall be utilized by the Government at the conclusion of each maintenance cycle, repair visit, or billing period to evaluate and document Contractor performance. The completed matrix will be retained within the official COR file and submitted with receiving documentation (PODCR).

# Performance Objective Standard & AQL Met?

10.1 Were all scheduled preventative maintenance inspections and system calibrations completed within the allowable timeframe (±30 days of annual date) according to MGC guidelines?

[ ] Yes [ ] No

10.2 Did the Contractor respond on-site within 24 to 48 hours for all unscheduled/emergency repair requests?

[ ] Yes [ ] No

10.3 Were 100% of replacement parts authentic OEM MGC parts with an active 1-year warranty provided?

[ ] Yes [ ] No

# Performance Objective Standard & AQL Met?

10.4 Was priority telephone and remote technical support accessible during the required operating window (6:00 AM – 8:00 PM CST, Monday–Friday)?

[ ] Yes [ ] No

10.5 Did the technician report to Biomedical Engineering prior to initiating service and comply with security check-in, smoking prohibitions, and safety policies?

[ ] Yes [ ] No

10.6 Were comprehensive, itemized service reports delivered to Biomedical Engineering and Respiratory Therapy within ten (10) calendar days of service?

[ ] Yes [ ] No

10.7 Did all service personnel hold current manufacturer-authorized qualifications and competencies?

[ ] Yes [ ] No

10.8 Were all customer inquiries, complaints, or equipment concerns addressed and resolved satisfactorily?

[ ] Yes [ ] No

Overall Performance Rating: [ ] Acceptable [ ] Unacceptable

Respiratory Therapy Supervisor Signature Date

Clinical / Department Director Signature Date

Rona S. Rupe -S Digitally signed by Rona S. Rupe -S Date: 2026.11.16 13:44:09 -06'00'

Rona S. Rupe -S Digitally signed by Rona S. Rupe -S Date: 2026.11.16 13:44:27 -06'00'

File details come from the government source that posted it. Updated .