This statement of work outlines requirements for a contractor to provide pipette calibration and preventative maintenance services to laboratories across the United States for the Department of Health and Human Services. The contractor must calibrate approximately 1100 single-channel and 500 multi-channel pipettes twice per year in accordance with ISO 17025 standards, including preventative maintenance, shipping, and a five day turnaround time for calibration or ten days for repairs. The contractor must also provide monthly status reports, invoices, and maintain ISO 17025 accreditation. The initial contract term is one year with four optional one-year extensions.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
1. SCOPE
The U.S. Department of Health and Human Services (DHHS) is seeking a qualified Contractor to provide offsite (mail in) pipette calibration with preventative maintenance to designated laboratories across the country. The contractor will provide timely pipette calibration, with 5 day or less turnaround time for each lab, regardless of lab location, that conforms to ISO 17025 calibration standards and meets manufacturer’s specifications unless otherwise specified. The contractor will provide quotations for pipette repairs and return repairs in 10 days or less to the participating laboratories. Contractor must be in good standing with and accredited under ISO 17025 by an accreditation body that is signatory to the International Laboratory Accreditation Cooperation (ILAC) Mutual Recognition Agreement (MRA) and operate in accordance with all applicable guidelines to maintain accreditation. Contractor must publish a 30, 60, and 90 day schedule of a reoccurring forecast of turn-ins for flexibility and budgeting purposes. Comment by FORSHEY, BRETT: Leslie – any concerns here?
2. BACKGROUND
The laboratories are the analytical components responsible for the detection and identification of infectious biological agents that may pose a threat to public health. Work performed by these laboratories requires the use of multiple, accurate, and reliable pipettes. Reasonable turn-around time is critical to avoid disruptions in daily operations and up-to-date status via tracking mechanisms is required. These laboratories are located in multiple sites across the continental United States.
Upon award, the contractor shall be required to sign a Non-Disclosure Agreement and instructions regarding the recipient for deliverables will be provided.
3. PERFORMANCE AND SURVEILLANCE PLAN REQUIREMENTS
3.1 Performance Requirements: The contractor must provide calibration and preventative maintenance services for repeater, single channel, and multi-channel pipettes from all major national brands including but not limited to Rainin, ThermoFisher, and Eppendorf as required.
1. Contractor must be ISO 17025:2005 accredited by an accreditation body that is signatory to the International Laboratory Accreditation Cooperation (ILAC) Mutual Recognition Agreement (MRA).
2. Pipette repair and calibration services for all laboratories must be coordinated through DHHS/DHS Logistical Management POC and laboratory requesting the service.
3. Information and status updates regarding calibration, repair, or replacement services for all laboratories must be communicated to DHHS and requesting lab as requested and when contractor has updates to services.
4. Contractor shall provide calibration services for each of approximately 1100 single-channel and 500 multi-channel pipettes (actual numbers may vary) in the system every six months (twice a year).
5. Calibration standards:
a. Calibration services must be performed twice a year for each pipette, approximately every 6 months
b. Calibrations shall provide pre- and post- calibration condition associated with each calibration according to manufacturer’s specifications unless otherwise approved by DHHS.
c. Preventative maintenance minimum: pipette inspection, replace worn seals, shafts and pistons, cleaning and greasing, pressurized leak test, minor repairs
d. Multichannel pipettes: each channel will be calibrated as individual channels.
e. Calibration label must be affixed to pipette with calibration date and next calibration due date.
f. Pipettes that fail calibration and are unrepairable must be clearly marked as such.
6. Repairs standards (repairs not included in preventative maintenance):
a. Repairs can be requested by laboratories any time during the contract
b. Repairs must be preapproved by DHHS
c. Quotes for repairs must be submitted to DHHS via electronically to recipients and designated contacts as instructed within 48 hrs of receiving pipettes.
d. Repaired pipettes must be calibrated as defined in Calibration Standards.
e. DHHS reserves the right to refuse repair.
7. Calibration certificate:
a. Calibration service and documentation must be ISO 17025 compliant.
b. For all calibration services, “as found” (pre-calibration) and “as left” (post-calibration) measurements must be included in the calibration certificate.
c. For all calibration services, pre- and post- conditions must be included in the calibration certificate.
d. Issue of service report and calibration certificate must be provided, electronically to recipient laboratories and listed contacts and as a hard copy with the return shipment, at each calibration service.
e. Contractor must maintain records of all certificates issued under this contract, preferably through electronic access with print capability.
8. Shipping
a. Shipment of the pipettes must be coordinated with the laboratories to avoid interruptions in daily operations.
b. Contractor must provide shipping to and from the laboratory facilities for pipettes submitted for calibration/repair services.
c. Contract shall use 3-day shipping, unless otherwise requested by DHHS.
d. Contractor must perform calibration and preventative maintenance on pipettes and shipped to requesting laboratory within 5 business days.
e. A pipette that requires repairs and calibration must be completed and shipped back to laboratory within 10 days of approval.
f. Contractor must provide service scheduling and tracking of service requests, preferably online.
g. Contractor shall provide notification by email or phone call if necessary (preferably as soon as an issue is identified) to DHHS, designated partners, and the affected Labs when schedules will not be met.
h. The contractor shall report any incidents or damage to pipettes sustained during shipment via email to the DHHS, requesting laboratory, and other points of contacts as identified by DHHS.
i. Contractor will coordinate with DHHS to develop a yearly schedule of pipette calibrations.
3.2 Surveillance Plan
1. A list of serial numbers for pipettes to be serviced under this contract will be generated by the laboratory requesting service and included in the shipment of pipettes to the contractor.
2. A complete inventory of all pipettes covered by this contract will be maintained by DHHS.
3. Only pipettes on this list are authorized for service under this agreement. If the contractor receives a request for service for a pipette not included on the most recent version of the list, the contractor shall provide notification as instructed and wait for approval from DHHS. If approval is not granted by DHHS, service is not authorized under this contract.
4. The contractor shall report any incidents or damage to pipettes sustained during shipment as instructed.
3.2 Performance Requirements and Surveillance Plan
Table 1. Performance Requirements and Surveillance Plan
| Performance Requirements |
| Performance Standards |
| Surveillance Method& Frequency |
| Contractor shall calibrate and/or repair pipettes and return to requesting laboratory |
| Within 5 business days (calibration only) or 10 business days (repair) of request for service |
Monthly inspection of service records
Monthly or as needed inspection of schedule and ability to adhere to schedule
| Contractor shall calibrate and/or repair pipettes |
| In a controlled environment and be ISO 17025:2005 accredited by an accreditation body that is signatory to the International Laboratory Accreditation Cooperation (ILAC) Mutual Recognition Agreement (MRA). |
| Proof of accreditation made available to DHHS and requesting laboratories |
4. DELIVERABLES
4.1. Documentation
1. Contractor will provide service and documentation as outlined in 3. PERFORMANCE AND SURVEILLANCE PLAN REQUIREMENTS.
2. A monthly status report to reflect work accomplished must be provided no later than the 5th of each month to DHHS and designated points of contact.
3. A monthly invoice report to reflect work accomplished and costs must be provided no later than the 5th of each month to DHHS Financial Management POC.
a. Shipping costs must be captured separately in invoice.
b. Repairs performed, with associated quote reference numbers, must be captured separately in invoice.
Table 2. In Appendix A outlines the Deliverables
5. CONTRACT AND PERIOD OF PERFORMANCE
5.1 Contract type
1. IDIQ
5.2 Period of Performance
1. The term of this contract shall be 1 year from date of award with 4 option years. A monthly status report to reflect work accomplished and cost information shall be provided no later than the 5th of each month to the designated financial points of contact using the format requested.
| Base Period: | March 2023 – February 2024 |
| Option Year 1: | March 2024 – February 2025 |
| Option Year 2: | March 2025 – February 2026 |
| Option Year 3: | March 2026 – February 2027 |
| Option Year 4: | March 2027 – February 2028 |
APPENDIX A
Invoicing: The contractor shall furnish an invoice monthly in accordance with instructions provided.
Table 2. Deliverables
| Site |
| Deliverable |
| Quantity |
| Format and |
Method of Delivery Due Date
DHHS Financial POC
| Monthly status report that contains invoices, shipping costs, and number of pipettes calibrated, requiring repair, or deemed irreparable and funding remaining on the contract |
| Once per month |
| Electronic delivery to recipients as instructed, DHHS financial point of contact and other designated partners. |
| NLT 5th of each month |
| DHHS and designated POCs |
| Monthly status report that contains number of pipettes calibrated, shipping, requiring repair, or deemed irreparable and funding remaining on the contract |
| Once per month |
| Electronic delivery to recipients as instructed, DHHS financial point of contact and DHHS and designated POCs |
| NLT 5th of each month |
| DHHS |
| ISO 17025 accreditation certificate |
| 1 |
| Electronic delivery to recipients as instructed |
| Availability of service begins immediately upon award |
| Laboratories located across the United States |
| Calibration and preventative maintenance as outlined in Performance and surveillance plan requirements |
| Twice a year, approximately every 6 months |
| Mail in service |
| Availability of service begins immediately upon award |
| Laboratories located across the United States |
| Provide ISO 17025 compliant calibration certificates |
| Each time service is requested |
| Electronic and hard copy delivery to requesting laboratory |
Electronic delivery to recipients as instructed Availability of service begins immediately upon award
| Laboratories located across the United States |
| Provide instructions on proper labeling and packaging of pipettes to be sent for calibration or repair |
| Each time service is requested |
| Electronic delivery to requesting laboratory |
| Availability of service begins immediately upon award |
Laboratories located across the United States
Return of calibrated pipettes within no more than 5 business days of receipt for calibration.
| Approximately 1100 single channel and 500 multichannel pipettes is anticipated |
| Shipping or delivery on demand during normal business hours; 3-day shipping or equivalent in speed and cost |
Availability of service begins immediately upon award
Laboratories located across the United States
| Quote, Diagnosis, anticipated repair completion date, cost estimate for repair, delivered within no more than 48 hours of receipt of pipette for repair. |
| Each time service is requested. |
| Electronic delivery to recipients as instructed |
| Availability of service begins immediately upon award |
Laboratories located across the United States
| Return of calibrated, functional pipette within no more than 10 business days of approval for repair. |
| Each time service is requested. |
Applies only when diagnosis deems repair is appropriate relative to replacement cost.
Shipping or delivery during normal business hours.
3-day shipping or equivalent in speed and cost
Availability of service begins immediately upon award
Laboratories located across the United States
| Contractor shall provide service scheduling outlined in Performance and surveillance plan requirements and tracking of existing service requests. |
| Beginning of year; updated as needed/requested |
| Online scheduling and tracking is preferred with access as instructed |
Electronic copy (minimum) provided to DHHS
Availability of service begins immediately upon award
Laboratories located across the United States
Status updates
| Each time service is requested or when new information is available from Contractor |
| DHHS and requesting lab |
| Availability of service begins immediately upon award |