This is a Minimum Requirements document for an automated inverted microscope system acquisition by the National Institute of Standards and Technology (NIST). NIST's Biophysical and Biomedical Measurement Group seeks a complete microscope system to support its Organ-on-a-Chip and Microphysiological Systems research program, with emphasis on live cell imaging capabilities and compatibility with existing Zeiss microscope infrastructure in the laboratory.
The system comprises eight line items with detailed technical specifications: (1) an inverted microscope body with motorized focus drive (20 nm minimum step width), motorized nosepiece, filter turret, and motorized X-Y scanning stage (100 mm travel range, 0.5 mm resolution) capable of bright field and reflection imaging plus phase contrast and differential interference contrast microscopy; (2) four microscope objectives ranging from 2x to 20x magnification with specified numerical aperture and working distance requirements; (3) a CMOS camera with greater than 15 megapixels resolution, forced-air cooling to 25°C, USB 3 and PCIe interfaces, 20+ fps frame rate, C-mount lens mount, and 1-16x signal amplification; (4) a multi-channel LED illumination source with integrated control unit and minimum 4-channel fluorescent capability (7-channel preferred) covering UV through far-red wavelengths (385-735 nm) with filter sets for standard fluorophores; (5) microscope control software with graphical user interface supporting image acquisition, processing, automation of motorized components, time-lapse imaging, z-stack capability, and compatibility with Zeiss software; (6) heating and temperature control unit (37°C for mammalian cultures) and CO2 control unit; (7) a workstation-class personal computer with Intel Xeon processor, 32 GB RAM minimum, dedicated Nvidia graphics card with 16 GB memory, 2 TB SSD storage, and 30-inch monitor; and (8) interchangeability of parts and software with existing Zeiss microscopes. All equipment must be new, manufacturer-packaged, and include original documentation and software.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
MINIMUM REQUIREMENTS
The Contractor shall provide a system that meets all technical specifications identified below. All items must be new. Used or remanufactured equipment will not be considered for award. Experimental, prototype, or custom items will not be considered. The use of “gray market” components are not acceptable. All line items shall be shipped in the original manufacturer’s packaging and include all original documentation and software, when applicable.
Line Item 001:
Description: Microscope Body Quantity: 1 A. Technical Specifications
a. Inverted microscope body with the following features:
i. One or two side ports to attach cameras
ii. Motorized focus drive with minimum step width of 20 nm or better
iii. Motorized nose piece to automatically switch objectives
iv. Motorized filter turret
v. All optics to allow bright field illumination in transmission AND reflection modes.
vi. Binocular tube and eyepieces when required.
vii. Motorized condenser capable of phase contrast and DIC microscopy.
b. Focus tracking: Ability to use an infrared light emitting diodes (LED) to independently and continually track microscope focus drift independent of imaging channels.
c. Sample Stage: Motorized X-Y scanning stage with the following features:
i. Travel range: ≥100 mm (X-axis) and ≥100 mm (Y-axis)
ii. Resolution: better than 0.5 μm along each axis.
iii. Mounting surface: adapt to allow standard microscope slides (25 mm x 75 mm) or multi-well plates.
Line Item 002:
Description: Microscope Objectives Quantity: Total 4, one each with the specifications below A. Technical Specifications All objectives shall be optimized for visible light illumination.
Line Item 003:
Description: Camera Quantity: 1 A. Technical Specifications
a. Detector: Complementary metal-oxide semiconductor (CMOS)
b. Resolution: > 15 Megapixels
c. Cooling method: forced air; stabilize to 25 °C.
d. Interface: USB 3 with a separate PCIe acquisition interface for fast frame rate readout.
e. Frame rate: fast frame readout > 20 fps.
f. Lens mount: C-mount
g. Exposure Time Range: 0.1 ms to 60 s
h. Hardware Trigger: Trigger-in for exposure time control, trigger-out for exposure time
i. Software: Fully integrated into microscope interface
j. Minimum pixel size: 3 μm x 3 μm or better
k. Spectral Range: 350 nm to 1000 nm
l. Adjustable Signal Amplification: from 1x to up to 16x
Line Item 0004:
Description: Illumination Sources and Filter Sets Quantity: 1 A. Technical Specifications
a. Multi-channel light emitting diode (LED) source that replaces traditional mercury vapor sources.
b. Integrated control unit for continuous brightness adjustment, quickly switchable, and stable for long term experiments.
| Magnification |
| Minimum Numerical Aperture (NA) |
| Minimum Working Distance (WD) |
| Immersion Media |
| 2 or 2.5x |
| 0.06 or higher |
| 9 mm or higher |
| Air |
| 5x |
| 0.06 or higher |
| 9 mm or higher |
| Air |
| 10X |
| 0.45 |
| 2 mm or higher |
| Air |
| 20x |
| 0.55 |
| 0.55 mm or higher |
| Water* or oil |
c. The combination of source and filters must provide flat illumination across channels.
d. Minimum of 4-channel fluorescent light source, but up to 7-channel fluorescent excitation lines are preferred.
e. The preferred 7 channels with excitation filters centered at the following wavelengths are:
i. UV (385 nm)
ii. Violet (423 nm)
iii. Blue (469 nm)
iv. Green (555 nm)
v. Yellow (591 nm)
vi. Red (631 nm)
vii. Red (735 nm)
f. Filter sets, matching filter cubes and all related accessories to measure commonly used fluorophores including: for UV (385 nm) - DAPI, Alexa 405, Hoechst 33258 and similar dyes; for Violet (423 nm) - eCFP, Lucifer Yellow, Alexa 430 and similar dyes; for Blue (469 nm) - eGFP, Fluo4, FITC and similar dyes; for Green(555 nm) - Cy3, TRITC, DsRed and similar dyes; for Yellow (591 nm) for excitation of mCherry, Alexa 568, mPlum and other similar dyes; for Red (631 nm) - Cy5, Alexa 631, TOTO-3 and similar dyes; and for Far Red (735 nm) for excitation of Cy7 and other similar dyes.
Line Item 0005:
Description: Software Quantity: 1 A. Technical Specifications
a. Software with graphical user interface to operate microscope, all imaging modalities, attached cameras, and accessories supplied by the vendor.
b. Image acquisition and processing to allow data capture, storage and post-processing of brightfield and fluorescence images.
c. Ability to export images to standard image formats (e.g., TIFF, JPEG, etc.)
d. Automation of all motorized components to allow time lapse imaging, sequencing and long-term measurements, z-stack, automated focus tracking, and image tiling.
e. High dynamic range imaging with the ability to merge images with different excitation intensities.
f. Ability to script or run microscope automation.
g. Compatibility of data processing software with Zeiss software microscopes.
Line Item 0006:
Description: Heating/Temperature Control and Carbon Dioxide (CO2) Control Units Quantity: 1 each A. Technical Specifications
a. Heating/Temperature Control Unit to support mammalian cell cultures at 37 °C for days and weeks.
b. CO2 Control Unit to maintain carbon dioxide conditions conventionally used in mammalian cell cultures.
Line Item 0007:
Description: Workstation Quantity: 1 A. Technical Specifications
a. Workstation class personal computer with Intel Xeon processor
b. 32 GB RAM or greater
c. Dedicated graphics card (Nvidia or equivalent) with 16 GB memory or greater.
d. A single monitor with a screen measuring 30” or greater diagonally.
e. All control cards necessary to operate microscope hardware.
f. SSD of 2 TB or more
Line item 0008:
Description: Interchangeability of parts and software We require that the system described above (automated inverted microscope system) be compatible with other existing Zeiss microscopes in the lab to enable exchanging components and analyzing data between microscopes (i.e., software must be compatible with Zeiss software for imaging acquisition and analysis).