RFI Attachment 1- Critical Operational Device Specifications (CODS) _ Analyzer Mercury Vapor_.pdf

PDF 138 KB Posted

Attached to
Request for Information (RFI) - ANALYZER MERCURY VAPOR 2024
Solicitation number
SMD_ANALYZER_MERCURY_VAPOR_2024
Issued by

About this file

This document is a Critical Operational Device Specification (CODS) for a mercury vapor analyzer system being sought by the Department of the Army. The specifications detail both minimum required and desired characteristics for a portable device capable of detecting and measuring toxic mercury vapor in the air.

Key operational requirements include detection sensitivity of 0.001 mg/m³ (minimum) to 0.1 μg/m³ (desired), measurement range of 0.1-500 μg/m³, gold film sensor technology (minimum) or atomic fluorescence/UV absorption technology (desired), 5-10 hours of rechargeable battery life, and field portability for single-operator use. The system must include a hardened transport case meeting MIL-STD 810H testing standards for environmental conditions, have Trade Agreements Act compliance, and include comprehensive documentation and training. Additional desired parameters specify maximum weights of 37 lbs packaged and 15 lbs operational, plus Defense Health Agency Authority to Operate certification if applicable. The contractor must provide at least one year of warranty (three years desired), repair parts support, and both physical and digital technical manuals.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Attachment 1

Critical Operational Device Specification (CODS)

1. Critical Specifications: The below are critical specifications and the minimum must be met for the device to be considered acceptable by the government. They are separated by “minimum” and “desired” columns.

2. Instructions: Indicate “yes” or “no” under both minimum and desired columns below. Additional comments may be added.

3. Critical Operational Device Specifications

Operational Characteristic Minimum Required Desired Characteristics

Meets Minimum Y/N?

Meets Desired Y/N? Comments

1 Detection Capability

The system shall be able to detect and measure toxic mercury vapor in the air with a minimal sensitivity of 0.001 mg/m³.

The system shall be able to detect and measure toxic mercury vapor in the air with a minimal sensitivity of of 0.1μg/m³.

N/A

2 Detection Range

The system shall be able to measure mercury vapor particles with a concentration as low as 0.1 μg/m³ and as high as 500 μg/m³.

Minimum Requirement N/A

3 Sensor Type

The system technology shall be based on gold film sensor technology.

The system technology shall be based on atomic fluorescence spectroscopy technology or UV absorption technology.

Calibration

The annual calibration shall be able to be performed at an Army Maintenance Depot Level. The manufacturer will provide the necessary instruction and equipment to allow the Army to calibrate the devices. The manufacturer shall not be required for periodic calibration or maintenance tasks.

The annual calibration shall be able to be performed at the Field Level. The manufacturer will provide the necessary instruction and equipment to allow the Army to calibrate the devices. The manufacturer shall not be required for periodic calibration or maintenance tasks.

5 Battery

The system shall be able to run off rechargeable battery power for no fewer than 5 hours of active use.

The system shall be able to run off rechargeable battery power for no less than 10 hours of active use.

6 Display The system shall have a display capable of operation in low-light conditions.

Minimum Requirement

7 Field Portability

The system shall be easily field portable and able to be used by one individual for surveying.

4. Critical Logistic Specifications

Operational Characteristic Minimum Required Desired Characteristics

Meets Minimum Y/N?

Meets Desired Y/N? Comments

1 Repair Parts Support

Contractor shall provide a full listing of repair parts and distinguish the items to be considered high-failure repair parts (identified by OEM and Government Subject Matter Experts).

The country of origin, availability, proprietary parts, and procurement accessibility of each item shall be provided on the list.

2 Warranty

The contractor shall provide a warranty for at least (1) year. This warranty shall include all parts, In- House labor, and technical support during the warranty period

Desired the contractor provides a warranty for at least 3 years.

This warranty shall include all parts, In-house labor, and technical support during the warranty period.

3 Manuals The vendor shall provide 1 physical and 1 digital User and Operator Service Manuals with each system

Trade Agreements Act (TAA) Compliance

The system shall be compliant with TAA regulations.

Harden Reusable Transport Case

A.) The system and all accessories should be contained in harden reusable transport cases (e.g.

Hardigg, Pelican, ECS Case or Equivalent).

B.) The system should be designed to withstand MIL-STD 810H (1) testing in accordance with (IAW) the following paragraphs:

500.6 Low Pressure (Altitude)

501.7 High Temperature

502.7 Low Temperature

507.6 Humidity

520.5 Temperature, Humidity, Vibration, and Altitude

6 Copyright/ Distribution Release

The vendors will provide an operational and Maintenance copyright/ distribution release for supporting Army Technical manuals to be distributed throughout the force.

7 Software update

The vendor will identify application software for their system. To include software updates. Minimum Requirement

8 Training

The vendor shall provide Operating an Analyzer Mercury Vapor training.

Proper handling of the device, sample preparation, and data interpretation.

5. ADDITIONAL PERFORMANCE PARAMETERS

These are desired Additional Performance Parameters (APPs) but are not essential.

# Additional Performance Parameters (APPs) Minimum Requirement

1 Weight, Packaged

The system and all accessories shall come in one hardened transport case weighing no more than 37lbs.

2 Weight, Operational

The analyzer device shall weigh less than 15lbs.

3 Authority to Operate (ATO)

If applicable, the system shall currently have an Authority to Operate from the Defense Health Agency (DHA) that will be supported for at least two (2) years from the date of award.

***End of CODS***