Specification and minimum requirement
| Part Number |
| Description |
| Quantity |
| 266002-1160-391 |
| ZEISS OS | CIRRUS HD-OCT (Model 5000) w/ |
AngioPlex OCT Angiography Version 9.0. Includes Smart HD Scans, FastTrac, Adv RPE,GCA, Printer & Table | EA
| 266002-1157-764 |
| Anterior Segment Premier Module |
| 1 |
Trade in of an existing Cirrus 4000 Machine
Additional Technical specification:
· Unit must have the capability to capture and provide OCT Angiography
· This capability for OCT Angiography is provided by a 1 single OCT scan that allows for visualization of both micro vascular and structural information in a noninvasive dye free scan by the OCT.
· Has the ability to import and integrate the Cirrus Model HD OCT 4000 raw data. The raw data must be interchangeable between the Cirrus 4000 and Cirrus 5000 unit
· The unit must have exact Macular Scan Pattern of 512x128 for retina scan with exact scanning speed at 27,000 A-scans a second within a 6mmx6mm
· Unit must produce exact scan pattern of 200x200 in a 6mmx6mm cube for Glaucoma Scan with exact scanning speed at 27,000 A-scans a second. The scan pattern must be able to capture and provide RNFL, ONH, Neuro Retinal Rim Tissue, and 3-D Cube capabilities all in one scan
· Scanning speeds: Scan at exactly 27,000 A-scans/sec, as well as the ability to scan at 68,000 A-scans/sec
· Cell Analysis that is derived from a scan must be comprised of 6mmx6mm cube scan with a minimum by 200x200 line scan pattern with no more than 30 microns between each slice
· Minimum Pupil Requirements must be 2.0mm or less.
· Unit must utilize an LSO Fundus Camera (Line Scanning Laser Ophthalmoscope). The unit must Automatically Track to Prior Scan: The unit must be able to do this without technician needing to align to prior scan:
· Anterior Segment scan that is at minimal 15.5 mm in length x 5.8 mm in depth. which is the full view of the Anterior Chamber. Must be able to provide on Company Literature.
· Global Pachymetry report that is comprised of minimum of 24 radial scans and must be at a minimum 9mm in length. The unit must be a 90 degree configuration, so that the tech (operator) must sit to the side (90 degrees)
· OCT Imaging: Methodology - Spectral domain OCT Optical source - Superluminescent diode (SLD), 840 nm Scan speed - 27,000 A-scans per second or 68,000 A-scans per second* *All existing scan patterns in version 6.5 software will run at 27,000 A-scans per second, which is the same scan speed as the CIRRUS 4000
· Axial resolution - 5 um (in tissue)
· Transverse resolution - 15 um (in tissue) Cube scan pattern - 512 x 126 with over >67 million data ports and 47um spacing between lines
· Cube scan pattern - 200 x 200 with over >40 million data ports and 30um spacing between lines
Fundus Imaging:
· Methodology - Line scanning ophthalmoscope (LSO)
· Live fundus image - during alignment and during OCT scan
· Optical source - Superluminescent diode (SLD), 750nm
· Field of view-36 degrees Wx 30 degrees H
· Frame rate - >20 Hz
· Transverse resolution - 25 um (in tissue)
Iris Imaging:
· Methodology - CCD Camera
· Resolution - 1280 x 1024
· Live iris image - During alignment
· Electrical and Physical:
· Weight - 80 lbs.
· Dimensions of instrument - 26L x 18Wx 21H (in)
· Dimensions of table - 39L X 22W(in)
Fixation:
· Internal fixation focus adjustment - -20D to +20D (diopters)
· Electrical rating (115V) - Single Phase, 100-120V~systems: 50/60Hz, 5A
· Electrical rating (230V) - Single Phase, 200-240V~systems: 50/60Hz, 2.5A
Internal Computer:
· Operating system/processor - Windows 7, i& Intel processor
Memory:
· Hard drive/internal storage - >=750 GB, > 80,000 scans
· Display - Integrated 19" color flat panel display USB ports - 6 ports
Proposal must also include the following:
· One-year manufacturer warranty
· delivery
· Training
· Two operators’ manuals
· Two service manuals
Delivery detail:
National Eye Institute / NIH 10 Center Drive
BLDG 10 RM 10C306
Bethesda, MD 20892