Technical Specifications: Master Oscillator Power Amplifier SLD System
Optical Specifications
Center Wavelength: 850 +/- 2 nm
Optical Bandwidth (FWHM): 10-20 nm
Spectral Ripple: < 2%
Output Power: 50 mW (45 mW min.)
Maximum Allowed Feedback: -10 dB (corresponds to 10%-feedback)
Ouput Power Stability: < 0.5%
Polarization Extinction Ratio: > 18 dB (20 dB typ.)
Optical Fiber Type: Corning PANDA PM 850
Polarization Orientation in the Output Fiber: Slow axis, Aligned with the connector key
Output Optical Connector: FC/APC type with narrow precision key (2.0 mm)
Electrical Specifications
Operating Mode, for SLD / SOA: Constant Power
Enable Control: 1) Manual: from the front panel 2) Electronical via: a) TTL pulses, b) RS-
232 port
Modulation Input: TTL, 0 – 5 V => Full Power – 0, RIN=1 kOhm
Rise Time / Fall Time: 2 s / 0.5 s
Modulation Input Connector: SMA-female (Referenced to Chassis)
I/O Interface: RS-232
Protective Measures
SLD Protective Measures: Soft start of the SLD pumping current, clamping limit circuit for the SLD current, SLD operation in constant power mode, fast switch off of an SLD module in case of a driver error, effective suppression of electronic transients and spikes.
Built-in Laser Safety Measures: Key-operated master control, remote interlock connector, manual reset in case of remote interlock activation, visual indication of laser output activation, laser safety information labels.
General Specifications
Power Requirements: 110 VAC, 50-60 Hz
Power Consumption: 20 VA Max
Warm-up Time: 10 min
Operating Temperature Range: +15 °C to + 30 °C
Storage Temperature Range: 0 °C to +40 °C
Continuous Operation: 16 hrs/day
Outline Dimensions (W × H × D): 257 × 170 × 325 mm
Approximate Weight: 7 kg
Warranty: 12 months
Accessories
The list of accessories below is required to be supplied with the instrument.
Item Quantity
PM Fiber patch cable of 1 m long for the device's optical output 2 pcs.
Key for the device's master control 2 pcs.
Short-circuited remote interlock connector (inserted in the rearpanel 1 pc.
socket of the device)
AC power cord 1 pc.
RS-232 null-modem cable 1 pc.
CD with the user's manual and Superlum's companion software 1 pc.
Acceptance test report (printed version) 1 pc.
Capabilities Statement Submission Process
Email Capabilities Statements to eric.chapman@nih.gov.
Capabilities Statements Submission Deadline
Tuesday, June 28, 2016, 11:59 PM EST
Acquisition Contact
Eric Chapman, Contract Specialist, 301-435-3910, eric.chapman@nih.gov mailto:eric.chapman@nih.gov mailto:eric.chapman@nih.gov
File details come from the government source that posted it. Updated .