RFI Attachment 1_ CODS _Sink Unit Scrub.pdf
PDF 136 KB Posted
- Attached to
- Request for Information - Sink Unit Scrub
- Solicitation number
- SinkUnitScrub_2024
About this file
This is a Critical Operational Device Specification (CODS) document for a Sink Unit Scrub system, detailing required and desired specifications for the Department of the Army. The document outlines 11 critical operational specifications including FDA regulatory compliance requirements, water supply capabilities, temperature controls (100-145°F), self-contained design with tankless water heater, hands-free operation, weight restrictions (maximum 550 lbs with desired 300 lbs), and power supply requirements supporting both 110-120 VAC and 220-240 VAC.
The specifications also include logistical requirements such as repair parts support documentation, minimum 1-year warranty (desired 3 years), user/operator manual requirements, and Trade Agreements Act compliance. Additional desired performance parameters specify multiple hands-free dispensing mechanisms, collapsible design reassembled within 5 minutes without tools, manufacturer training requirements, and hardened transport case specifications meeting multiple MIL-STD testing requirements (including temperature, humidity, shock, and vibration testing). The vendor must provide service and installation training at 3 events for 5 people.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| RFI_ Sink Unit Scrub.pdf | ||
| RFI Attachment 2_Company and Product Information _ Sink Unit Scrub.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment 1
Critical Operational Device Specification (CODS)
Critical Specifications: The below are critical specifications and the minimum must be met in order for the device to be considered acceptable by the government. They are separated by “minimum” and “desired” columns.
Instructions: Indicate “yes” or “no” under both minimum and desired columns below.
Additional comments may be added.
1. Critical Operational Device Specifications
Number
Operational
Characteristic Minimum Required Desired
Characteristics
Meets Minimum
Y/N?
Meets Desired
Y/N? Comments
Food and Drug Administration (FDA) Regulatory Oversight
The Sink Unit Scrub shall be cleared by the FDA with a 510(k) approval if it is a Class II medical device. If the system is a Class I medical device, it shall be registered with the FDA for human use.
Minimum Requirement
N/A
2 Water Supply
The system shall supply both hot and cold water.
Self- Contained
The system shall be self-contained with a tankless electric water heater pump
4 Temperature
The system shall provide adjustable water temperature from 100° Fahrenheit (F) to 145°F.
5 Accessories
The system shall include the following unattached accessories:
• Mixing faucet with gooseneck for hot and cold water
• Hot and cold faucet handles
Desired additional accessories:
• External
Water supply
• Electrical Supply
• Waste hose assembly
• Water supply hose
• Filter assembly
• Electric Inlet Cord
• Surgical Soap Dispenser
Hands-free Operation
The system shall have hands-free water control capabilities.
7 Weight weigh no more than 550 lbs.
Desired weight of no more than 300lbs.
8 Pressure automatically repressurize the sink when the tank pressure is reduced to a preset lower limit, approximately 30 psi minimum.
9 Power
The power supply shall provide a balanced power source for the device, supporting both 110-120 VAC, 60Hz, and 220-240 VAC, 50Hz. It must include auto-detection and simultaneous switching capability between both voltage ranges.
Additionally, if battery charging is necessary, the device must support charging for both voltage ranges.
Water Weight Accommodati on
The system shall accommodate water weight of 100lbs.
Desired to accommodate water weight of 182 lbs.
Sink Bowl Volume Capacity
The sink bowl shall have a capacity of 4032 cubic inches.
Desired sink bowl capacity of 5880 cubic inches
2. Critical Logistic Specifications
Operational
Characteristic Minimum Required Desired
Characteristics
Meets Minimum
Y/N?
Meets Desired
Y/N? Comments
1 Repair Parts Support
Contractor shall provide a full listing of repair parts and distinguish the items to be considered high-failure repair parts (identified by OEM and Government Subject Matter Experts). The country of origin, availability, proprietary parts, and procurement accessibility of each item shall be provided on the list.
2 Warranty
The contractor shall provide a warranty for at least (1) year.
This warranty shall include all parts, In- House labor, and technical support during the warranty period
Desired the contractor provides a warranty for at least 3 years.
This warranty shall include all parts, In-house labor, and technical support during the warranty period.
3 Manuals
The vendor shall provide 1 physical and 1 digital User and Operator Service Manuals with each system
Minimum Requirement
N/A
4 Trade Agreements Act (TAA) Compliance
The system shall be compliant with TAA regulations.
3. ADDITIONAL PERFORMANCE PARAMETERS
These are desired Additional Performance Parameters (APPs) but are not essential.
Additional Performance
Parameters (APPs) Minimum Requirement
The system should offer multiple mechanisms to trigger soap and water dispensing
Multiple mechanisms should trigger hands-free hot and cold water and soap dispensing.
The system should be readily collapsible, and reassembly should require no tools
The sink should be able to collapse to less than one half of its cubic size for transport and be reassembled in less than five (5) minutes without external tools.
3 Manufacturer Training
The manufacturer should train both the program management team and selected Service Members (Schoolhouse trainers) on the operation and maintenance of the system.
4 Harden Reusable Transport Case
A.) The system and all accessories should be contained in harden reusable transport cases (e.g. Hardigg, Pelican, ECS Case or Equivalent).
B.) The system should be designed to withstand MILSTD testing in accordance with (IAW) the following paragraphs:
501.7 High Temperature
502.7 Low Temperature 507.6 Humidity
511.7 Explosive Atmosphere
514.8 Vibration
516.8 Shock
517.3 Pyroshock
528.1 Mechanical Vibrations of Shipboard Equipment
(Type I- Environmental and Type II- Internally Excited)
5 Training Events The vendor provide service and installation training at 3 training events for 5 people
(End of CODS)