This document is a Sources Sought notice for a 100MHz benchtop Nuclear Magnetic Resonance (NMR) spectrometer for the U.S. Army Dugway Proving Ground West Desert Test Center in Utah. The government intends to award a sole source contract for this requirement but is seeking information from interested vendors, including small businesses, to support a competitive procurement.

The key specifications for the NMR instrument include a 100MHz permanent magnet, capability to analyze proton and phosphorous nuclei, flow cell capability, autosampler, compatibility with NMR analysis software, and ability to perform various NMR experiments. The government is seeking information on pricing, delivery, commercial availability, and recommendations to improve the acquisition approach. Responses are due by September 9, 2024.

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Product Description

U.S. ARMY DUGWAY PROVING GROUND

WEST DESERT TEST CENTER

DUGWAY, UTAH 84022-5000

1. Subject: Benchtop Nuclear Magnetic Resonance (NMR) Instrument

2. Final Delivery Location: West Desert Test Center, U.S. Army Dugway Proving Ground, Utah.

3. Key Personnel from Dugway Proving Ground: TBD

Scope Of Work Required:

· A benchtop NMR instrument is needed to upgrade the current NMR analysis capability.

· Procure one benchtop NMR instrument with the following specifications:

· 100MHz permanent magnet that does not require cryogens

· Perform analysis with a minimum of proton and phosphorous nuclei (1H/31P)

· Flow cell capability to automate sample analysis

· Autosampler that can hold a capacity of 25 standard 5mm NMR tubes

· Compatibility with Mnova and TopSpin NMR software

· Ability to perform the following experiments: 1D, T1, T2, COSY, JRES, TOCSY, HSQC, DEPT, APT, HECTOR, and HMBC

· Built-in touchscreen and capability of connecting to an external computer for operation

· Resolution of < 0.007 ppm and a sensitivity of > 220:1 based on 1 scan of 1% Ethylbenzene

· Recommended dimensions of 17 x 15.25 x 32 inches (must be able to fit inside the hood)

· Weight < 250 lbs