Technical Specifications: Master Oscillator Power Amplifier SLD System

Optical Specifications

Center Wavelength: 850 +/- 2 nm

Optical Bandwidth (FWHM): 10-20 nm

Spectral Ripple: < 2%

Output Power: 50 mW (45 mW min.)

Maximum Allowed Feedback: -10 dB (corresponds to 10%-feedback)

Ouput Power Stability: < 0.5%

Polarization Extinction Ratio: > 18 dB (20 dB typ.)

Optical Fiber Type: Corning PANDA PM 850

Polarization Orientation in the Output Fiber: Slow axis, Aligned with the connector key

Output Optical Connector: FC/APC type with narrow precision key (2.0 mm)

Electrical Specifications

Operating Mode, for SLD / SOA: Constant Power

Enable Control: 1) Manual: from the front panel 2) Electronical via: a) TTL pulses, b) RS-

232 port

Modulation Input: TTL, 0 – 5 V => Full Power – 0, RIN=1 kOhm

Rise Time / Fall Time: 2 s / 0.5 s

Modulation Input Connector: SMA-female (Referenced to Chassis)

I/O Interface: RS-232

Protective Measures

SLD Protective Measures: Soft start of the SLD pumping current, clamping limit circuit for the SLD current, SLD operation in constant power mode, fast switch off of an SLD module in case of a driver error, effective suppression of electronic transients and spikes.

Built-in Laser Safety Measures: Key-operated master control, remote interlock connector, manual reset in case of remote interlock activation, visual indication of laser output activation, laser safety information labels.

General Specifications

Power Requirements: 110 VAC, 50-60 Hz

Power Consumption: 20 VA Max

Warm-up Time: 10 min

Operating Temperature Range: +15 °C to + 30 °C

Storage Temperature Range: 0 °C to +40 °C

Continuous Operation: 16 hrs/day

Outline Dimensions (W × H × D): 257 × 170 × 325 mm

Approximate Weight: 7 kg

Warranty: 12 months

Accessories

The list of accessories below is required to be supplied with the instrument.

Item Quantity

PM Fiber patch cable of 1 m long for the device's optical output 2 pcs.

Key for the device's master control 2 pcs.

Short-circuited remote interlock connector (inserted in the rearpanel 1 pc.

socket of the device)

AC power cord 1 pc.

RS-232 null-modem cable 1 pc.

CD with the user's manual and Superlum's companion software 1 pc.

Acceptance test report (printed version) 1 pc.

Capabilities Statement Submission Process

Email Capabilities Statements to eric.chapman@nih.gov.

Capabilities Statements Submission Deadline

Tuesday, June 28, 2016, 11:59 PM EST

Acquisition Contact

Eric Chapman, Contract Specialist, 301-435-3910, eric.chapman@nih.gov mailto:eric.chapman@nih.gov mailto:eric.chapman@nih.gov

File details come from the government source that posted it. Updated .