Pipette Calibration Minimum Specifications:

The Environmental Chemistry Laboratory requires an automated or semi-automated system for calibrating microliter pipettes and syringes. The system shall have the following characteristics and attributes:

1. The detector shall be a spectrometric detector utilizing dual-dye, ratiometric photometry.

2. The detector must exhibit an extremely high, signal-to-noise ratio and high accuracy and precision at volumes as low as 100 nL.

3. The system must be able to complete a 10 point calibration, with documentation in less than 5 minutes.

4. Built in barcode printer and scanner to track pipette and syringes history and to identify reagents.

5. The system shall provide complete documentation, data management, and audit trail in compliance with 21 CFR Part 11.

6. The calibration method must be compliant with ISO 8655-7