This request for quotes solicitation seeks a cell and virion analysis equipment system capable of reliably detecting and isolating viral particles. The National Institute of Allergy and Infectious Diseases requires the brand name Beckman Coulter/Life Sciences CytoFLEX S V4-B2-Y4-R3 Flow Cytometer and CytoFLEX SRT Cell Sorter, as well as a Haake Refrigerated Water Bath and CytoFLEX service installation. The solicitation is issued using simplified acquisition procedures under NAICS code 334516 with a size standard of 1,000 employees and PSC code 6640. Quotes are due by the date specified and award will be made to the responsible offeror providing the requested equipment and meeting stated requirements for a firm-fixed-price purchase order.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| RFQ Attachment 2 - Bill of Materials.xlsx | XLSX spreadsheet | |
| RFQ - CytoFlex.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Needs – CytoFLEX S Flow Cytometer & CytoFLEX SRT Cell Sorter
This statement of work is for the procurement of two instruments capable of: a) detecting and measuring traits of cell and virus populations, and b) separating cells and virus subpopulations from a sample based on their properties. These are our specific requirements:
Flow Cytometer
· Must have 4 lasers, including a violet (405 nm) laser for small (<100 nm) particle detection and 3 additional lasers (488, 561, and 638 nm) for fluorescence applications.
· Must have Violet Side Scatter Resolution (VSSC) and the capability to detect 80 nm polystyrene particles.
· Fluorescence resolution: the instrument can achieve <3% rCV.
· Sensitivity parameters:
-FITC: <30 molecules of equivalent soluble fluorochrome (MESF‐FITC) from the 488 nm laser -PE: <10 molecules of equivalent soluble fluorochrome (MESF‐PE) from the 488 nm laser (this property is essential for detecting small particles, such as viruses)
· Fixed integrated optics and quartz flow cell design with >1.3 numerical aperture.
· Proprietary fluorescence and side scatter light delivery by fiber optics design to ensure high performance, high efficiency, and low-noise signal detection.
· Must have adjustable flow rate options that range from 10 to 240 µl/min to provide high sensitivity (in both small particle and low concentration regimes)
· Must have plate loader options that can accommodate standard and deep well 96 well plates.
· Must have compact size, up to 17x17x13.5 in. Must include 10 L Sheath and Waste tank options.
Cell Sorter
· Must have 4 lasers, including a violet (405 nm) laser for small (≤200 nm) particle detection and 3 additional lasers (488, 561, and 638 nm) for fluorescence applications.
· Must have Violet Side Scatter Resolution (VSSC) and the capability to detect particles ≤200 nm.
· Fluorescence resolution: the instrument can achieve <3% rCV.
· Must be capable of processing samples at a range of flow rates between approximately 10 μL and 100 μL per minute providing high sensitivity (in both small particle and low concentration regimes).
· Must have increased accuracy of population identification and high sensitivity to measure events with low staining index for rare event detection and application with dim fluorescent staining. Sensitivity parameters:
-FITC: <30 molecules of equivalent soluble fluorochrome (MESF‐FITC) from the 488 nm laser -PE: <10 molecules of equivalent soluble fluorochrome (MESF‐PE) from the 561 nm laser -APC: <25 molecules of equivalent soluble fluorochrome (MESF-APC) from the 638 nm laser
· Fixed integrated optics and quartz flow cell design.
· Accepts the following collection devices: 6-, 24-, 48-, 96- and 384-well plates, 96-deep well plate.
· Must be able to perform system startup, QC and sort calibration to be ready to sort within 30 minutes. All sort settings can be defined automatically.
· Must be able to generate 30,000 Hz to 35,000 Hz droplets and optimize the settings automatically for sorting.
· Must have technology used for automatic droplet optimization, monitor and maintain, and drop delay determination.
· Must be capable of simultaneously bulk sorting up to 4 defined populations of particles >99% purity with 5% target population, sample threshold ≤10,000 events per second, while the yield rate is above 80% theoretic rate.
· Must have up to four-way sorting with mixed modes to 5 mL and 15 mL tubes; one-way sorting can be conducted with up to 96 deep well plates and standard 384 micro-well plates or plates.
· Must have compact size, up to 29x19x18 in.