Requirements and Specifications.docx

DOCX document 15 KB Posted

Attached to
Liquid Chromatography Mass Spectrometer Triple Quad System
Solicitation number
005-RFB-2276-2026
Issued by
Story County, Iowa

About this file

This is a Requirements and Specifications document for the State of Iowa detailing procurement specifications for a Liquid Chromatography Mass Spectrometer Triple Quad System. The document specifies that bidders must provide a complete system inclusive of columns, computer, installation, power supply, software package, and a five-year maintenance agreement. The instrumentation and software must be compatible with Iowa DCI Labs' current validated methods, requiring only partial validation rather than full validation or new standard operating procedures to bring the system online.

The system specifications include mandatory technical features: StepWave ion transfer optics, two high resolution and high stability quadrupole analyzers, a T-Wave collision cell for MS/MS performance, a low noise off-axis long life photomultiplier detector, software with a HUB design and connected apps for user experience consistency, a Z-spray ion source, rapid ionization mode switching capability for polarity switching, an atmospheric pressure ionization source, and MaxPeak High Performance Surfaces technology. No specific pricing terms, set-asides, funding amounts, response dates, or contract award dates are provided in this specifications document.

View the file

Other files

File Type Posted
1350382-event.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Requirements and Specifications Bidder must offer a complete Liquid Chromatography Mass Spectrometer Triple Quad System.

· The system must complete with columns, computer, installation, power supply, software package and a five (5) year maintenance agreement.

· New instrumentation and software must not require a full validation or new standard operating procedures.

· Instrumentation and software that is consistent with the Iowa DCI Labs current validated methods, requiring only a partial validation to get the instrument online. The system must:

· Have StepWave ion transfer optics

· Utilize two high resolution, high stability quadrupole analyzers

· Have a T-Wave collision cell enabled for optimal MS/MS performance

· Offer a low noise, off axis, long life photomultiplier detector

· Utilize software with a HUB design and apps that provide a consistent connected user experience across all applications.

· Have a Z-spray ion source

· Be capable of rapid ionization mode switching for polarity

· Have an atmospheric pressure ionization source

· Utilize MaxPeak High Performance Surfaces (HPS) technology

File details come from the government source that posted it. Updated .