Purchase Description (W80RA6-08-26C).pdf
PDF 101 KB Posted
- Attached to
- Optical Spectrum Analyzer
- Solicitation number
- PANRSA-26-P-0000-03396901
About this file
This is a Purchase Description for the procurement of one (1) Optical Spectrum Analyzer by the U.S. Army TMDE Activity (USATA) for the Army Primary Standards Laboratory (APSL) located at Building 5435 Fowler Road, Redstone Arsenal, AL. The procurement specifies a brand name or equivalent high-performance optical spectrum analyzer, designated as the Yokogawa AQ6370E-20 Performance Model, to replace aging equipment.
The required equipment must meet the following technical specifications: wavelength range of 600 to 1700 nm, wavelength accuracy of ±0.01 nm, wavelength resolution of 0.02 nm, close-in dynamic range of 78 dB typical, and level range of +20 to -90 dBm. Additional features include up to 2x faster measurement capability using SMSR mode, sharper spectrum measurement through HCDR mode, and a large touchscreen LCD display. The unit shall be delivered to U.S. Army TMDE Activity, Attention: Jonathan Larson, Building 5435, Redstone Arsenal, AL 35898. The procurement quantity is one unit.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 033969 MFR for Optical Spectrum Analyzer (W80RA6-08-26C)_Redacted.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Purchase Description Procurement of Optical Spectrum Analyzer
The U.S. Army TMDE Activity (USATA) has a requirement to purchase a brand name purchase of one (1) Optical Spectrum Analyzer High Performance model or equivalent that will be a suitable replacement for an aging equipment. This item shall be delivered to the Army Primary Standards Laboratory (APSL), Building 5435 Fowler Road, Redstone Arsenal, AL. The Optical Spectrum Analyzer shall include the following specifications or equivalent (if not brand name):
Requirements:
Yokogawa AQ6370E-20 Optical Spectrum Analyzer Qty:1 Performance Model Wavelength range:600 to 1700 nm Wavelength accuracy:±0.01 nm Wavelength resolution:0.02 nm Close-in Dynamic range:78 dB typ.
Level range:+20 to -90 dBm Up to 2x faster measurement: SMSR mode Sharper spectrum measurement: HCDR mode Large touchscreen LCD
1.0 Delivery Requirements
Shipping Address: U.S. Army TMDE Activity, Attn: Jonathan Larson, Building 5435, Redstone Arsenal, AL 35898