Purchase specification for Fourier Transform Infrared Spectrometer (FTIR) per MXDTA-PO-91-539
1. FTIR Physical Characteristics
a. Spectrometer Dimensions:
i. Maximum: 25” W x 27” D x 11” H (62.6cm x 69.8cm x 27.6cm)
b. Spectrometer Weight:
i. Maximum: 132 lbs (60kg)
c. Sample Compartment Dimensions
i. Maximum: 8.3” H x 10.2” D X 5.9” H (21cm x 26cm x 15cm)
2. FTIR Spectrometer Characteristics
i. High Stability, Long Lifetime Mid-IR Source
ii. Continuously Variable Iris Aperture
iii. Gold Optics for far IR, mid IR, and near IR operation
iv. DLaTGS Detector
v. Attenuation Wheel
vi. Validation Wheel
vii. A/D Converter of 24 bit
viii. English Language kit
ix. North American Power Cord
x. Ge on KBR beam splitter
xi. Is50 Built in Diamond ATR Module: monolithic diamond ATR crystal, all reflective optics permitting wide spectral range performance, DLaTGS detector
3. FTIR Spectrometer Performance Specifications
i. Spectral Range, Standard System: 7800 – 350 cm-1
ii. Spectral Range, CSI Optics: 6400 – 200 cm-1
iii. Spectral Range, Multi-Range Optics: 27,000-20 cm-1
iv. Optical Resolution, Mid-IR: Less than 0.09 cm-1
v. Signal-to-Noise, 1 minute scan: 55,000:1
vi. Signal-to-Noise, 5 second scan: 13,000:1
vii. Ordinate Linearity: 0.07% T
viii. Wavenumber Precision: Better than 0.01 cm-1
ix. Scan Velocity (15 values): 0.158-6.25 cm/sec
x. Rapid Scan, Spectra per Second: 65 (at 16 cm-1), 95 (at 32 cm-1)
xi. MCT Dewar LN2 Hold Time: 18 Hours
4. FTIR Operational Software
i. OMNIC Software
5. Attenuated Total Reflectance (ATR) Module Characteristics:
i. Diamond Crystal
ii. Spectral Range: 5000-80 cm-1
iii. Down Force of Pressure Device: 60 lbs
iv. Detector: DLaTGS
6. Computer
i. Dell Optiplex Workstation Windows 10 Professional 64 Bit Loaded (or similar model with similar capabilities)
ii. Dell UltraSharp 22” Flat Panel Widescreen Monitor (or equivalent)
7. OMNIC spectra Nicolet IS50 Edition Cap
i. OMNIC Spectra – FTIR Standard
ii. Georgia State Crime Lab Drug
iii. Polymer Additives and Plasticizers Library
iv. Hummel Polymer and Additives Library
v. Common Materials and White powders library
vi. Polymer and Plasticizer by ATR library
vii. Comprehensive Forensic Library
viii. Inorganics Library
ix. Nicolet Vapor Phase Library