Minimum Technical Requirements for Air Data Controller
Two independent absolute pressure transducers, Ps and Pt (Qc derived) Aeronautic and engineering units, including psia (Ps,Pt), psid (Qc) and User Defined Ps > 100,000 ft. alt. to minimum 36 inHga, Pt < 1.0 inHga, to > 900 kts / 50 inHga Capable of concurrent independent Measure/Control of Ps/Pt RVSM Compliant Multi-gas species capable, minimum GN2 and Air Must pressurize and control with stability a 5187 cu. In. / 87-liter volume and provide documentation demonstrating this capability (such as testing output results) UUT head correction input IEEE-488 bus SCPI for automated operation Provide for in-house Ps/Pt calibration by owner Must be 19” rack mountable