5 Each Animal Intensive Care System Units.
Solicitation number
NIH-OLAO-OD3-NOI-5435228
Issued by

About this file

Required items description

View the file

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Animal Intensive Care System Units Specification 5 Each Animal Intensive Care System Units.

Specs:

· ICS-I (I.D):

· 41"w x 28"d x 31.5"h

· 104,14 x 78,12 x 80 cm

· O.D. *:

· 59"w x 34.5"d x 74.25"h

· 149,86 x 87,63 x 188,6 cm

Features

· Each unit is enclosed within stainless steel for durability, ease of cleaning, and sanitation

· The entire unit is mounted on a steel frame for structural integrity

· The animal compartment is a separate unit attached to the frame which provides for replacement should the cage be damaged

· An oxygen analyzer is included in the package

· Heavy duty stainless steel grounded casters with brakes, locks, and zerk fittings are standard

· Operators' manual and limited warranty Performance Specifications

· Rh level increased only. Not reduced.

· Animal capacity - up to 125 pounds (ICS-l)

· Cage temperature - variable from 55 degrees F to 120 degrees F

· Temperature accuracy - plus/minus 2 degrees F

· Air change rate - 1 cage volume/minute

· C02 absorption capacity - 24 hours

· Temperature stabilization - approximately 15 minutes heat/20 minutes cool

· Power requirements: 220-240 Volts

· Amperes - 15 max.

· Decibel level: Inside: 72dba / Outside: 82 dba