Ion Chromatograph (IC) System

The IC will be used for atmospheric research conducted by the Atmospheric Chemistry Group at NOAA/Pacific Marine Environmental Lab

1. Description: Ion Chromatograph system, including autosampler(s), degasser, columns, carbonate suppression system, conductivity detectors, and software for managing instrumentation and rapid analysis of chromatograms for anions and cations.

2. Requirement: All components necessary for dual detection of ion concentrations of Anions [acetate, formate, methane sulfonic acid, chloride, bromide, nitrate, sulfate, oxalate] and Cations [sodium, ammonium, potassium, magnesium, calcium] simultaneously

3. Requirement: Automated sampling system

a. injection volume of 50ul with the ability to change the volume from 10-250ul using a single, variable loop or multiple hardware configurations.

b. sample delivery via syringe

c. injection methods needed: multiple injection volume options for samples that fall outside of standard curve (i.e. full loop/partial loop injection)

d. Sample analysis: At least 90 samples can be loaded at once for concurrent/parallel anion/cation determination

e. Connection to PC via USB or RS-232

f. Compatible with 0.3/0.7 polypropylene vials or microtiter plates

4. Requirement: Concentration range of detection for all ions: 10ppb to 10ppm; must be able to handle sample matrix of water:methanol in 5:1 ratio as well as diluted seawater matrix.

5. Requirement: Columns: Vendor to provide 1 setup column plus one spare column each for cations and anions (4 columns total) plus 2 sets of appropriately matched guard columns each for cations and anions

a. Requirement: Analysis time: <=30 minutes for anions, <=15 minutes for cations

b. Required precision:<10% RSD

6. Requirement: Eluent must have flexibility to be prepared by user from commonly available lab chemicals (non-proprietary); 2 or 4 liter eluent vessels included for user generated eluent.

7. Requirement: Suppression system for anion determination without interference by carbonate (May be chemical + carbonate suppression)

a. Suppressor regeneration from non-proprietary chemicals/consumables

b. 10 year suppressor warranty

8. Requirement: Software:

a. Automated peak detection and quantification based on standard curve

b. Chromatogram overlay capability

c. Standard curve comparison analysis (to monitor for drift)

9. Eluent Degassing capability (either with inline or supplied degasser)

10. Ability to operate on a moving platform: can handle motion and temperature/pressure (15-30C) fluctuations commonly found at sea.

11. Requirement: System can be shut down, transported, and restarted on a semi-annual basis without major impact to performance or start up costs due to being unused for a prolonged period (>1 month)

12. Heated and thermostatic conductivity cell which permits conductivity measurements unaffected by temperature variations for improved reproducibility

13. 10 year guarantee of available replacement parts for entire system.

14. Installation by qualified service technician includes delivery of system, complete setup, including multiple standard curve runs with accuracy and precision of system verified by standards and multiple injections with adequate software familiarization of PMEL technician that instrument can be operated independently upon departure.

a. Installer should inform the PMEL technician of the required preventative maintenance for weekly, monthly and annual intervals for optimal performance.

15. IC should have automated regeneration of suppressor, and inert, non metallic components such as peek throughout system for solvent compatibility.

16. Vendor to include consumables for the analysis of at least 1000 samples (sample vials and standards).