Portable oil condition/analysis and viscosity combination testing kit

Salient Characteristics o Viscosity testing must meet the minimum criteria below:

Accuracy +/-3% Relative Standard Deviation to NIST viscosity standards Solvent free operation: No solvent needed to perform the test or clean the test cell.

Minimal amount of fluid required to test without impacting lubrication level Must be ASTM D8092 Compliant o Oil condition Monitor Infrared Spectrometer used for oil contamination testing Ability to calibrate to unique oil-specific algorithms Solvent free operation: No solvent needed to perform the test or clean the test cell.

Minimal amount of fluid required to test without impacting lubrication level Ability to measure engine oil total base number (TBN), water, glycol, soot, additive depletion, and oxidation Ability to measure total acid number (TAN), water, and oxidation in hydraulic and gearbox oils Must be ASTM D7889 compliant

• System must be portable, each individual piece of equipment with case should weigh approximately 20lb or less

• Stored in a weather resistant transportable case

• System must be battery operated with a minimum 6hrs of run time between charges

• Kit must include battery charger, USB cables, support software, user’s manual/programing instructions for both devices.

• Include pipettes and cleaning pads for a minimum of 100 samples