Summary: Requirements/Specifications Document for AlH+ Photodissociation Laser
This is a Requirements/Specification document issued by the National Institute of Standards and Technology (NIST) Physical Measurements Lab, Time and Frequency Division, Ion Storage Group (Reference: NB688000-26-02196). The Ion Storage Group requires one AlH+ photodissociation laser to support development of optical frequency standards and precision measurements with trapped ions.
The laser must meet the following technical specifications: continuous-wave (CW) operation at a wavelength of 360.4 nm ± 0.1 nm with minimum power output of 20 mW at that wavelength. Power stability must not exceed 2% over 8 hours in a temperature-controlled laboratory environment (dT < 0.5°C). The beam must be a free-space beam in TEM00 spatial mode with M² ≤ 1.1 and divergence ≤ 1.0 mrad. The system must include all necessary cables and heat sinks, incorporate USB and RS-232 ports for computer control, and feature a compact size with plug-and-play design requiring minimal user alignment. Delivery of goods, installation, and training must be completed within 06 weeks ARO. The system shall be installed by a vendor technician to confirm performance, and government acceptance testing is expected to require 14 business days. Software and electronic manuals must be provided. Base manufacturer warranty is acceptable. Laboratory users require training in alignment, optimization, and operation of the frequency doubling unit. Standard power availability of 60 Hz, 120 V is noted as typical for the facility.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| NB688000-26-02196 RFQ LPTA.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Re qu ire m en ts
Sp ec ifi ca tio n Do cu m en t
Requirements / Specifications: AlH+ Photodissociation Laser 1 Version 1.0 (2019) Double-click on for further assistance.
Use this document to detail specifications/requirements for an item and/or ancillary services.
TITLE: AlH+ Photodissociation Laser Requesting Lab/Division/Group: Physical Measurements Lab/Time and Frequency Division/ Ion Storage Group
NB688000-26-02196
General Statement of Need The Ion storage group requires AlH+ photodissociation laser to support development of optical frequency standards and precision measurements with trapped ions.
Items and/or Ancillary Services
Item 1: AlH+ Photodissociation Laser Quantity: One (1) Specifications:
1.1 - CW laser with the following specifications
1.2 - Wavelength: 360.4 nm +/- 0.1 nm
1.3 - Power: ≥ 20 mW @ 360.4 nm
1.4 - Power stability: < 2% over 8 hours in a temperature controlled (dT < 0.5 C) laboratory environment
1.5 - Beam: free space beam in TEM00 spatial mode with M2 ≤ 1.1 and divergence ≤ 1.0 mrad
1.6 - Includes all necessary cables and heat sinks
1.7 - USB and RS-232 ports for computer control
1.8 - Compact size for integrating onto existing optics table
1.9 - Plug-and-play design with minimal to no alignment by the user
Separately Priced Options and Services:
Item 2: [List minimum specifications.
Quantity: [Enter quantity here.] Specifications: [Enter quantity here.]
[List any other item(s) or service requirements here using the same format above.]
Service Requirements / Item Specifications
Section Requirements/Specifications Planning Considerations ☒ N/A
☐ List any visits required to allow potential vendors to see site specific constraints.
[Explain requirement here] ☐ US citizen required on-site (foreign nationals require preregistration 30 days prior) ☐ Vendor rep on-site more than three days ☐ Other considerations: [List other considerations to ensure an accurate quote]
Shipping & Delivery in 06 weeks ARO preferred Shipping criteria:
ire m en ts
Sp ec ifi ca tio n Do cu m
Requirements / Specifications: AlH+ Photodissociation Laser 2
Delivery ☐ N/A
☐ Partial delivery acceptable ☐ Direct delivery to building required [Explain need for requirement here] Building has: ☐ Loading dock ☐ Containerization preference: [Choose container] Other requirements: [List any other requirements here] Delivery date criteria:
☒ Delivery of goods, installation, and training shall be completed no later than:
06 weeks ARO ☐ Deliver no earlier than: [Insert date here] ☐ Other: [List other shipping, delivery, or specil requirements]
Electronic Media ☐ N/A
[Enter electronic media requirements by item here.] ☒ Software ☒ Electronic manuals ☐ License Key ☐ Other: [List any other Electronic Media requirements here] [Enter email address for software delivery here]
Installation ☒ N/A
Vendor technician will be on site to install system and confirm performance ☐ Vendor technician on site ☐ Rigging ☐ Uncrating / Unpackaging ☐ Removal of packaging material ☐ Equipment set up ☐ Start-up services ☐ Turn key installation ☐ Other [Enter details here.] Note: When installation is required, TPOC or COR shall be the coordination point.
Facility / Utility Considerations ☒ N/A
Typical 60 Hz, 120 V power available ☐ OFPM Work Order Number(s) [List any Work Order Number(s) here] ☐ Power available Voltage: [Enter voltage here] (V) Max Current: [Enter max current draw here] (A) Phase: [Choose phase] Frequency: [Typically 60Hz] (Hz) NEMA Plug Type: [Enter plug type here] (i.e.: NEMA L5-20P) Environment: [Choose environment] Other power consideration: [List any other power considerations here] ☐ Utility connections: [List relevant available utilities, pressure(s), and flowrate(s)] ☐ Other considerations: [List other considerations to ensure an accurate quote] ire m en ts
Sp ec ifi ca tio n Do cu m
Requirements / Specifications: AlH+ Photodissociation Laser 3
Warranty ☐ N/A
Base Manufacturer Warranty is acceptable [If you require an Extended Manufacturer Warranty, describe the warranty requirements per item here.] What is included?
☐ Labor ☐ Parts ☐ Travel ☐ Cost and liability for returns ☐ Other: [Enter other requirements here] ☒ Warranty length: Base Manufacturer Period Length:[Ex. month, 6 month, year etc.]
Training ☒ N/A
Training of laboratory users in alignment, optimization and operation of the frequency doubling unit ☐ On-site ☐ Off-site location: [Enter location(s) here] ☐ Number of trainees: [Enter number of trainees here] Scope of training:
☐ Operations ☐ Maintenance ☐ Troubleshooting ☐ Safety ☐ Other: [Enter other scope requirements here] ☐ Training materials to be provided: [Describe requirement here] ☐ Other: [Enter other training requirements here] Note: When training is required, TPOC or COR shall be the coordination point.
Inspection and Acceptance ☐ N/A
[Enter inspection requirements per item here] ☐ Planned government acceptance testing (7 business days) ☒ Gov’t acceptance expected to take longer than 7 business days: 14 days ☐ Inspection and acceptance plan: [Explain here] ☐ Other: [Enter other requirements here]
Maintenance ☒ N/A
[Describe your maintenance requirements per item here] Period Length: [Ex. month, 6 month, year etc.] What is included?
☐ Labor ☐ Spare parts ☐ Consumables ☐ Travel ☐ Cost and liability for returns ☐ Software updates ☐ Manuals ☐ Service plan length: [Enter service plan length here] ☐ Other: [Enter other requirements here] ☐ Requirements: [Enter requirements here] (i.e.: onsite within 72 hours, etc.)
☐ Option Period 1 ☐ Option Period 2 ☐ Option Period 3 ☐ Option Period 4
Government Furnished Material / Equipment ☒ N/A
[List any government furnished material/equipment per item here] ☐ Samples for testing ☐ Government equipment which will leave gov't site (TPOC must coordinate with division property custodian) Serial Number: [Enter Serial Number here] Description: [Enter description here] ire m en ts
Sp ec ifi ca tio n Do cu m
Requirements / Specifications: AlH+ Photodissociation Laser 4
NIST property number: [Enter NIST property number here] ☐ Other: [Enter other details here]
Travel ☒ N/A
[Describe any travel requirements per item here] ☐ Explain: [Explain here] ☐ Other: [Enter other requirements here] Vendor: To the maximum extent possible, include as part of appropriate sections above (maintenance, repair, etc.). Otherwise, it will be a separate line item, reimbursable per the Federal Travel Regulation.
Provide suggestions to improve this form by clicking here.
File details come from the government source that posted it. Updated .