NB688000-19-02847

Specifications for a Magnetic Declinometer/Inclinometer System

A magnetic Declinometer/Inclinometer System (D/I System) is a specialized measurement tool allowing for accurate measurements the declination and inclination (the direction) and magnitude of the earth’s magnetic field.

This instrument is needed as part of a calibration system for sensitive and accurate magnetometers made by our group at NIST. It is necessary to perform some calibrations at remote sites away from magnetic contamination due to buildings, electrical systems, automobiles, etc. It become necessary to accurately measure the magnetic field magnitude and direction at the test site to compensate for any bias.

A D/I system consists of a single-axis magnetometer mounted on a steel-free theodolite and steel-free tripod. The theodolite is a surveying instrument that provides measurements of the declination and inclination of the magnetic field. The steel-free (non-magnetic) components are crucial otherwise the magnetic field due to the surveying instruments will corrupt the measurements.

Detailed Specifications

Single axis Magnetometer and Probe System

Measurement Range +/- 0.2 milliTesla

Bandwidth DC to 10 Hz

Offset in Zero Field +/-1 nanoTesla

Operating Temperature Range 0 to 50 Celsius

Maximum Resolution 0.1 nanoTesla

Calibration Uncertainty 0.25%

Steel-Free Theodolite and Tripod

All non-magnetic components in construction

Optical/Magnetic Collimation error <20 sec

Directional Accuracy 3 sec