required specifications.docx
DOCX document 221 KB Posted
- Attached to
- Ultrapure Water System
- Solicitation number
- N0018926QL274
About this file
This is a required specifications document for an ultrapure water system procurement by the Department of the Navy Naval Supply Systems Command under solicitation N0018926QL274. The system must be capable of producing Type I water that meets the stringent purity requirements necessary for use in Inductively-Coupled Plasma Optical Emission Spectrometry (ICP-OES) and Dissolved Organic Carbon (DOC) analysis applications. Type I water specifications typically require resistivity of 18.2 megohms-cm or greater, total organic carbon content below 5 ppb, and absence of particulates and ionic contaminants that would compromise analytical instrument performance.
The procurement seeks a complete ultrapure water system solution suitable for laboratory analytical operations. Prospective contractors should ensure their proposed systems meet all specified water quality parameters and are compatible with the demanding requirements of both ICP-OES instrumentation and DOC analysis methodologies. The system should include appropriate pretreatment, purification, and storage components to maintain water quality standards throughout the system.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Combined synopsis solicitation d.pdf | ||
| Solicitation Amendment-post.pdf | ||
| Combined Synopsis Solicitation Final.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
REQUIRED SPECIFICATIONS
1. To support its academic mission of educating undergraduate U.S. Navy Midshipmen and conducting sponsored research in the field of chemical oceanography, the Department of Ocean and Atmospheric Sciences at the United States Naval Academy (USNA) requires an ultrapure water system.
The ultrapure water system will be capable of producing Type I water suitable for use in Inductively-Coupled Plasma Optical Emission Spectrometry (ICP-OES) and Dissolved Organic Carbon (DODC) with the following minimum characteristics: Resistivity:
per billion); Bacteria: < 1 CFU/mL (colony-forming units); and
2. System Technical Specifications and Requirements: The Ultrapure Water System will be capable of producing Type 2 for storage in a holding tank between 20-30 L. The tank will have a vent filter with CO2 absorbent to prevent airborne contamination and will recirculate water automatically through a UV lamp to maintain purity. The UV lamp will be mercury-free.
The system will purify Type II water for point of dispensing (POD) of Type 1 water at a rate 2L/minute or greater. Daily output will be 100L/day or greater. Power will be 120 VAC, 60 Hz. System will come with a remote dispensing wand. A full set of filters and a set a set of spare filters will be supplied with the system. The system will come with all connectors and accessories required for installation and operation. All freight charges and shipping will be included in system cost. System delivery will come with a minimum of 1 day of vendor support for installation and training for 3 or more users.
image6.png image96.png image97.png image98.png image99.png image100.png image101.png image102.png image103.png image104.png image105.png image7.png image106.png image107.png image108.png image109.png image110.png image111.png image112.png image113.png image114.png image115.png image8.png image9.png image10.png image11.png image12.png image13.png image14.png image15.png image16.png image17.png image18.png image19.png image20.png image21.png image22.png image23.png image24.png image25.png image26.png image27.png image28.png image29.png image30.png image31.png image32.png image33.png image34.png image35.png image36.png image37.png image38.png image39.png image40.png image41.png image42.png image43.png image44.png image45.png image1.png image46.png image47.png image48.png image49.png image50.png image51.png image52.png image53.png image54.png image55.png image2.png image56.png image57.png image58.png image59.png image60.png image61.png image62.png image63.png image64.png image65.png image3.png image66.png image67.png image68.png image69.png image70.png image71.png image72.png image73.png image74.png image75.png image4.png image76.png image77.png image78.png image79.png image80.png image81.png image82.png image83.png image84.png image85.png image5.png image86.png image87.png image88.png image89.png image90.png image91.png image92.png image93.png image94.png image95.png