Interference Filters and Blocking Filters Specifications

1) Narrowband Blocker Pair for 290.785 nm Qty = 2 Part A: Narrowband Filter

CWL: 290.785+0.050/-0 nm (in vacuum), FWHM: 1.0 ± 0.1 nm CWL: 290.700+0.050/-0 nm (as measured in air)

Peak Transmission: > 60% (Goal = 65%) Size: 25.4 mm +0, -0.25 mm diameter, mounted in black anodized aluminum ring Clear Aperture: 21.5 mm minimum (goal = 23 mm) Substrate Material = Fused Silica

CWL shift due to temperature </= 0.004 nm/deg.C Thickness including ring: ≤ 3.5 mm Substrate thickness: 3.0 ± 0.25 mm Angle of Incidence: 0º-2º (angle-tunable), with full cone angle of 2º

Operating Temperature: 23ºC

Part B: Blocker

Transmission @ 290.785nm: > 50% (Goal > 55%), (290.700+.050/-0 nm as measured in air)

Clear Aperture: 21.5 mm minimum (goal = 23 mm) Substrate Material = Fused Silica Thickness including ring: ≤ 3.5 mm Substrate thickness: 3.0 ± 0.25 mm Angle of Incidence: 0º-2º (angle-tunable) Operating Temperature: 23ºC

Part A and Part B series performance: Combined blocking of both parts: 1E-5 transmission for 260-700 nm wavelength range. Each part will be measured separately. Parts will not be measured in series. Specs will be met at 0° AOI.

Deliverable Data:

1) In-band evaluation per filter, confirm peak T%, CWL, FWHM

2) Out-of-band blocking on 1 filter per batch ONLY

3) Spectral data supplied in both graphic and digital formats

4) Inspection report (measured dimensions)

2) Narrowband Blocker Pair for 300.488 nm Qty = 2 Part A: Narrowband Filter

CWL: 300.488+0.050/-0 nm (in vacuum), FWHM: 1.0 ± 0.1 nm

CWL: 300.400+0.050/-0 nm (as measured in air) Peak Transmission: > 65% (Goal > 70%) Size: 25.4 mm +0, -0.25 mm diameter, mounted in black anodized aluminum ring

Clear Aperture: 21.5 mm minimum (goal = 23 mm)

Substrate Material = Fused Silica

CWL shift due to temperature </= 0.004 nm/deg.C Thickness including ring: ≤ 3.5 mm

Transmission @ 300.488 nm: > 70% (Goal > 75%), (300.400+.050/-0 nm as measured in air)

Clear Aperture: 21.5 mm minimum (goal = 23 mm)

Substrate Material = Fused Silica Thickness including ring: ≤ 3.5 mm Substrate thickness: 3.0 ± 0.25 mm Angle of Incidence: 0º-2º (angle-tunable) Operating Temperature: 23ºC Part A and Part B series performance: Combined blocking of both parts: 1E-5 transmission for 260-700 nm wavelength range. Each part will be measured separately. Parts will not be measured in series. Specs will be met at 0° AOI.

Deliverable Data:

1) In-band evaluation per filter, confirm peak T%, CWL, FWHM

2) Out-of-band blocking on 1 filter per batch ONLY

3) Spectral data supplied in both graphic and digital formats

4) Inspection report (measured dimensions)

3) Narrowband Blocker Pair for 304.092 nm Qty = 2

CWL: 304.092+0.050/-0 nm (in vacuum), FWHM: 1.0 ± 0.1 nm

CWL: 304.003+0.050/-0 nm (as measured in air) Peak Transmission: > 65% (Goal > 70%) Size: 25.4 mm +0, -0.25 mm diameter, mounted in black anodized aluminum ring

Clear Aperture: 21.5 mm minimum (goal = 23 mm)

Substrate Material = Fused Silica

CWL shift due to temperature </= 0.004 nm/deg.C Thickness including ring: ≤ 3.5 mm Substrate thickness: 3.0 ± 0.25 mm

Angle of Incidence: 0º-2º (angle-tunable), with full cone angle of 2º

Transmission @ 304.092 nm: > 70% (Goal > 75%), (304.003+.050/-0 nm as measured in air)

Clear Aperture: 21.5 mm minimum (goal = 23 mm)

Substrate Material = Fused Silica Thickness including ring: ≤ 3.5 mm Substrate thickness: 3.0 ± 0.25 mm Angle of Incidence: 0º-2º (angle-tunable) Operating Temperature: 23ºC Part A and Part B series performance: Combined blocking of both parts: 1E-5 transmission for 260-700 nm wavelength range. Each part will be measured separately. Parts will not be measured in series. Specs will be met at 0° AOI.

Deliverable Data:

1) In-band evaluation per filter, confirm peak T%, CWL, FWHM

2) Out-of-band blocking on 1 filter per batch ONLY

3) Spectral data supplied in both graphic and digital formats

4) Inspection report (measured dimensions)

4) Narrowband Blocker Pair for 316.418 nm Qty = 2

CWL: 316.418+0.050/-0 nm (in vacuum), FWHM: 1.0 ± 0.1 nm

CWL: 316.326+0.050/-0 nm (as measured in air) Peak Transmission: > 65% (Goal > 70%) Size: 25.4 mm +0, -0.25 mm diameter, mounted in black anodized aluminum ring

Clear Aperture: 21.5 mm minimum (goal = 23 mm)

Substrate Material = Fused Silica

CWL shift due to temperature </= 0.004 nm/deg.C Thickness including ring: ≤ 3.5 mm

Transmission @ 316.418 nm: > 70% (Goal > 75%), (316.326+.050/-0 nm as measured in air)

Clear Aperture: 21.5 mm minimum (goal = 23 mm)

Substrate Material = Fused Silica Thickness including ring: ≤ 3.5 mm Substrate thickness: 3.0 ± 0.25 mm Angle of Incidence: 0º-2º (angle-tunable) Operating Temperature: 23ºC Part A and Part B series performance: Combined blocking of both parts: 1E-5 transmission for 260-700 nm wavelength range. Each part will be measured separately. Parts will not be measured in series. Specs will be met at 0° AOI.

Deliverable Data:

1) In-band evaluation per filter, confirm peak T%, CWL, FWHM

2) Out-of-band blocking on 1 filter per batch ONLY

3) Spectral data supplied in both graphic and digital formats

4) Inspection report (measured dimensions)

5) Narrowband Blocker Pair for 354.862 nm Qty = 1

Part A: Narrowband Filter

CWL: 354.862 +0.050/-0 nm (in vacuum), FWHM: 0.3 ± 0.1 nm

CWL: 354.761 +.050/-0 nm (as measured in air)

Peak Transmission: > 80% (Goal > 85%) Size: 25.4 mm +0, -0.25 mm diameter, mounted in black anodized aluminum ring

Clear Aperture: 21.5 mm minimum (goal = 23 mm)

CWL shift due to temperature = 0.002 nm/deg.C

Thickness including ring: ≤ 3.5 mm

Substrate thickness: 3.0 ± 0.25 mm

Angle of Incidence: 0º-2º (angle-tunable), with full cone angle of 2º

Operating Temperature: 23ºC

Part B: Blocker

Transmission @ 354.862 nm: > 90% (Goal > 95%), (354.761 +.050/-0 nm as measured in air)

Size: 25.4 mm +0, -0.25 mm diameter, mounted in black anodized aluminum ring

Clear Aperture: 21.5 mm minimum (goal = 23 mm)

Thickness including ring: ≤ 3.5 mm

Substrate thickness: 3.0 ± 0.25 mm

Angle of Incidence: 0º-2º (angle-tunable), with full cone angle of 2º

Operating Temperature: 23ºC

Part A and Part B series performance: Combined blocking of both parts: 1E-5 transmission for 200-700 nm wavelength range. Each part will be measured separately. Parts will not be measured in series. Specs will be met at 0° AOI.

Deliverable Data:

1) In-band evaluation per filter, confirm peak T%, CWL, FWHM

2) Out-of-band blocking on 1 filter per batch ONLY

3) Spectral data supplied in both graphic and digital formats

4) Inspection report (measured dimensions)

6) Narrowband Blocker Pair for 532.293 nm Qty = 1

Part A: Narrowband Filter CWL: 532.293 +0.050/-0 nm (in vacuum), FWHM: 0.4 ± 0.1 nm

CWL: 532.145 +.050/-0 nm (as measured in air)

Peak Transmission: > 85% (Goal > 90%) Size: 25.4 mm +0, -0.25 mm diameter, mounted in black anodized aluminum ring

Clear Aperture: 21.5 mm minimum (goal = 23 mm)

CWL shift due to temperature = 0.003 nm/deg.C

Thickness including ring: ≤ 3.5 mm

Substrate thickness: 3.0 ± 0.25 mm

Angle of Incidence: 0º-2º (angle-tunable), with full cone angle of 2º

Operating Temperature: 23ºC

Part B: Blocker

Transmission @ 532.293 nm: > 95% (532.145 +.050/-0 nm as measured in air)

Size: 25.4 mm +0, -0.25 mm diameter, mounted in black anodized aluminum ring

Clear Aperture: 21.5 mm minimum (goal = 23 mm)

Thickness including ring: ≤ 3.5 mm

Substrate thickness: 3.0 ± 0.25 mm

Angle of Incidence: 0º-2º (angle-tunable), with full cone angle of 2º

Operating Temperature: 23ºC

Part A and Part B series performance: Combined blocking of both parts: 1E-5 transmission for 200-950 nm wavelength range. Each part will be measured separately. Parts will not be measured in series. Specs will be met at 0° AOI.

Deliverable Data:

1) In-band evaluation per filter, confirm peak T%, CWL, FWHM

2) Out-of-band blocking on 1 filter per batch ONLY

3) Spectral data supplied in both graphic and digital formats

4) Inspection report (measured dimensions)

7) Narrowband Blocker Pair for 1064.585 nm Qty = 1

Part A: Narrowband Filter CWL: 1064.585 +0.050/-0 nm (in vacuum), FWHM: 0.4 ± 0.1 nm CWL: 1064.293 +.050/-0 nm (as measured in air) Peak Transmission: > 85% (Goal > 90%) Size: 38.1 mm +0, -0.25 mm diameter, mounted in black anodized aluminum ring Clear Aperture: 34 mm minimum Thickness including ring: ≤ 3.5 mm Substrate thickness: 3.0 ± 0.25 mm

Angle of Incidence: 0º-2º (angle-tunable), with full cone angle of 2º Operating Temperature: 23ºC

CWL shift due to temperature: < 0.005 nm/°C

Transmission @ 1064.585 nm: > 95% (1064.293 +.050/-0 nm as measured in air) Size: 38.1 mm +0, -0.25 mm diameter, mounted in black anodized aluminum ring Clear Aperture: 34 mm minimum Thickness including ring: ≤ 3.5 mm

Substrate thickness: 3.0 ± 0.25 mm

Angle of Incidence: 0º-2º (angle-tunable), with full cone angle of 2º

Operating Temperature: 23ºC

Part A and Part B series performance: Combined blocking of both parts: 1E-5 transmission for 200-1150 nm wavelength range. Each part will be measured separately. Parts will not be measured in series. Specs will be met at 0° AOI.

Deliverable Data:

1) In-band evaluation per filter, confirm peak T%, CWL, FWHM

2) Out-of-band blocking on 1 filter per batch ONLY

3) Spectral data supplied in both graphic and digital formats

4) Inspection report (measured dimensions)

8) Dichroic Beamsplitter for 290.785 / 300.488 nm Qty = 1 Angle of Incidence = 45 deg, with a full-field of view of 2° (35 mrad).

Reflect 290.785 nm (vacuum) at R>95% (Goal > 99%) for S and P polarizations.

Transmit 300.488 nm (vacuum) at T> 85% (Goal > 90%) for S and P polarizations.

Substrate diameter: 50.8 mm diameter +0/-0.25 mm, unmounted Substrate thickness: 6.35 mm +/- 0.4 mm Clear Aperture: 45 mm minimum

Substrate Material = Fused Silica The operating and measurement evaluation temperature of the filter shall be 23ºC+/- 4ºC.

Vendor shall provide modeled data showing temperature extremes and 2 deg FOV.

Surface quality: F-F per MIL-C_48497A

Deliverable Data:

1) Measured transmission at 45°

2) Reflection verified by analysis, based on T data

3) Spectral data supplied in both graphic and digital formats

4) Inspection report (measured dimensions)

5) Certificate of Conformance

9) Dichroic Beamsplitter for 304.092 / 316.418 nm Qty = 1

Angle of Incidence = 45 deg, with a full-field of view of 2° (35 mrad).

Reflect 304.092 nm (vacuum) at R>95% (Goal > 99%) for S and P polarizations.

Transmit 316.418 nm (vacuum) at T> 85% (Goal > 90%) for S and P polarizations.

Substrate diameter: 50.8 mm diameter +0/-0.25 mm, unmounted Substrate thickness: 6.35 mm +/- 0.4 mm Clear Aperture: 45 mm minimum

Substrate Material = Fused Silica The operating and measurement evaluation temperature of the filter shall be 23ºC+/- 4ºC.

Vendor shall provide modeled data showing temperature extremes and 2 deg FOV.

Surface quality: F-F per MIL-C_48497A

Deliverable Data:

1) Measured transmission at 45°

2) Reflection verified by analysis, based on T data

3) Spectral data supplied in both graphic and digital formats

4) Inspection report (measured dimensions)

5) Certificate of Conformance

10) Dichroic Beamsplitter for UV / VIS-NIR

Angle of Incidence: 45°, ± 1° field of view

R > 95% (goal > 99%) for S and P polarizations at the following wavelengths & ranges in vacuum: 290.8 nm, 300.5 nm, 304.1 nm, 316.4 nm, 354.9 nm, 394-410 nm, 435-455 nm.

T > 90% (goal > 95%) for S and P polarizations at the following wavelengths & ranges in vacuum: 532.3 nm, 640-660 nm, 670-690 nm, and at 1064.6 nm.

Substrate diameter: 50.8 mm diameter +0/-0.25 mm, unmounted

Clear Aperture: 45 mm minimum

Thickness: 6.35 ± 0.4 mm

Substrate material: Fused silica

The operating and measurement evaluation temperature of the filter shall be 23ºC+/- 4ºC.

Vendor shall provide modeled data showing temperature extremes and 2 deg FOV.

Surface quality: F-F per MIL-C_48497A

Deliverable Data:

1) Measured transmission at 45°

2) Reflection verified by analysis, based on T data

3) Spectral data supplied in both graphic and digital formats

4) Inspection report (measured dimensions)

5) Certificate of Conformance

File details come from the government source that posted it. Updated .