06/10/2021

RFP A21-1328 Two Person Glove Box Addendum 1

Questions Submitted and Answers Provided

1. Chamber dimension, as least Depth: 30” (762mm), our glovebox is 29.5”(750mm), is it acceptable?

Answer: If this is an inside dimension I don't see a problem, but if this is an outside dimension then we might have an issue depending on actual interior clear space.

2. Interior LED light, our VTI is exterior LED light, which is located outside the glovebox, on the top of the glovebox, more convenient to replace the light.

Majority glovebox vendor mounts the light outside the glovebox. Please let me know if it’s acceptable? If not, would you please let us know why they want a interior LED light? Where is the location? We are able to custom design.

Answer: The main requirement for the interior lighting is to avoid glare on the front panel and provide an even light inside the working area. Exterior lights should be acceptable as long as they meet both conditions.

3. Do you have a specific requirement for the delivery time?

Answer: The glove box must be delivered to Ames Laboratory by 09/15/2021.

Installation would be expected by 09/30/2021

End of Addendum 1