This is a Statement of Work (SOW) for repair and reconditioning services of the Cooperative Observer Program (COOP) Thermometer with Shelter Assembly (ASN: C450-2, NSN: 6685-01-285-4537), version dated September 1, 2022.
The contractor shall repair and recondition the thermometer units to "Like New" operating condition, meeting all electrical, mechanical, and cosmetic specifications per National Weather Service requirements. The scope includes disassembly, inspection, replacement of damaged or corroded parts, cleaning and repainting of shelter louvers and plates using Krylon 1315 white primer and Krylon 1501 high gloss white paint, thermistor testing and replacement if defective, and reassembly. The contractor must perform final temperature verification within ±0.5 degrees Fahrenheit of a known standard across the operational range of -56.0 to +122.0 degrees Fahrenheit using an NRC-supplied C451-N1 temperature display. All repair costs including parts, materials, labor, inspection, and return shipping are the contractor's responsibility, except thermistors which are supplied by NRC upon request and shelter louvers, insulated plates, and top plates which are government-furnished if requested. The contractor must develop detailed repair procedures and contact the Contracting Official immediately if unforeseen problems arise. Defective thermistors must be returned to NRC for validation. Units must be returned to the National Reconditioning Center in Grandview, Missouri, properly packaged to withstand transportation by ship, rail, truck, and aircraft. The contractor provides a one-year warranty against material and workmanship defects beginning after NRC final acceptance and inspection. Any units failing NRC Performance Quality Assurance testing must be returned to the contractor for rework at no cost to the government.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| CAR 1352.209-74.docx | DOCX document | |
| 52.222-48 Exemption from Appl of the SC Labor Standards.pdf | ||
| Evaluation Factors.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Statement of Work
Repair and Recondition: Cooperative Observer Program (COOP)
Thermometer with Shelter Assembly
ASN: C450-2, NSN: 6685-01-285-4537
Version: September 1, 2022
Description: The Contractor shall repair and recondition the Thermometer with Shelter
Assembly to "Like New" operating condition, including cosmetic appearance in accordance with
National Weather Service specifications described in the Description of Work – Work
Procedures. The repair and reconditioning price shall include the parts, materials, labor, inspection, and return shipping necessary to return the Thermometer with Shelter Assembly to operational specifications. Repair materials or specifications differing from the NWS source specification must be approved by the National Weather Service prior to use.
Definition: “Like New” Condition means reconditioned to the listed specification to meet all applicable electrical, mechanical, and cosmetic specifications and engineering documentation, including the replacement of damaged, missing, or non-functioning parts.
Government Furnished Equipment: The NRC will provide the C450-2 and C451-N1 MMTS
“Gold Standard” and any shelter louvers, insulated plates and top plates, if requested. Thermistors will be supplied in accordance with item 6.b. The vendor is responsible for all other parts and materials.
Government Furnished Data: Any documents or drawings identified in the Description of Work –
Work Procedures (upon request)
Vendor Furnished Equipment: Vendor shall provide all parts and supplies as required to accomplish the repair and reconditioning actions, expect where noted in the Description of Work
– Work Procedures.
Recommended Equipment and Supplies: The primer used should be Krylon 1315 white primer or equivalent and the paint used should be similar to High Gloss Krylon 1501.
Description of Work – Work Procedures:
The following steps may be accomplished in any order deemed necessary by the vendor. Steps may be skipped if not required. This is not intended to be a step by step description of the repair process.
The vendor will need to develop their own detailed repair procedures.
1. Unit shall be disassembled by removing the (4) M8 bolts, washer and lock washer that secure
(4) M8 x 270mm threaded rods running from top to bottom thru the assembly.
a. Care must be taken to not misplace the spacers that go between each section of louvers as the assembly is being taken apart.
b. Care must be taken to not damage thermistor wires during disassembly.
i. Thermistor is mounted to an inner thermistor mount using 5/16-18 stainless steel hardware. This hardware must be removed to allow the thermistor to be pulled out. Best practice would be to remove each section from the top plate down
(widest section w/acorn style nuts) so that the thermistor hardware can be removed and the thermistor and associated wiring will not be damaged.
ii. Connector terminals attached to the thermistor wiring will need to be removed from the connector to facilitate removing the thermistor completely from the unit.
These wires and pins are routed through the bottom three sections of louvers/plates. Pin removal tool can be provided upon request.
2. All stainless steel parts will be thoroughly inspected for wear and corrosion.
a. Parts showing wear or corrosion will be replaced with new stainless steel hardware of 18-
8 or 316 grade stainless steel.
3. Shelter louvers, insulated plates and top plates will be inspected for cracks, chips and any other visual signs of damage.
a. Parts showing such damage will be replaced by NRC at vendor’s requests.
4. Pipe flange (DWG: C450-A100-DR013) will be removed and inspected for corrosion and damage to pipe thread. If pipe flange is not original cast iron 2" pipe flange or replacement cast iron 2" pipe flange that has been modified to meet specifications of referenced drawing, vendor will replace at vendors expense.
a. If existing pipe cast iron flange is reused, it will need to be stripped, primed with Krylon
1315 white or equivalent and repainted with two coats of High Gloss Krylon 1501 white or equivalent. Pipe thread will be free of paint and have no defects in threads themselves.
b. If new pipe flange is needed, vendor shall supply cast iron pipe flange in accordance with
DWG: C450-A100-DR013, which can be substituted with colors stated in section 4.a.
above.
5. As needed, shield assembly plates and louvers will need to be cleaned and/or repainted to uniform color and color shall be Krylon 1501 white or equivalent.
a. These pieces are made of polycarbonate and proper preparation to surfaces should be made in order to prevent premature peeling of newly painted items.
6. Thermistor should be checked while disassembled from assembly for proper operation.
a. This is a resistive device and without a calibrated temperature source to use, an approximation of its reading and effectiveness will have to be done.
i. A resistive chart that can be supplied by the NRC can be used in conjunction with a volt-ohm meter to see if the thermistor is reading appropriately.
b. If thermistor is checked and found to be defective, NRC will supply a replacement thermistor to the vendor to use during reassembly. NRC will not pre-ship thermistors prior to determination of their operational status. Defective thermistors need to be returned to the NRC for validation of defective status with the returned units.
7. Once all parts are checked, replaced, painted, etc. proceed by reassembling the unit back to its operational state.
a. If needed, a drawing can be supplied that depicts a side cut away of the device in the way it should be assembled.
8. A final temperature checkout should be performed on each unit prior to sending it back to the
NRC using an NRC supplied C451-N1 temperature display.
a. If the temperature display reads a temperature that is +/- 0.5 degrees Fahrenheit compared to a known good temperature standard, then the unit can be sent back. (This measurement should be accomplished between the unit’s operational range: -56.0 degrees Fahrenheit to
+122.0 degrees Fahrenheit)
b. If the display unit reads 255.5 or -99.9, then there is a defect in the wiring or the thermistor itself and should be investigated.
Return fully functional and reconditioned Thermometer with Shelter Assemblies to NRC for evaluation by the NRC’s Technical Inspection and Material Reception Branch.
Warranty Terms: The Contractor will provide a 1-year warranty against material workmanship.
This 1-year warranty will begin after passing the final equipment inspection and acceptance for stock.
NRC Final Acceptance: All repaired and reconditioned Thermometer with Shelter Assemblies are subject to Performance Quality Assurance testing/inspection and will be returned for corrections if discrepancies are found. To meet the material workmanship requirements, the Thermometer with
Shelter Assemblies will need to be in a “Like New” operating condition and good appearance to include the unit’s paint job, cables, connectors, and associated hardware will be inspected. Shelter louvers, insulated plates and top plates will be inspected for cracks, chips and any other visual signs of damage. The NRC’s Technical Inspection and Material Reception Branch will verify units meet operational requirements by testing each unit for proper and accurate functionality in the NRC
Instrument Inspections Unit’s system. Any inspected item that does not meet the acceptance criteria will be returned to the vendor for rework at no additional cost to the government.
Shipping Address:
NATIONAL RECONDITIONING CENTER (NRC)
14200 MERRITT RD.
GRANDVIEW, MO. 64030
Transportability: Vendor is responsible for returning repaired products in a shipping container designed to withstand the vibration and shock encountered during transportation by ship, rail, truck, and aircraft. The units shall be packaged to prevent damage to include: physical, electrical, mechanical, cosmetic, and the units shall not be damaged or lose performance capability.
Final Note: When unforeseen problems arise, that may need clarification, please contact the
Contracting Official immediately. If the unit has any problem that prevents a quality repair, do not work on the item until the issue can be resolved.
NRC Reviewed:
WSRB Repair Branch Review: Jerry Forste, Branch Chief, September 1, 2022
TI&MR Branch Review: Gregory Flores, Branch Chief, September 1, 2022